  "results": [
    {
      "status": "Terminated",
      "city": "Brea",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "63143",
      "recalling_firm": "Beckman Coulter Inc.",
      "address_1": "250 S Kraemer Blvd",
      "address_2": "N/A",
      "postal_code": "92821-6232",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA including TX and OH. Interntionally to China and Turkey",
      "recall_number": "Z-0123-2013",
      "product_description": "UniCel¿ DxC 600/800 with  Software v5.0.11, Part Number B12218.    The UniCel¿ DxC Synchron¿ Clinical Systems are fully automated, computercontrolled clinical chemistry analyzers designed for the in vitro determination of a variety of general chemistries, therapeutic drugs, and other chemistries. Analysis can be performed on serum, plasma, urine, or cerebrospinal fluid (CSF) and whole blood (sample type is chemistry dependent).",
      "product_quantity": "7 units total (both US and International)",
      "reason_for_recall": "Beckman Coulter has confirmed that the Shuttle to Barcode Reader and Cap Piercer alignment  procedure cannot be performed on certain DxC configurations with v5.0.11 software installed. If the Shuttle to Barcode Reader becomes misaligned, the Operator will not be able to perform this  alignment procedure on the DxC systems listed above. Software v5.0 .11 does not allow access to the  Shuttle to Bar",
      "recall_initiation_date": "20120903",
      "center_classification_date": "20121024",
      "termination_date": "20140214",
      "report_date": "20121031",
      "code_info": "Part Number B12218. Serial numbers 2329, 5477, 5479."
    }
  ]
}