{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Williamston",
      "state": "MI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "83774",
      "recalling_firm": "Centurion Medical Products Corporation",
      "address_1": "100 Centurion Way",
      "address_2": "N/A",
      "postal_code": "48895-9086",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide",
      "recall_number": "Z-0120-2020",
      "product_description": "Centurion  CIRCLAMP W/1.1CM BELL - single-use, disposable instrument used to compress the foreskin of the penis during circumcision of a  male infant.  ReOrder: 310CR",
      "product_quantity": "1530 each",
      "reason_for_recall": "A  mismatch between the angle of the hole in the base and the angle of the bell, which may cause the clamp to remove foreskin without the use of a scalpel.",
      "recall_initiation_date": "20190828",
      "center_classification_date": "20191010",
      "termination_date": "20200723",
      "report_date": "20191016",
      "code_info": "Batch/Lot:  2019061801"
    }
  ]
}