  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Norcross",
      "address_1": "4775 Peachtree Industrial Blvd",
      "reason_for_recall": "During the planning process the beam weights can be changed by the customers. When the beam weights are changed and renumbered and this happens during the dose calculation it is possible to save and reopen the plan without any error messages.",
      "address_2": "Bldg 300, #300",
      "product_quantity": "52 units",
      "code_info": "None as the product is software",
      "center_classification_date": "20121023",
      "distribution_pattern": "Nationwide Distribution-including the states of AL, CA, DC, KS, LA, MO, NB, NY, NC, ND, OH, OR, PA, RI, SC, TX, and WA.",
      "state": "GA",
      "product_description": "Oncentra External Beam 4.1.    Radiation treatment planning software designed to analyze and plan radiation treatments in their dimension for the purpose of treating patients with cancer.",
      "report_date": "20121031",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Elekta, Inc.",
      "recall_number": "Z-0118-2013",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "63405",
      "termination_date": "20141103",
      "recall_initiation_date": "20120821",
      "postal_code": "30092-3011",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}