  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Bothell",
      "state": "WA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "88728",
      "recalling_firm": "Philips Ultrasound Inc",
      "address_1": "22100 Bothell Everett Hwy",
      "address_2": "N/A",
      "postal_code": "98021-8431",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US: MI, AZ, IA, TN, OH, CO, WI, NY, NJ, NC, LA, IL, CA, KY, IN, MD, TX, PA.    OUS: Denmark, Latvia, Belgium, Hungary, Argentina, Canada, Spain, Germany, Greece, Korea, Republic of Japan, France, Australia, Egypt, Switzerland, Italy, Senegal, Israel, Indonesia, Netherlands, Hong Kong, United Kingdom, Poland, Finland, India, New Zealand",
      "recall_number": "Z-0123-2022",
      "product_description": "EPIQ CVxi Diagnostic Ultrasound Systems, Product Code: 795232",
      "product_quantity": "107",
      "reason_for_recall": "Ultrasound system software issue can cause an EchoNavigator error notification, leading to an unresponsive system, necessitating system restart, which could lead to therapy/treatment delay and/or unnecessary therapy/treatment. Can occur during ultrasound-guided TEE with EchoNavigator, if secondary screen capture is enabled and user presses other system buttons before secondary capture completed.",
      "recall_initiation_date": "20210921",
      "center_classification_date": "20211015",
      "termination_date": "20240412",
      "report_date": "20211027",
      "code_info": "Software Versions: 6.0, 7.0, 7.0.3"
    }
  ]
}