  "results": [
    {
      "status": "Terminated",
      "city": "Plainsboro",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "66243",
      "recalling_firm": "Integra Limited",
      "address_1": "311 Enterprise Dr",
      "address_2": "N/A",
      "postal_code": "08536-3344",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide Distribution",
      "recall_number": "Z-0117-2014",
      "product_description": "Brand Names: Integra¿ Camino 110-4 Intracranial Pressure (ICP) Monitoring  Catheters.  Product / Catalogue No.s: 110-4B, 110-4BC, 110-4G, 110-4H, 110-4HC, 110-4L, 110-4LC.  The fully packaged 110-4B, 110-4H, and 110-4L catheters are  also provided as a component in kits that contain other cranial  access devices. These kits are Catalogue No. 110-4BC, 110-  4HC, and 110-4LC respectively.    Common Name: Intracranial Pressure and Pressure-Temperature Monitoring Kits  Classification Name: Device, Monitoring, Intracranial Pressure  510(k) Number: K102875    The Camino 110-4 Intracranial Pressure monitoring catheters are sterile transducer-tipped pressure monitoring catheters that are used by neurosurgeons to rapidly determine and continuously monitor intracranial pressure or pressure/temperature. \u0001 The Camino 110-4B catheter is used for Intracranial Pressure Monitoring in the parenchyma or subarachnoid space. \u0001 The Camino 110-4G catheter is used for Post Craniotomy Subdural Pressure Monitoring. \u0001 The Camino 110-4H catheter is intended for Intracranial Pressure Monitoring and Cerebrospinal fluid drainage. \u0001 The Camino 110-4L catheter is intended for use with the Licox¿ Brain Oxygen Monitoring System to measure intracranial pressure in the parenchyma. \u0001 The fully packaged 110-4B, 110-4H, and 110-4L catheters are also provided as a component in kits that contain other cranial access devices. These kits are Catalogue No. 110-4BC, 110- 4HC, and 110-4LC respectively.",
      "product_quantity": "79,887",
      "reason_for_recall": "Integra LifeSciences has identified through an internal evaluation that in order to comply with US FDA  requirements, certain Camino¿ Intracranial Pressure Monitoring Kits require additional labelling to indicate  they are UNSAFE in an MR environment.  Only the following Camino¿ Intracranial Pressure Monitoring Kits are affected:  110-4B, 110-4BC, 110-4G, 110-4H, 110-4HC, 110-4L, 110-4LC.",
      "recall_initiation_date": "20130910",
      "center_classification_date": "20131030",
      "termination_date": "20140604",
      "report_date": "20131106",
      "code_info": "Product / Catalogue No.s: 110-4B, 110-4BC, 110-4G, 110-4H, 110-4HC, 110-4L, 110-4LC.    Lot Numbers:  Catalog Number 110-4B: 3050RX195484, 3050RX198473, 305000205426, 3050RX197641, 3050RX201421, 305000205648, 305000205652, 3050RX199149, 3050RX201407, 3050RX199850, 305000205976, 3050RY189979, 3050RX200862, 305000206178, 3050RY193648, 3050RX200486, 3050RX181026, 3050RX185115, 3050RY195484, 305000207061, 305000207062, 3050RY198473, 305000207330, 305000207334, 3050RY197641, 3050RY200862, 305000207580, 305000207892, 3050RY200486, 3050RY181026, 3050RY185115, 3050RX207330, 3050RX207334, 3050RX207580, 305000208008, 305000208510, 3050RY207892, 305000210060, 3050RX208008, 3050RX208510, 3050RX205976, 3050RX206178, 305000210563, 305000211723, 3050RX210060, 305000212616, 305000213007, 305000213191, 30500X210563, 3050RX207061, 305000214590, 3050RX205426, 30500X215502, 3050RX207062, 305000215531, 30500X216410, 3050RY199850, 305000216132, 30500X216627, 30500X217119, 3050RY217112, 305000216718, 3050RX217751, 3050RX217840, 3050RY217837, 305000217858, 30500X218369, 30500X218391, 30500X219054, 3050RY206178, 305000220708, 305000218578, 30500X220430, 305000219261, 30500X221221, 305000221226, 305000219989, 30500X222569, 305000221953, 305000221955, 30500X223153, 30500X223709, 305000223275, 305000223280, 305000222640, 30500X224142, 305000223867, 305000223866, 30500X224692, 305000224352, 305000224353, 30500X225386, 305000225045, 305000225044, 30500X225992, 305000225632, 