  "results": [
    {
      "country": "United States",
      "city": "Tarrytown",
      "address_1": "511 Benedict Ave",
      "reason_for_recall": "Customer complaints were received for positive bias reported on the controls used on the IMMULITE /IMMULITE 1000 on the Cortisol Assay reagent lots IMMULITE /IMMULITE 1000 (LKCO1) Lots 381,382 and 383 .  Subsequently, the firm confirmed a positive bias on the IMMULITE/IMMULITE 1000 (LKCO1) Lots 381, 382 and 383.",
      "address_2": "",
      "product_quantity": "US: 772 units, Canada: 43 units",
      "code_info": "Lots: 381 (exp. 10/31/2014), 382 (exp. 10/31/2014) and 383 (exp. 11/30/2014)",
      "center_classification_date": "20141023",
      "distribution_pattern": "Worldwide Distribution: US (nationwide) and internationally to Canada.",
      "state": "NY",
      "product_description": "IMMULITE /IMMULITE 1000 Systems COR Cortisol, REF/Catalog Number LKC01, SMN 10381388, IVD.    For in vitro diagnostic use with the IMMULITE 1000 Systems Analyzers for the quantitative measurement measurement of cortisol (hydrocortisone, Compound F) in serum, as an aid in the clinical assessment of adrenal status.",
      "report_date": "20141029",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Siemens Healthcare Diagnostics",
      "recall_number": "Z-0122-2015",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "69321",
      "termination_date": "20151118",
      "more_code_info": "",
      "recall_initiation_date": "20140919",
      "postal_code": "10591-5005",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}