  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Santa Clara",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "90840",
      "recalling_firm": "Abbott Vascular",
      "address_1": "3200 Lakeside Dr",
      "address_2": "N/A",
      "postal_code": "95054-2807",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide including in the states of PR, WA, NE, MO, IA, KS, TN, MN, ND, MD, VA, MA, PA, DC, LA, KY, CA, NV, IL, WI, TX, CT, VT, AL, AR, GA, MS, MI, OH, FL, NC, NJ, CO, NY, DE, OR, SC, IN, HI, AZ, NH, ME, WV, OK, ID, UT, MT, NM, RI, AK, SD and the countries of AE, AR, AT, AU, BE, BG, BR, BY, CA, CH, CL, CN, CO, CR, CY, CZ, DE, DK, EC, EE, EG, ES, FI, FR, GB, GR, HK, HU, ID, IE, IL, IN, IR, IT, JO, JP, KR, KW, LB, LU, MT, MX, MY, NL, NO, NZ, OM, PE, PH, PK, PL, PT, QA, RE, RO, RU, SA, SE, SG, SI, SK, TH, TN, TR, TW.",
      "recall_number": "Z-0117-2023",
      "product_description": "Product/Part: MitraClip G4 CDS NT REPAIR-MR Clinical/CDS0704-NT,  MitraClip G4 CDS NTW REPAIR-MR Clinical/CDS0704-NTW,  MitraClip G4 CDS XT REPAIR-MR Clinical/CDS0704-XT,  MitraClip G4 CDS XTW REPAIR-MR Clinical/CDS0704-XTW,  TriClip NT Clip Delivery Sys CLIN/TCDS0203-NT,  TriClip XT Clip Delivery Sys CLIN/TCDS0203-XT,  TriClip G4 NT Delivery System CLIN/TCDS0303-NT,  TriClip G4 NTW Delivery System CLIN/TCDS0303-NTW,  TriClip G4 XT Delivery System CLIN/TCDS0303-XT,  TriClip G4 XTW Delivery System CLIN/TCDS0303-XTW,  TriClip NT Clip Delivery Sys CLIN/TCDS0203-NT,  TriClip XT Clip Delivery Sys CLIN/TCDS0203-XT.  For cardiac procedures.",
      "product_quantity": "767",
      "reason_for_recall": "An increase in clip locking malfunctions, Establish Final Arm Angle and Clip Opening While Locked, have been observed. Contributing causes may be clip locking component material properties and certain use conditions. Unsuccessful clip procedure can lead to recurrence of severe mitral regurgitation, device embolization, or tissue injury.",
      "recall_initiation_date": "20220908",
      "center_classification_date": "20221020",
      "report_date": "20221026",
      "code_info": "Part/UDI-GTIN: CDS0704-NT/08717648234576, CDS0704-NTW/08717648234590, CDS0704-XT/08717648234583, CDS0704-XTW/08717648234606, TCDS0203-NT/08717648229992, TCDS0203-XT/08717648229961, TCDS0303-NT/08717648334276, TCDS0303-NTW/08717648334290, TCDS0303-XT/08717648334283, TCDS0303-XTW/08717648334306, TCDS0203-NT/08717648229992, TCDS0203-XT/08717648229961  All lots.",
      "more_code_info": ""
    }
  ]
}