305000225633, 30500X226540, 30500X227059, 305000226844, 305000226846, 305000226127, 30500X227564, 305000226836, 30500X227820, 305000227127, 305000227128, 30500X228194, 305000227790, 305000227789, 30500X228787, 305000228522, 305000228521, 305000228971, 305000228972, 30500X229560, 305000229328, 305000229329, 30500X229766, 30500X230087, 305000229825, 305000229826, 30500X230565, 305000230303, 305000230302, 30500X231053, 305000230780, 305000230781, 30500X231786, 305000231461, 305000231462, 30500X232331, 305000232098, 305000232099, 30500X232960, 305000232605, 305000232604, 305000233245, 305000233244, 305000233243, 305000233242, 305000233241, 305000233240, 30500X233415, 30500X234183, 305000233607, 305000233608, 30500X234806, 305000234303, 305000234304, 305000235022, 305000235017, 30500X235732, 305000235598, 30500X236328, 305000236105, 305000236104, 305000236655, 305000236653, 30500Y237087, 30500X237090, 305000237232, 30500X237761, 305000237824, 305000237825, 30500X238273, 305000238263, 305000238264, 30500X239095, 305000238818, 305000238819, 30500X239497, 30500X239748, 30500X239958, 30500X240334, 30500X240668, 30500X241368, 305000241516, 305000242236, 305000242792, 305000242795, 305000242796, 305000242794, 305000242793, 305000242791, 3050RY243869, 305000243251, 305000243882, 305000243881, 305000243888, 305000243889, 305000243883, 30500X244281, 30500X244880, 305000244513,30500X245306, 305000244481, 305000244480, 305000244934, 30500X245937, 30500X246067, 305000245408, 30500X246872, 305000245835, 30500X246951, 30500X247608, 305000246460, 30500X248284, 305000247007, 3050RX227127, 3050RX234303, 305000247717, 305000249072, 3050RX242794, 3050RX242793, 3050RX237825, 3050RX241516, 30500X249488, 305000248479, 30500X250423, 305000249743, 30500X250964, 305000250456, 30500X251060, 305000251055, 30500X251560, 30500X252133, 3050RY252136, 305000251453, 30500X252771, 305000252058, 3050RX244513, 305000252646, 305000253785, 30500X254929, 3050RY255145, 30500X255141, 305000254589, 305000255549, 305000255551, 305000255038, 30500X256614, 305000256192, 305000257281, 305000256709, 305000257280, 305000258048, 30500X259385, 305000258379, 305000258921, 305000258922, 30500X259557, 305000259431, 305000259432, 30500X260159, 305000259966, 305000259960, 305000260514, 305000260527, 30500Y261748, 3050RX253785, 305000261325, 305000261326, 305000261737, 305000261736, 3050RX256192, 30500X262441,   305000262328, 305000262327, 305000262978, 305000262977, 3050RY263508, 30500X263559, 305000263504, 305000264073, 305000264074, 305000264625, 305000264624, 3050RY265348, 30500X265422, 30500X265976, 3050RY265975, 305000265530, 305000265527, 30500X266142, 305000266134, 305000266135, 30500X266789, 305000267193, 305000267194, 3050RX261325, 30500X267468, 305000267421, 305000267422, 30500X267990, 305000267673, 305000267674, 305000268294, 305000268293, 305000266784, 305000268914, 305000268915, 3050RX261736, 3050RY269955, 30500X269963, 305000269220, 305000269221, 305000270081, 3050RY271045, 30500X271047, 305000271042, 305000271716, 305000271715, 30500X272367, 305000272273, 305000272274, 305000272852, 305000272851, 30500X274204, 305000273466, 305000273463, 305000274200, 305000274201, 30500X275297, 30500X275393, 305000274816,  30500X276594, 3050RX268914, 305000277761, 3050RY282196, 305000282304, 30500X282815, 3050RY283371, 3050RX283315, 3050RX283314      Catalog Number 110-4G: 3050RX195650, 3050RX198690, 3050RY193969, 3050RY197837, 3050RX199151, 3050RX199942, 3050RX201552, 3050RY185259, 3050RX200538, 305000206177, 3050RX200989, 305000206556, 305000206557, 3050RX200990, 3050RY199151, 3050RY199942, 305000207465, 3050RY200538, 3050RY200989, 305000208007, 3050RY200990, 3050RX207465, 3050RX206557, 3050RX206556, 305000209417, 305000209105, 3050RX208007, 305000210891, 305000211419, 3050RX206177, 30500X209105, 3050RX209417, 305000212298, 30500X210891, 305000213240, 305000214103, 30500X215501, 30500X215577, 3050RX202201, 305000215628, 3050RY195650, 3050RY198690, 30500X216621, 305000216246,3050RY217113, 30500X217118, 305000217140, 30500X217835, 