  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Fairfield",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "90913",
      "recalling_firm": "Datascope Corporation",
      "address_1": "15 Law Dr",
      "address_2": "N/A",
      "postal_code": "07004-3206",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "No US distribution. Foreign distribution to Argentina, Brazil, Canada, Chile, Columbia, Ecuador, Egypt, France, Indonesia, Israel, Italy, Jordan, Kenya, Malaysia, Mexico, Netherlands, Pakistan, Poland, Portugal, South Korea, Russia, Spain, Turkmenistan, United Arab Emirates, United Kingdom, Uruguay.",
      "recall_number": "Z-0131-2023",
      "product_description": "LINEAR 7.5Fr. 40cc IAB with Accessories (APA), Model No. 0684-00-0480-02.  intra-aortic balloon catheter.",
      "product_quantity": "4414 OUS",
      "reason_for_recall": "The balloon volume is incorrectly described as 34cc instead of 40cc on one of the three internal labels enclosed for hospital use. A labelling error may result in user confusion and/or delay in initiating therapy.  Therapy delays may impact those patients who are more hemodynamically unstable.",
      "recall_initiation_date": "20220805",
      "center_classification_date": "20221024",
      "report_date": "20221102",
      "code_info": "UDI-DI: 10607567106571  Lot codes: 3000210759 3000210758 3000203002 3000198455 3000144227 3000142368 3000141069 3000140532 3000134522 3000130423 3000126591",
      "more_code_info": ""
    }
  ]
}