  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Vista",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72168",
      "recalling_firm": "DJO, LLC",
      "address_1": "1430 Decision St",
      "address_2": "N/A",
      "postal_code": "92081-8553",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide in the states of WY, OH, WA, NC, FL, NE, GA, KS CA, MO, IN, MA, IL, OK, AZ, TX, AR, UT, NJ, WI, PA, SC, WV, NY, MI, MN, VA, AK, TN, OR, NC, NH, KY, CO, NV, ID and the countries of El Salvador, China, Ecuador, South Korea, Chile, Israel, Peru, Nicaragua, Bolivia, Indonesia",
      "recall_number": "Z-0121-2016",
      "product_description": "Monode, Shortwave Accessory  Model #: 02200002    Used in conjunction with the Intelect Shortwave Diathermy 100 and Intelect Shortwave Diathermy 400 units.  For the clinical application of electromagnetic energy in the radio frequency of 27.12 megahertz to generate deep heat within body tissues.",
      "product_quantity": "2887",
      "reason_for_recall": "Overheating and burning of the wire insulation inside the applicator.",
      "recall_initiation_date": "20150821",
      "center_classification_date": "20151014",
      "termination_date": "20161024",
      "report_date": "20151021",
      "code_info": "Model #: 02200002    All units received from the manufacture prior to the week of July 12, 2015"
    }
  ]
}