3050RY217839, 305000217464, 30500X218393, 305000218009, 30500X219053, 305000218225, 305000219118, 30500X220001, 30500X220429, 305000219990, 30500X221225, 305000221227, 30500X222567, 305000222642, 305000223279, 30500X223707, 305000223868, 30500X224685, 305000224354, 30500X225383, 305000225046, 305000225634, 30500X226000, 30500X226541, 305000226731, 305000226869, 305000226867, 30500X227567, 305000227129, 30500X228784, 305000228523, 305000228973, 30500X229561, 305000229330, 30500X230084, 305000229827, 305000230304, 30500X231054,305000230782, 30500X231787, 305000231463, 30500X232333, 305000232101, 30500X232776, 305000232606, 30500X233419, 305000233246, 305000233609, 30500X234690, 30500X234807, 305000234306, 305000235566, 30500X236329, 30500Y237088, 30500X237767, 305000237233, 305000238820, 30500X239496, 305000239476, 30500X240336, 30500X241357, 305000241518, 30500X242172, 305000242249, 305000243252, 305000244364, 30500X245304, 305000244935, 305000245409, 305000245838, 30500X246522, 305000246308, 30500X246873, 30500X247609, 30500X248281, 305000249073, 3050RX241518, 3050RX233246, 3050RX242249, 3050RX228973, 30500X251061, 305000251056, 30500X251563, 3050RY252137, 305000252059, 305000252648, 305000254591, 30500X254931, 30500X255143, 305000255039, 305000255548, 305000256193, 305000256737, 305000257244, 305000257246, 305000257921, 305000257920, 305000258303, 305000258290, 30500X259383, 305000258793, 305000259573, 30500X260163, 305000259958, 305000259959, 3050RX254591, 3050RX255039, 305000260513, 305000260512, 30500Y261771, 3050RX257244, 3050RX257920, 305000261159, 305000261160, 3050RY262444, 30500X262443, 305000261734, 305000261732, 305000262325, 305000262326, 3050RX258303, 305000262976, 305000262975, 3050RY263510, 30500X263560, 305000263507, 305000263506, 305000264075, 305000264076, 305000264627, 305000264628, 30500X265424, 305000265525, 305000265524, 30500X265978, 30500X266143, 305000266063, 305000266064, 30500X266787, 305000266638, 305000266637, 305000267191, 30500X267470, 305000267423, 305000267675, 3050RY269957, 30500X274206    Catalog Number 110-4H: 3050RX198281, 3050RY194144, 3050RY198281, 3050RY200861, 30500X223154, 305000220948, 305000223622, 30500X225385, 30500X233417, 305000231714, 305000241350, 30500X243874, 305000250959, 30500X254932, 305000263557    Catalog Number 110-4L: 305000221383, 30500X222566, 305000222762, 30500X223971,305000223969, 305000225078, 30500X226539, 305000226221, 30500X227058, 305000226848, 305000226749, 305000226847, 305000226871, 305000226,870, 30500X228783, 305000229450, 305000229909, 305000230871  30500X232775, 305000232199, 30500X232961, 305000233311, 305000234438, 30500X235731, 305000235181, 305000236201, 305000237368, 305000237369, 305000237799, 305000238539, 305000239102, 30500X240245, 30500X240246, 30500X240247, 305000239736, 305000239737, 30500X241365, 30500X241522, 305000241082, 30500X242170, 30500X245765, 30500X246521, 3050RX237799, 3050RX226749, 3050RX229450, 3050RX241082, 3050RX232199, 3050RY252775, 3050RX240446, 3050RY255148, 305000259553, 305000260626, 305000261835, 305000262436, 3050RX259553, 305000263146, 3050RX260626, 305000264077, 3050RX262436, 3050RX261835, 305000264175, 3050RY265351, 3050RY265766, 305000265531, 30500X266144, 3050RY266146, 30500X266522, 305000266062, 305000267297, 30500X267473, 30500X268133, 305000267784, 305000268412, 3050RX265531, 305000268916, 3050RY271046, 305000271872, 30500X272374, 305000272591, 30500X274210, 30500X274963, 305000281820     Catalog Number 110-4BC: 176044, 177018, 178013, 178898, 179757, 180139, 180955, 182279, 182908, 184079, 184687, 185284, 186277, 187293, 190452, 187780, 190451, 191891, 193414, 194074, 195161, 195608, 196987, 197492, 198189, 198512, 200069, 200490, 199609, 201039, 201595, 202423, 203565, 207077, 208532    Catalog Number 110-4HC: 179758, 180956, 183071,   184686, 186278, 187781, 190450, 192246, 194993, 195450, 198191, 199708, 201040, 201596, 202009     Catalog Number 110-4LC: 162523, 167450, 167786, 175720, 177188, 178547, 182130, 187495, 195245, 197493."
    }
  ]
}