{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Hoffman Estates",
      "address_1": "2501 Barrington Rd",
      "reason_for_recall": "Reports of loose bolts in the detector support structure of Symbia and e.cam systems.",
      "address_2": "",
      "product_quantity": "4,862 devices",
      "code_info": "*** 1) Part Number: 10151531; Serial Numbers: 1002 ,  1003 ,  1004 ,  1005 ,  1006 ,  1007 ,  1008 ,  1009 ,  1010 ,  1011 ,  1012 ,  1013 ,  1014 ,  1015 ,  1016 ,  1017 ,  1018 ,  1019 ,  1020 ,  1021 ,  1022 ,  1023 ,  1024 ,  1025 ,  1026 ,  1027 ,  1028 ,  1029 ,  1030 ,  1031 ,  1032 ,  1033 ,  1034 ,  1035 ,  1036 ,  1037 ,  1038 ,  1039 ,  1040 ,  1041 ,  1042 ,  1043 ,  1044 ,  1045 ,  1046 ,  1047 ,  1048 ,  1049 ,  1050 ,  1051 ,  1052 ,  1053 ,  1054 ,  1055 ,  1056 ,  1057 ,  1058 ,  1059 ,  1060 ,  1061 ,  1062 ,  1063 ,  1064 ,  1065 ,  1066 ,  1067 ,  1068 ,  1069 ,  1070 ,  1071 ,  1072 ,  1073 ,  1074 ,  1075 ,  1076 ,  1077 ,  1078 ,  1079 ,  1080 ,  1081 ,  1082 ,  1083 ,  1084 ,  1085 ,  11128 ,  11131 ,  11144 ,  70001 ,  70002 ;     *** 2) Part Number: 10151532; Serial Numbers: 1001 ,  1003 ,  1004 ,  1005 ,  1006 ,  1007 ,  1008 ,  1009 ,  1010 ,  1011 ,  1013 ,  1014 ,  1015 ,  1016 ,  1017 ,  1018 ,  1019 ,  1020 ,  1021 ,  1022 ,  1023 ,  1024 ,  1025 ,  1026 ,  1027 ,  1028 ,  1029 ,  1030 ,  1031 ,  10319 ,  1032 ,  1033 ,  1034 ,  1035 ,  1036 ,  1037 ,  1038 ,  1039 ,  1040 ,  1041 ,  1042 ,  1043 ,  1044 ,  1045 ,  1047 ,  1048 ,  1049 ,  1050 ,  1051 ,  1052 ,  1053 ,  1054 ,  1055 ,  1056 ,  1057 ,  1058 ,  1059 ,  1060 ,  1061 ,  1062 ,  1063 ,  1064 ,  1065 ,  1066 ,  1067 ,  1068 ,  1069 ,  1070 ,  1071 ,  1072 ,  1073 ,  1074 ,  1075 ,  1076 ,  1077 ,  1078 ,  1079 ,  1080 ,  1081 ,  1082 ,  1083 ,  1084 ,  1085 ,  1086 ,  1087 ,  1088 ,  1089 ,  1090 ,  1091 ,  1092 ,  1093 ,  1094 ,  1095 ,  1096 ,  1097 ,  1098 ,  1099 ,  1100 ,  1101 ,  1102 ,  1103 ,  1104 ,  1105 ,  1106 ,  1107 ,  1108 ,  1109 ,  1110 ,  1111 ,  1112 ,  1113 ,  1114 ,  1115 ,  1116 ,  1117 ,  1118 ,  1119 ,  1120 ,  1121 ,  1122 ,  1123 ,  1124 ,  1125 ,  1126 ,  1127 ,  1128 ,  1129 ,  1130 ,  1131 ,  1132 ,  1133 ,  1134 ,  1135 ,  1136 ,  1137 ,  1138 ,  1139 ,  1140 ,  1141 ,  1142 ,  1143 ,  1144 ,  1145 ,  1146 ,  1147 ,  1148 ,  1149 ,  1150 ,  1151 ,  1152 ,  1153 ,  1154 ,  1155 ,  1156 ,  1157 ,  1158 ,  1159 ,  1160 ,  1161 ,  1162 ,  1163 ,  1164 ,  1165 ,  1166 ,  1167 ,  1168 ,  1169 ,  1170 ,  1171 ,  1172 ,  1173 ,  1174 ,  1175 ,  1176 ,  1177 ,  1178 ,  1179 ,  1180 ,  1181 ,  1182 ,  1183 ,  1184 ,  1185 ,  1186 ,  1187 ,  1188 ,  1189 ,  1190 ,  1191 ,  1192 ,  1193 ,  1194 ,  1195 ,  1196 ,  1198 ,  1199 ,  1200 ,  1201 ,  1202 ,  1203 ,  1204 ,  1205 ,  1206 ,  1207 ,  1208 ,  1209 ,  1210 ,  1211 ,  1212 ,  1213 ,  1214 ,  1215 ,  1216 ,  1217 ,  1218 ,  1219 ,  1220 ,  1221 ,  1222 ,  1223 ,  1224 ,  1225 ,  1226 ,  1227 ,  1228 ,  1229 ,  1230 ,  1231 ,  1232 ,  1233 ,  1234 ,  1235 ,  1236 ,  1237 ,  1238 ,  1239 ,  1240 ,  1241 ,  1242 ,  1243 ,  1244 ,  1245 ,  1246 ,  1247 ,  1248 ,  1249 ,  1250 ,  1251 ,  1252 ,  1253 ,  1254 ,  1255 ,  1256 ,  1257 ,  1258 ,  1259 ,  1260 ,  1261 ,  1263 ,  1264 ,  1265 ,  1266 ,  1267 ,  1268 ,  1269 ,  1270 ,  1271 ,  1272 ,  1273 ,  1274 ,  1275 ,  1276 ,  1277 ,  1278 ,  1279 ,  1280 ,  1281 ,  1282 ,  1283 ,  1284 ,  1285 ,  1286 ,  1287 ,  1288 ,  1289 ,  1290 ,  1291 ,  1292 ,  1293 ,  1294 ,  1295 ,  1296 ,  1297 ,  1298 ,  1299 ,  1300 ,  1301 ,  1302 ,  1303 ,  1304 ,  1305 ,  1306 ,  1307 ,  1308 ,  1309 ,  1310 ,  1311 ,  1312 ,  1313 ,  1314 ,  1315 ,  1316 ,  1317 ,  1318 ,  1319 ,  1320 ,  1321 ,  1322 ,  1323 ,  1324 ,  1325 ,  1326 ,  1327 ,  1328 ,  1329 ,  1330 ,  1331 ,  1332 ,  1333 ,  1334 ,  1335 ,  1336 ,  1337 ,  1338 ,  1339 ,  1340 ,  1341 ,  1342 ,  1343 ,  1344 ,  1345 ,  1346 ,  1347 ,  1348 ,  1349 ,  1350 ,  1351 ,  1352 ,  1353 ,  1354 ,  1355 ,  1356 ,  1357 ,  1358 ,  1359 ,  1360 ,  1361 ,  1362 ,  1363 ,  1364 ,  1365 ,  1366 ,  1367 ,  1368 ,  1369 ,  1370 ,  1371 ,  1372 ,  1373 ,  1374 ,  1375 ,  2107 ,  70001 ,  70002 ,  70003 ,  70004 ,  70005 ,  70006 ,  70007 ,  70008 ,  70009 ,  7197 ,  9989 ;     *** 3) Part Number: 4380213; Serial Numbers: 1 ,  1002 ,  1003 ,  1004 ,  1005 ,  1006 ,  1007 ,  1008 ,  1009 ,  1010 ,  1011 ,  1012 ,  1013 ,  1014 ,  1015 ,  1016 ,  1017 ,  1018 ,  1019 ,  1020 ,  1021 ,  1022 ,  1065 ,  2 ,  2001 ,  2002 ,  2003 ,  2004 ,  2005 ,  2006 ,  2007 ,  2008 ,  2009 ,  2010 ,  2011 ,  2012 ,  2013 ,  2014 ,  2015 ,  2016 ,  2017 ,  2018 ,  2019 ,  2020 ,  2021 ,  2022 ,  2023 ,  2024 ,  2025 ,  2026 ,  2027 ,  2028 ,  3001 ,  3002 ,  3003 ,  3004 ,  3005 ,  3006 ,  3007 ,  3008 ,  3009 ,  3010 ,  3011 ,  3012 ,  3013 ,  3014 ,  3015 ,  3016 ,  3017 ,  3018 ,  3019 ,  3020 ,  3021 ,  3022 ,  3023 ,  3024 ,  3024// ,  3025 ,  3026 ,  3027 ,  3028 ,  3029 ,  3030 ,  3030// ,  3031 ,  3032 ,  3033 ,  3034 ,  4 ,  4001 ,  4002 ,  4003 ,  4004 ,  4005 ,  4006 ,  4007 ,  4008 ,  4009 ,  4010 ,  4011 ,  4012 ,  4013 ,  4014 ,  4015 ,  4016 ,  4219 ,  49518 ,  5 ,  6104 ,  6134R ,  6303 ,  7912 ,  96302 ,  9965 ;     *** 4) Part Number: 4380221; Serial Numbers: 01008A ,  1002 ,  1004 ,  1005 ,  1006 ,  1007 ,  1008 ,  1009 ,  1010 ,  1011 ,  1012 ,  1013 ,  1014 ,  1015 ,  1016 ,  1017 ,  1018 ,  1019 ,  1020 ,  1021 ,  1022 ,  1023 ,  1024 ,  1025 ,  1026 ,  1027 ,  1028 ,  1029 ,  1030 ,  1031 ,  1032 ,  1033 ,  1034 ,  1035 ,  10351 ,  10352 ,  1036 ,  1037 ,  1038 ,  1039 ,  1040 ,  1041 ,  1042 ,  1043 ,  1044 ,  1045 ,  1046 ,  1047 ,  1048 ,  1049 ,  1051 ,  1052 ,  1053 ,  1054 ,  1055 ,  1056 ,  1057 ,  1058 ,  1059 ,  1060 ,  1061 ,  1062 ,  1063 ,  1064 ,  1065 ,  1066 ,  1067 ,  1068 ,  1069 ,  1070 ,  1071 ,  1072 ,  1073 ,  1074 ,  1075 ,  1076 ,  1077 ,  1078 ,  1079 ,  1080 ,  1081 ,  1082 ,  1084 ,  1085 ,  1086 ,  1087 ,  1088 ,  1089 ,  1090 ,  1091 ,  1092 ,  1093 ,  1094 ,  1095 ,  1096 ,  1097 ,  1098 ,  1099 ,  11 ,  1100 ,  1101 ,  1102 ,  1103 ,  1104 ,  1105 ,  1106 ,  1107 ,  1108 ,  1109 ,  1111 ,  1112 ,  1113 ,  1114 ,  1115 ,  1116 ,  1117 ,  1118 ,  1119 ,  1120 ,  1121 ,  1122 ,  1123 ,  1124 ,  1125 ,  1126 ,  1127 ,  1128 ,  1129 ,  1130 ,  1132 ,  1133 ,  1134 ,  1135 ,  1136 ,  1137 ,  1140 ,  1141 ,  1142 ,  1143 ,  1144 ,  1145 ,  1146 ,  1147 ,  1148 ,  1149 ,  1150 ,  1151 ,  1152 ,  1153 ,  1154 ,  1155 ,  1156 ,  1157 ,  1158 ,  1159 ,  1160 ,  1161 ,  1162 ,  1163 ,  1164 ,  1165 ,  1166 ,  1167 ,  1169 ,  1170 ,  1171 ,  1172 ,  1174 ,  1176 ,  1180 ,  1181 ,  1182 ,  1183 ,  1184 ,  1185 ,  1186 ,  1187 ,  1188 ,  1189 ,  1190 ,  1191 ,  1192 ,  1193 ,  1195 ,  1196 ,  1197 ,  1198 ,  1199 ,  1200 ,  1201 ,  1202 ,  1203 ,  1204 ,  1205 ,  1206 ,  1207 ,  1208 ,  1209 ,  1211 ,  1212 ,  13 ,  2050 ,  2089 ,  2100 ,  3005 ,  3066 ,  3111 ,  3125 ,  31827 ,  4003 ,  4052 ,  4069 ,  4081 ,  4117 ,  4166 ,  49641 ,  5 ,  5009 ,  5090 ,  5145 ,  7030 ,  7582 ,  8149 ,  8462 ,  8635 ,  86439 ,  9 ,  9601 ,  NU6457-01 ,  O1135 ;     *** 5) Part Number: 5242826; Serial Numbers: 01009R ,  1001 ,  1002 ,  1003 ,  1004 ,  1005 ,  1006 ,  1007 ,  1009 ,  1010 ,  1011 ,  1012 ,  1013 ,  1014 ,  1015 ,  1016 ,  1018 ,  1019 ,  1020 ,  1021 ,  1022 ,  1023 ,  1024 ,  1025 ,  1026 ,  1027 ,  1028 ,  1029 ,  1030 ,  1031 ,  1032 ,  1034 ,  1035 ,  1036 ,  1037 ,  1038 ,  1039 ,  1040 ,  1041 ,  1042 ,  1043 ,  1044 ,  1045 ,  1046 ,  1047 ,  1048 ,  1049 ,  1050 ,  1051 ,  1052 ,  1054 ,  1056 ,  1057 ,  1058 ,  1059 ,  1060 ,  1062 ,  1063 ,  1064 ,  1065 ,  1066 ,  1067 ,  1068 ,  1069 ,  1070 ,  1071 ,  1072 ,  1073 ,  1073DEL ,  1074 ,  1075 ,  1076 ,  1077 ,  1078 ,  1079 ,  1080 ,  1081 ,  1082 ,  1083 ,  1084 ,  1087 ,  1088 ,  1089 ,  1090 ,  1091 ,  1093 ,  1094 ,  1095 ,  1097 ,  1099 ,  1100 ,  1101 ,  1102 ,  1103 ,  1104 ,  1105 ,  1107 ,  1109 ,  1110 ,  1111 ,  1112 ,  1113 ,  1114 ,  1116 ,  1117 ,  1118 ,  1119 ,  1121 ,  1129 ,  1130 ,  1131 ,  1134 ,  1135 ,  1136 ,  1137 ,  1138 ,  1139 ,  1140 ,  1141 ,  1142 ,  1143 ,  1144 ,  1203 ,  1210 ,  2001 ,  2002 ,  2003 ,  2004 ,  2005 ,  2006 ,  2007 ,  2008 ,  2009 ,  2010 ,  2011 ,  2012 ,  2013 ,  2014 ,  2015 ,  2016 ,  2017 ,  2018 ,  2019 ,  2020 ,  2021 ,  2022 ,  2023 ,  2024 ,  2025 ,  2026 ,  2027 ,  2028 ,  2029 ,  2030 ,  2031 ,  2032 ,  2034 ,  2035 ,  2036 ,  2037 ,  2038 ,  2039 ,  2040 ,  2041 ,  2042 ,  2043 ,  2044 ,  2045 ,  2046 ,  2047 ,  2048 ,  2049 ,  2050 ,  2051 ,  2052 ,  2053 ,  2054 ,  2055 ,  2056 ,  2057 ,  2058 ,  2059 ,  2060 ,  2061 ,  2062 ,  2063 ,  2064 ,  2065 ,  2065XX ,  2066 ,  2067 ,  2068 ,  2069 ,  2070 ,  2071 ,  2072 ,  2073 ,  2074 ,  2075 ,  2076 ,  2077 ,  2078 ,  2079 ,  2080 ,  2081 ,  2082 ,  2083 ,  2084 ,  2085 ,  2086 ,  2089 ,  2090 ,  2091 ,  2092 ,  2093 ,  2094 ,  2095 ,  2096 ,  2097 ,  2098 ,  2099 ,  2100 ,  2101 ,  2102 ,  2107 ,  2108 ,  2109 ,  2110 ,  2111 ,  2114 ,  2115 ,  2117 ,  2118 ,  2119 ,  2120 ,  2121 ,  2122 ,  2123 ,  2124 ,  2125 ,  2126 ,  2127 ,  2128 ,  2129 ,  2130 ,  3001 ,  3002 ,  3003 ,  3004 ,  3005 ,  3006 ,  3007 ,  3008 ,  3009 ,  3010 ,  3011 ,  3012 ,  3013 ,  3014 ,  3015 ,  3016 ,  3017 ,  3018 ,  3019 ,  3020 ,  3021 ,  3022 ,  3023 ,  3024 ,  3025 ,  3026 ,  3027 ,  3028 ,  3029 ,  3030 ,  3031 ,  3032 ,  3033 ,  3034 ,  3035 ,  3036 ,  3037 ,  3038 ,  3039 ,  3040 ,  3041 ,  3043 ,  3044 ,  3045 ,  3046 ,  3047 ,  3048 ,  3049 ,  3050 ,  3051 ,  3052 ,  3053 ,  3054 ,  3055 ,  3056 ,  3057 ,  3058 ,  3059 ,  3060 ,  3061 ,  3062 ,  3063 ,  3064 ,  3065 ,  3066 ,  3067 ,  3068 ,  3069 ,  3070 ,  3070// ,  3071 ,  3072 ,  3073 ,  3074 ,  3075 ,  3076 ,  3077 ,  3078 ,  3079 ,  3080 ,  3081 ,  3082 ,  3083 ,  3084 ,  3085 ,  3086 ,  3087 ,  3088 ,  3089 ,  3090 ,  3091 ,  3092 ,  3093 ,  3094 ,  3095 ,  3096 ,  3097 ,  3098 ,  3099 ,  3100 ,  3101 ,  3102 ,  3103 ,  3104 ,  3105 ,  3106 ,  3107 ,  3108 ,  3109 ,  3110 ,  3111 ,  3112 ,  3113 ,  3114 ,  3115 ,  3116 ,  3117 ,  3118 ,  3119 ,  3120 ,  3121 ,  3122 ,  3123 ,  3124 ,  3125 ,  3126 ,  3127 ,  3128 ,  3129 ,  3130 ,  3131 ,  3132 ,  3133 ,  3134 ,  3135 ,  3136 ,  3137 ,  3138 ,  3139 ,  3140 ,  3141 ,  4001 ,  4002 ,  4003 ,  4004 ,  4005 ,  4006 ,  4007 ,  4008 ,  4009 ,  4010 ,  4011 ,  4012 ,  4013 ,  4014 ,  4015 ,  4016 ,  4017 ,  4018 ,  4019 ,  4020 ,  4021 ,  4022 ,  4023 ,  4024 ,  4025 ,  4026 ,  4027 ,  4028 ,  4029 ,  4030 ,  4031 ,  4032 ,  4033 ,  4034 ,  4035 ,  4036 ,  4037 ,  4038 ,  4039 ,  4040 ,  4041 ,  4042 ,  4043 ,  4044 ,  4045 ,  4046 ,  4047 ,  4048 ,  4049 ,  4049XXX ,  4050 ,  4051 ,  4052 ,  4053 ,  4054 ,  4055 ,  4056 ,  4057 ,  4058 ,  4059 ,  4060 ,  4061 ,  4062 ,  4063 ,  4064 ,  4065 ,  4066 ,  4067 ,  4068 ,  4069 ,  4070 ,  4071 ,  4072 ,  4073 ,  4074 ,  4075 ,  4076 ,  4077 ,  4078 ,  4090 ,  4111 ,  7501 ,  CAR-557 ,  SHOW1 ;     *** 6) Part Number: 5977066; Serial Numbers: 1006 ,  2002 ,  4079 ,  4080 ,  4081 ,  4082 ,  4083 ,  4084 ,  4084A ,  4085 ,  4086 ,  4087 ,  4088 ,  4089 ,  4091 ,  4092 ,  4093 ,  4094 ,  4095 ,  4096 ,  4097 ,  4098 ,  4099 ,  4100 ,  4101 ,  4102 ,  4103 ,  4104 ,  4105 ,  4106 ,  4107 ,  4108 ,  4109 ,  4110 ,  4112 ,  4113 ,  4114 ,  4115 ,  4116 ,  4117 ,  4118 ,  4119 ,  4120 ,  4121 ,  4122 ,  4123 ,  4123DEL ,  4124 ,  4125 ,  4126 ,  4127 ,  4128 ,  4129 ,  4130 ,  4131 ,  4132 ,  4133 ,  4134 ,  4135 ,  4136 ,  4137 ,  4138 ,  4139 ,  4141 ,  4142 ,  4143 ,  4144 ,  4145 ,  4146 ,  4147 ,  4148 ,  4149 ,  4150 ,  4151 ,  4152 ,  4153 ,  4154 ,  4155 ,  4156 ,  4157 ,  4158 ,  4159 ,  4160 ,  4161 ,  4162 ,  4163 ,  4164 ,  4165 ,  4166 ,  4167 ,  4168 ,  4169 ,  4170 ,  4171 ,  4172 ,  4173 ,  4174 ,  4176 ,  4177 ,  4178 ,  4179 ,  4180 ,  4181 ,  4182 ,  4183 ,  4184 ,  4185 ,  4186 ,  4187 ,  4188 ,  4189 ,  4190 ,  4192 ,  4194 ,  4195 ,  4196 ,  4197 ,  4198 ,  4199 ,  4202 ,  4205 ,  4206 ,  4208 ,  4209 ,  4214 ,  4215 ,  4216 ,  4217 ,  4218 ,  4219 ,  4220 ,  4221 ,  4222 ,  4223 ,  4224 ,  4225 ,  4226 ,  5001 ,  5002 ,  5003 ,  5004 ,  5005 ,  5006 ,  5007 ,  5008 ,  5009 ,  5010 ,  5011 ,  5012 ,  5013 ,  5014 ,  5015 ,  5016 ,  5017 ,  5018 ,  5019 ,  5020 ,  5021 ,  5022 ,  5023 ,  5024 ,  5025 ,  5026 ,  5027 ,  5027A ,  5028 ,  5029 ,  5030 ,  5031 ,  5032 ,  5033 ,  5034 ,  5036 ,  5037 ,  5038 ,  5039 ,  5040 ,  5041 ,  5042 ,  5043 ,  5044 ,  5045 ,  5046 ,  5047 ,  5048 ,  5049 ,  5050 ,  5051 ,  5052 ,  5053 ,  5054 ,  5055 ,  5056 ,  5057 ,  5058 ,  5059 ,  5060 ,  5061 ,  5062 ,  5063 ,  5064 ,  5065 ,  5066 ,  5067 ,  5068 ,  5069 ,  5070 ,  5071 ,  5102 ,  5103 ,  5106 ,  5107 ,  5108 ,  5109 ,  5114 ,  5117 ,  5123 ,  5164 ,  6000 ,  6001 ,  6648 ,  6654 ,  7503 ;     *** 7) Part Number: 5977074; Serial Numbers: 1001 ,  4017 ,  4018 ,  4019 ,  4020 ,  4021 ,  4022 ,  4023 ,  4024 ,  4025 ,  4026 ,  4027 ,  4028 ,  4029 ,  4030 ,  4031 ,  4032 ,  4033 ,  4034 ,  4035 ,  4036 ,  4037 ,  4040 ,  4041 ,  4042 ,  4043 ,  4044 ,  4045 ,  4046 ,  4047 ,  4048 ,  4049 ,  4050 ,  4090 ,  4091 ,  5001 ,  5002 ,  5003 ,  5004 ,  5005 ,  5006 ,  5007 ,  5008 ,  5009 ,  5010 ,  5011 ,  5012 ,  5013 ,  5014 ,  5015 ,  5016 ,  5017 ,  5018 ,  5019 ,  5020 ,  5021 ,  5022 ,  5023 ,  5024 ,  5025 ,  5026 ,  5027 ,  5028 ,  5029 ,  5030 ,  5031 ,  5032 ,  5033 ,  5034 ,  5035 ,  5036 ,  5037 ,  5038 ,  5039 ,  5040 ,  5041 ,  5042 ,  5043 ,  701001 ,  7862 ;     *** 8) Part Number: 5984005; Serial Numbers: 1 ,  2 ,  3 ,  4 ,  6001 ,  6002 ,  6003 ,  6004 ,  6005 ,  6006 ,  6007 ,  6008 ,  6009 ,  6010 ,  7001 ,  7002 ,  7003 ,  7004 ,  7005 ,  7006 ,  7008 ,  7501 ,  7502 ,  7503 ,  7505 ,  7507 ,  7508 ,  7509 ,  8001 ;     *** 9) Part Number: 5989079; Serial Numbers: 08501/7823920 ,  1 ,  100100018 ,  4001 ,  4366618 ,  44001 ,  5224741 ,  5652 ,  6501 ,  7057 ,  71101 ,  71102 ,  72102 ,  7501 ,  7502 ,  7503 ,  7504 ,  7505 ,  7506 ,  7507 ,  7508 ,  7509 ,  7510 ,  7512 ,  7513 ,  7514 ,  7515 ,  7516 ,  7517 ,  7518 ,  7519 ,  7520 ,  7521 ,  7522 ,  7523 ,  7524 ,  7525 ,  7526 ,  7527 ,  7528 ,  7529 ,  7530 ,  7531 ,  7532 ,  7533 ,  7534 ,  7535 ,  7536 ,  7537 ,  7538 ,  7539 ,  7540 ,  7541 ,  7542 ,  7543 ,  7544 ,  7545 ,  7546 ,  7547 ,  7548 ,  7549 ,  7550 ,  7551 ,  7552 ,  7553 ,  7554 ,  7555 ,  7556 ,  7557 ,  7558 ,  7560 ,  7561 ,  7562 ,  7563 ,  7564 ,  7565 ,  7566 ,  7567 ,  7568 ,  7569 ,  7570 ,  7571 ,  7572 ,  7573 ,  7574 ,  7575 ,  7576 ,  7577 ,  7578 ,  7579 ,  7580 ,  7581 ,  7582 ,  7583 ,  7584 ,  7585 ,  7586 ,  7587 ,  7588 ,  7589 ,  7590 ,  7591 ,  7592 ,  7593 ,  7594 ,  7595 ,  7596 ,  7597 ,  7598 ,  7599 ,  7600 ,  7601 ,  7602 ,  7604 ,  7605 ,  7606 ,  7607 ,  7608 ,  7609 ,  7610 ,  7611 ,  7612 ,  7613 ,  7614 ,  7615 ,  7616 ,  7617 ,  7618 ,  7618* ,  7619 ,  7620 ,  7621 ,  7622 ,  7623 ,  7624 ,  7625 ,  7626 ,  7627 ,  7628 ,  7629 ,  7630 ,  7631 ,  7632 ,  7633 ,  7634 ,  7635 ,  7636 ,  7637 ,  7638 ,  7651 ,  7823920../08504 ,  7823920..08502 ,  7861 ,  7889 ,  8001 ,  8002 ,  8003 ,  8004 ,  8005 ,  8006 ,  8007 ,  8008 ,  8009 ,  8010 ,  8011 ,  8012 ,  8013 ,  8014 ,  8015 ,  8016 ,  8017 ,  8018 ,  8019 ,  8020 ,  8021 ,  8022 ,  8023 ,  8024 ,  8025 ,  8026 ,  8027 ,  8028 ,  8029 ,  8030 ,  8031 ,  8032 ,  8033 ,  8034 ,  8035 ,  8036 ,  8037 ,  8038 ,  8039 ,  8040 ,  8042 ,  8043 ,  8044 ,  8045 ,  8046 ,  8047 ,  8048 ,  8049 ,  8050 ,  8051 ,  8052 ,  8053 ,  8054 ,  8055 ,  8056 ,  8057 ,  8058 ,  8059 ,  8060 ,  8061 ,  8062 ,  8063 ,  8064 ,  8065 ,  8066 ,  8067 ,  8068 ,  8069 ,  8070 ,  8071 ,  8072 ,  8073 ,  8074 ,  8075 ,  8076 ,  8077 ,  8078 ,  8079 ,  8080 ,  8081 ,  8082 ,  8083 ,  8084 ,  8085 ,  8086 ,  