  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "West Chester",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72134",
      "recalling_firm": "Synthes (USA) Products LLC",
      "address_1": "1301 Goshen Pkwy",
      "address_2": "N/A",
      "postal_code": "19380-5986",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide Distribution",
      "recall_number": "Z-0134-2016",
      "product_description": "5.0mm Unit Rod 270mm, 5.0mm Unit Rod 290mm, 5.0mm Unit Rod 310mm,   5.0mm Unit Rod 330mm, 5.0mm Unit Rod 350mm, 5.0mm Unit Rod 370mm,   5.0mm Unit Rod 390mm, 5.0mm Unit Rod 410mm, 5.0mm Unit Rod 430mm,   5.0mm Unit Rod 450mm; Orthosis, Spinal, Pedicle fixation    Intended to provide immobilization and stabilization of spinal segments in skeletally mature patients.",
      "product_quantity": "492",
      "reason_for_recall": "This product was produced using a finishing process not identified as part of the manufacturing specification.  The process used with the lots subject to this Recall was a bead blast process. (Bead Blasting vs. Shot Peened).",
      "recall_initiation_date": "20150904",
      "center_classification_date": "20151016",
      "termination_date": "20160826",
      "report_date": "20151028",
      "code_info": "Part numbers: 298.269  298.270  298.271  298.272  298.273  298.274  298.275  298.276  298.277  298.278  lot numbers: 4729951; 4841209; 4923651; 4923652; 4987688; 4987750; 5350635; 2002330; 3000595; 4729952; 4841210; 4923653; 4923654; 4987683; 4856248; 4987755; 5153848; 5153853; 1602596; 1880489; 4729953; 4841211; 4923666; 4923667; 4987778; 4856275; 4987761; 5159826; 5157329; 3016224; 4729954; 4841212; 4919162; 4923669; 4987779; 4856250; 4987766; 5153796; 3093619;   4729956; 4923674; 4987825; 4987780; 5066363; 4729958; 4923615; 4987826; 4987781; 5066364; 4729959; 4841213; 4923655; 4936270; 4987684; 4987771; 5153849; 5153854; 4729960; 4835367; 4923656; 4923657; 4984982; 4856246; 4987772; 5159827; 5157330; 4729961; 4835368; 4923658; 4923659; 4987685; 4856249; 4987773; 1602617; 3080451; 4729962; 4835378; 4919165; 4923660; 4987686; 4856252; 4987776; and 5153851."
    }
  ]
}