  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Irvine",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "83701",
      "recalling_firm": "Horiba Instruments Incorporated",
      "address_1": "34 Bunsen",
      "address_2": "N/A",
      "postal_code": "92618-4210",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US: AL,  AZ,  CA,  CO,  CT,  FL,  GA,  IA,  ID,  IL,  KS,  KY,  LA,  MA,  MD,  MI,  MN,  MO,  MS,  MT,  NC,  NE,  NJ,  NM,  NV,  NY,  OH,  OK,  OR,  PA,  RI,  SC,  TN,  TX,  UT,  VA,  WA,  WI,  WV,  WY",
      "recall_number": "Z-0119-2020",
      "product_description": "Clinical Chemistry, ABX Pentra Lipase CP,  ABX Pentra 400/Pentrac400,  REF: A11A01631, Model/Part # 1220001631, UDI # 03610230004944",
      "product_quantity": "1539 Chemistry Analyzers",
      "reason_for_recall": "Potential for chemical analyzer to report  abnormal results  due to an interference with triglycerides CP and Cholesterol CP. An over estimation of the lipase concentration could be observed and a patient result could be declared abnormal by the instrument due to these interferences instead of normal result < 38 U/L (or < reference value used in your laboratory).",
      "recall_initiation_date": "20181030",
      "center_classification_date": "20191010",
      "termination_date": "20200616",
      "report_date": "20191016",
      "code_info": "All Lots; All Software versions"
    }
  ]
}