  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Sunnyvale",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "66101",
      "recalling_firm": "Bayer Healthcare Llc",
      "address_1": "510 Oakmead Pkwy",
      "address_2": "N/A",
      "postal_code": "94085-4022",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) and Internationally to China, Argentina, Uruguay, Japan.",
      "recall_number": "Z-0136-2014",
      "product_description": "Bayer brand A1CNow+ MULTI-TEST A1C SYSTEM TEST KITS (PART NUMBER 3021/08842610); Hemoglobin A1c Monitor; Glycosylated Hemoglobin Assay; Product is manufactured and distributed by   Bayer HealthCare LLC, Sunnyvale, CA       The A1CNow+ test provides quantitative measurement of the percent of glycated hemoglobin (%HbA1c or %A1C) levels in capillary (fingerstick) or venous whole blood samples. The test is for professional use to monitor glycemic control in people with diabetes.",
      "product_quantity": "138 Kits in US (Total 2452)",
      "reason_for_recall": "Test results from these lots may produce a high-bias reading (by up to 20%) that is outside of limits defined for stability and product accuracy when product is stored at room temperature four (4) to five (5) months after release.",
      "recall_initiation_date": "20130826",
      "center_classification_date": "20131101",
      "termination_date": "20140114",
      "report_date": "20131113",
      "code_info": "Lot: 1300733 (US, China);   Lot: 1300712 (Argentina, Uruguay);  Lots: 1300711, 130012 (Japan)"
    }
  ]
}