  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Wilsonville",
      "state": "OR",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "69395",
      "recalling_firm": "Sam Medical Products",
      "address_1": "27350 SW 95th Ave Ste 3038",
      "address_2": "N/A",
      "postal_code": "97070-7709",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Press Release",
      "distribution_pattern": "Worldwide Distribution: US (nationwide) in states of Ohio, Oregon, New Mexico, North Carolina, Texas and to the following countries: Australia, Canada, Hong Kong, Iceland, Japan, Lebanon, Singapore, Thailand, and United Kingdom.",
      "recall_number": "Z-0133-2015",
      "product_description": "SAM Junctional Tourniquet (SJT) Auxiliary, Model SJT 102 and SJT 112.    The  SJT models specific intended uses are to control difficult bleeds in the inguinal area and to immobilize a pelvic fracture.",
      "product_quantity": "total 3633 units (3342 units in US and 291 units Outside US)",
      "reason_for_recall": "SAM Junctional Tourniquet (SJT) Auxiliary strap, model SJT 102 and SJT 112, is recalled because the snap hook on auxiliary strap broke during training demonstration.",
      "recall_initiation_date": "20140814",
      "center_classification_date": "20141024",
      "termination_date": "20150721",
      "report_date": "20141105",
      "code_info": "SJT 112 Lot number (represented as date codes): 6/5/14;    SJT 102 Lot numbers (represented as date codes):    3/29/2013, 5/15/2013, 6/11/2013, 7/16/2013, 8/27/2013, 9/3/2013, 9/10/2013, 11/4/2013, 12/9/2013, 12/11/2013, 1/10/2014, 1/30/2014, 2/25/2014, 3/4/2014, 3/5/2014, 3/19/2014, 3/21/2014, 3/24/2014, 3/31/2014, 4/1/2014, 4/2/2014, 4/4/2014, 4/7/2014, 4/9/2014, 4/11/2014, 4/22/2014, 4/28/2014, 4/29/2014, 5/5/2014, and 6/12/2014.",
      "more_code_info": ""
    }
  ]
}