8087 ,  8088 ,  8089 ,  8090 ,  8091 ,  8092 ,  8093 ,  8094 ,  8095 ,  8096 ,  8097 ,  8098 ,  8099 ,  8100 ,  8101 ,  8102 ,  8103 ,  8104 ,  8105 ,  8106 ,  8107 ,  8108 ,  8109 ,  8110 ,  8111 ,  8112 ,  8113 ,  8114 ,  8115 ,  8116 ,  8117 ,  8118 ,  8119 ,  8120 ,  8121 ,  8122 ,  8123 ,  8124 ,  8125 ,  8126 ,  8127 ,  8128 ,  8130 ,  8131 ,  8132 ,  8133 ,  8134 ,  8135 ,  8136 ,  8137 ,  8138 ,  8139 ,  8140 ,  8141 ,  8142 ,  8143 ,  8144 ,  8145 ,  8146 ,  8147 ,  8501 ,  8502 ,  8503 ,  8504 ,  8505 ,  8506 ,  8507 ,  8508 ,  8509 ,  8510 ,  8511 ,  8512 ,  8513 ,  8514 ,  8515 ,  8516 ,  8517 ,  8518 ,  8519 ,  8520 ,  8521 ,  8522 ,  8523 ,  8524 ,  8525 ,  8526 ,  8527 ,  8528 ,  8529 ,  8530 ,  8531 ,  8532 ,  8533 ,  8534 ,  8535 ,  8536 ,  8537 ,  8538 ,  8539 ,  8540 ,  8541 ,  8542 ,  8543 ,  8544 ,  8545 ,  8546 ,  8547 ,  8548 ,  8549 ,  8550 ,  8551 ,  8552 ,  8553 ,  8554 ,  8555 ,  8556 ,  8557 ,  8558 ,  8559 ,  8560 ,  8561 ,  8562 ,  8563 ,  8564 ,  8565 ,  8566 ,  8567 ,  8568 ,  8569 ,  8570 ,  8571 ,  8572 ,  8573 ,  8574 ,  8575 ,  8576 ,  8577 ,  8578 ,  8579 ,  8580 ,  8581 ,  8582 ,  8583 ,  8584 ,  8585 ,  8586 ,  8587 ,  8588 ,  8589 ,  8590 ,  8591 ,  8592 ,  8593 ,  8594 ,  8595 ,  8596 ,  8597 ,  8598 ,  8599 ,  8600 ,  8601 ,  8602 ,  8603 ,  8604 ,  8605 ,  8606 ,  8607 ,  8608 ,  8609 ,  8610 ,  8611 ,  8612 ,  8613 ,  8614 ,  8615 ,  8616 ,  8617 ,  8618 ,  8619 ,  8620 ,  8621 ,  8622 ,  8623 ,  8624 ,  8625 ,  8626 ,  8627 ,  8628 ,  8629 ,  8630 ,  8631 ,  8632 ,  8633 ,  8634 ,  8635 ,  8636 ,  8637 ,  8638 ,  8639 ,  8640 ,  8641 ,  8642 ,  8643 ,  8644 ,  8645 ,  8646 ,  8647 ,  8648 ,  8649 ,  8650 ,  8651 ,  8652 ,  8653 ,  8654 ,  8655 ,  8656 ,  8657 ,  8658 ,  8659 ,  8660 ,  8661 ,  8662 ,  8663 ,  8664 ,  8665 ,  8666 ,  8667 ,  8668 ,  8669 ,  8670 ,  8671 ,  8672 ,  8673 ,  8674 ,  8675 ,  8676 ,  8677 ,  8678 ,  8679 ,  8680 ,  8681 ,  8682 ,  8683 ,  8684 ,  8685 ,  8686 ,  8687 ,  8688 ,  8689 ,  8690 ,  8691 ,  8692 ,  8693 ,  8694 ,  8695 ,  8696 ,  8697 ,  8698 ,  8699 ,  8700 ,  8701 ,  8702 ,  8703 ,  8704 ,  8705 ,  8706 ,  8707 ,  8708 ,  8709 ,  8710 ,  8711 ,  8712 ,  8713 ,  8714 ,  8715 ,  8716 ,  8717 ,  8718 ,  8719 ,  8720 ,  8721 ,  8722 ,  8723 ,  8724 ,  8725 ,  8726 ,  8727 ,  8728 ,  8729 ,  8730 ,  8731 ,  8732 ,  8733 ,  8734 ,  8735 ,  8736 ,  8737 ,  8738 ,  8739 ,  8740 ,  8741 ,  8742 ,  8743 ,  8744 ,  8745 ,  8746 ,  8747 ,  8748 ,  8749 ,  8750 ,  8751 ,  8752 ,  8753 ,  8754 ,  8755 ,  8756 ,  8757 ,  8758 ,  8759 ,  8760 ,  8761 ,  8762 ,  8763 ,  8764 ,  8765 ,  8766 ,  8767 ,  8768 ,  8769 ,  8770 ,  8771 ,  8772 ,  8773 ,  8774 ,  8775 ,  8776 ,  8777 ,  8778 ,  8779 ,  8780 ,  8781 ,  8782 ,  8783 ,  8784 ,  8785 ,  8786 ,  8787 ,  8788 ,  8789 ,  8790 ,  8791 ,  8792 ,  8793 ,  8794 ,  8794DEL ,  8795 ,  8796 ,  8797 ,  8798 ,  8799 ,  8800 ,  8801 ,  8802 ,  8803 ,  8804 ,  8805 ,  8806 ,  8807 ,  8808 ,  8809 ,  8810 ,  8811 ,  8812 ,  8813 ,  8814 ,  8815 ,  8816 ,  8817 ,  8818 ,  8819 ,  8820 ,  8821 ,  8822 ,  8823 ,  8824 ,  8825 ,  8826 ,  8827 ,  8828 ,  8829 ,  8830 ,  8831 ,  8832 ,  8833 ,  8834 ,  8835 ,  8836 ,  8837 ,  8838 ,  8839 ,  8840 ,  8841 ,  8842 ,  8843 ,  8844 ,  8847 ;     *** 10) Part Number: 5989087; Serial Numbers: 5045 ,  7501 ,  7502 ,  7503 ,  7504 ,  7505 ,  7506 ,  7507 ,  7508 ,  7509 ,  7510 ,  7511 ,  7512 ,  8001 ,  8002 ,  8003 ,  8004 ,  8005 ,  8006 ,  8007 ,  8008 ,  8009 ,  8010 ,  8011 ,  8012 ,  8013 ,  8014 ,  8015 ,  8016 ,  8017 ,  8018 ,  8019 ,  8020 ,  8021 ,  8022 ,  8023 ,  8024 ,  8025 ,  8026 ,  8027 ,  8028 ,  8029 ,  8030 ,  8031 ,  8032 ,  8033 ,  8501 ,  8502 ,  8503 ,  8504 ,  8505 ,  8506 ,  8507 ,  8508 ,  8509 ,  8510 ,  8511 ,  8512 ,  8513 ,  8514 ,  8515 ,  8516 ,  8517 ,  8518 ,  8519 ,  8520 ,  8521 ,  8522 ,  8523 ,  8524 ,  8525 ,  8526 ,  8527 ,  8528 ,  8529 ,  8530 ,  8532 ,  8533 ,  8534 ,  8535 ,  8536 ,  8537 ,  8538 ,  8539 ,  8540 ,  8541 ,  8542 ,  8543 ,  8544 ,  8545 ,  8546 ,  8547 ,  8548 ,  8549 ,  8550 ,  8551 ,  8552 ,  8553 ,  8554 ,  8555 ,  8556 ;     *** 11) Part Number: 5989095; Serial Numbers: 7090 ,  7501 ,  7502 ,  7504 ,  7505 ,  8001 ,  8002 ,  8003 ,  8004 ,  8005 ,  8021 ,  8501 ,  8502 ,  8503 ,  8504 ,  8505 ,  8506 ,  8507 ;     *** 12) Part Number: 5991109; Serial Numbers: 05173/5992099.. ,  1001 ,  5039 ,  5161 ,  5501 ,  5503 ,  5504 ,  5505 ,  5506 ,  5507 ,  5508 ,  5509 ,  5510 ,  5511 ,  5512 ,  5513 ,  5514 ,  5515 ,  5516 ,  5517 ,  5518 ,  5519 ,  5520 ,  5521 ,  5522 ,  5523 ,  5524 ,  5525 ,  5526 ,  5527 ,  5528 ,  5529 ,  5530 ,  5531 ,  5532 ,  5533 ,  5534 ,  5535 ,  5536 ,  5537 ,  5538 ,  5539 ,  5540 ,  5541 ,  5542 ,  5543 ,  5544 ,  5545 ,  5546 ,  5547 ,  5548 ,  5549 ,  5550 ,  5551 ,  5552 ,  5553 ,  5554 ,  5555 ,  5556 ,  5557 ,  5558 ,  5559 ,  5560 ,  5561 ,  5562 ,  5563 ,  5564 ,  5565 ,  5566 ,  5567 ,  5568 ,  5569 ,  5570 ,  5571 ,  5572 ,  5573 ,  5575 ,  5576 ,  5577 ,  5578 ,  5579 ,  5580 ,  5581 ,  5582 ,  5583 ,  5584 ,  5585 ,  5586 ,  5587 ,  5588 ,  5589 ,  5590 ,  5591 ,  5592 ,  5593 ,  5594 ,  5595 ,  5596 ,  5597 ,  5598 ,  5599 ,  5600 ,  5601 ,  5602 ,  5603 ,  5604 ,  5605 ,  5606 ,  5607 ,  5608 ,  5609 ,  5610 ,  5611 ,  5612 ,  5613 ,  5614 ,  5615 ,  5616 ,  5617 ,  5618 ,  5619 ,  5620 ,  5621 ,  5622 ,  5623 ,  5624 ,  5625 ,  5626 ,  5627 ,  5628 ,  5629 ,  5631 ,  5632 ,  5633 ,  5634 ,  5635 ,  5636 ,  5637 ,  5638 ,  5639 ,  5640 ,  5641 ,  5642 ,  5643 ,  5644 ,  5645 ,  5646 ,  5647 ,  5649 ,  5650 ,  5652 ,  5655 ,  5658 ,  5671 ,  5672 ,  5673 ,  5674 ,  5675 ,  5676 ,  5677 ,  5678 ,  5679 ,  7001 ,  7002 ,  7003 ,  7004 ,  7005 ,  7006 ,  7007 ,  7008 ,  7009 ,  7010 ,  7011 ,  7012 ,  7013 ,  7014 ,  7016 ,  7017 ,  7018 ,  7019 ,  7020 ,  7021 ,  7022 ,  7023 ,  7024 ,  7025 ,  7026 ,  7027 ,  7028 ,  7029 ,  7030 ,  7031 ,  7032 ,  7033 ,  7034 ,  7035 ,  7036 ,  7037 ,  7038 ,  7039 ,  7040 ,  7041 ,  7042 ,  7043 ,  7044 ,  7045 ,  7046 ,  7047 ,  7048 ,  7049 ,  7050 ,  7051 ,  7052 ,  7053 ,  7054 ,  7055 ,  7056 ,  7057 ,  7058 ,  7059 ,  7060 ,  7061 ,  7062 ,  7063 ,  7065 ,  7067 ,  7068 ,  7069 ,  7070 ,  7071 ,  7072 ,  7073 ,  7074 ,  7075 ,  7076 ,  7077 ,  7078 ,  7079 ,  7080 ,  7081 ,  7082 ,  7083 ,  7084 ,  7085 ,  7086 ,  7088 ,  7090 ,  7091 ,  7092 ,  7093 ,  7094 ,  7095 ,  7096 ,  7097 ,  7097. ,  7098 ,  7099 ,  7101 ,  7102 ,  7103 ,  7104 ,  7105 ,  7109 ,  7110 ,  7111 ,  7112 ,  7114 ,  7115 ,  7116 ,  7117 ,  7118 ,  7119 ,  7120 ,  7121 ,  7122 ,  7123 ,  7124 ,  7125 ,  7126 ,  7127 ,  7129 ,  7130 ,  7178 ,  7179 ,  7501 ,  7502 ,  7503 ,  7504 ,  7505 ,  7506 ,  7507 ,  7508 ,  7509 ,  7510 ,  7511 ,  7512 ,  7513 ,  7514 ,  7515 ,  7516 ,  7517 ,  7518 ,  7519 ,  7520 ,  7521 ,  7522 ,  7523 ,  7524 ,  7525 ,  7526 ,  7527 ,  7528 ,  7529 ,  7530 ,  7531 ,  7532 ,  7533 ,  7534 ,  7535 ,  7536 ,  7537 ,  7538 ,  7539 ,  7540 ,  7541 ,  7542 ,  7543 ,  7544 ,  7545 ,  7546 ,  7547 ,  7549 ,  7550 ,  7551 ,  7552 ,  7553 ,  7554 ,  7555 ,  7556 ,  7557 ,  7558 ,  7559 ,  7560 ,  7561 ,  7562 ,  CAX271/U2B004 ;     *** 13) Part Number: 5991117; Serial Numbers: 4005 ,  5501 ,  5502 ,  5503 ,  5504 ,  5505 ,  5506 ,  5507 ,  5508 ,  5509 ,  5510 ,  5511 ,  5512 ,  5513 ,  5514 ,  5515 ,  5516 ,  5517 ,  5518 ,  5519 ,  5520 ,  5521 ,  5522 ,  5523 ,  5524 ,  5525 ,  5526 ,  5527 ,  5528 ,  5529 ,  5626 ,  7001 ,  7002 ,  7003 ,  7004 ,  7005 ,  7006 ,  7007 ,  7008 ,  7009 ,  7010 ,  7011 ,  7012 ,  7013 ,  7013-SMS ,  7014 ,  7015 ,  7016 ,  7017 ,  7018 ,  7019 ,  7020 ,  7021 ,  7022 ,  7023 ,  7024 ,  7025 ,  7026 ,  7501 ,  7502 ,  7503 ,  7504 ,  7505 ,  7505X ,  7506 ,  7507 ,  7508 ,  7509 ,  7510 ,  7511 ,  7512 ,  7513 ,  7514 ;     *** 14) Part Number: 5992099; Serial Numbers: 5101 ,  5104 ,  5107 ,  5109 ,  5110 ,  5111 ,  5112 ,  5113 ,  5115 ,  5116 ,  5118 ,  5119 ,  5120 ,  5121 ,  5122 ,  5123 ,  5124 ,  5125 ,  5126 ,  5127 ,  5128 ,  5129 ,  5131 ,  5132 ,  5133 ,  5134 ,  5135 ,  5136 ,  5137 ,  5138 ,  5139 ,  5140 ,  5141 ,  5142 ,  5143 ,  5144 ,  5145 ,  5146 ,  5147 ,  5148 ,  5149 ,  5150 ,  5151 ,  5152 ,  5153 ,  5154 ,  5155 ,  5156 ,  5157 ,  5158 ,  5159 ,  5160 ,  5161 ,  5162 ,  5163 ,  5165 ,  5166 ,  5167 ,  5168 ,  5169 ,  5170 ,  5171 ,  5172 ,  5173 ,  5174 ,  5211 ,  5502 ,  5578 ,  6510 ;     *** 15) Part Number: 7324119; Serial Numbers: 1003 ,  520 ,  542 ,  548 ,  550 ,  558 ,  564 ,  579 ,  581 ,  585 ,  588 ,  589 ,  590 ,  592 ,  593 ,  595 ,  596 ,  604 ,  9164 ,  RC494 ,  RC518 ,  RC528 ,  RC537 ,  RC539 ,  RC549 ;     *** 16) Part Number: 7324135; Serial Numbers: 551 ,  574 ,  584 ,  591 ,  RC521 ,  RC536 ;     *** 17) Part Number: 7324143; Serial Numbers: 510 ,  529 ,  530 ,  557 ,  559 ,  563 ,  565 ,  573 ,  575 ,  580 ,  582 ,  587 ,  594 ,  7127 ,  9325 ,  RC495 ,  RC498 ,  RC543 ;     *** 18) Part Number: 7324150; Serial Numbers: 556 ,  572 ,  RC492 ,  RC493 ,  RC524 ,  RC538 ,  RC540 ,  RC541 ,  RC545 ,  RC546 ;     *** 19) Part Number: 7760809; Serial Numbers: 8501 ,  8502 ,  8503 ,  8504 ,  8505 ,  8506 ,  8507 ,  8508 ,  8509 ,  8510 ,  8511 ;     *** 20) Part Number: 7760932; Serial Numbers: 8501 ,  8502 ,  8503 ,  8505 ,  8506 ,  8507 ,  8508 ,  8509 ,  8510 ,  8511 ,  8512 ,  8513 ,  8514 ,  8515 ,  8516 ,  8517 ,  8518 ,  8519 ,  8520 ,  8521 ,  8522 ,  8523 ,  8524 ,  8525 ,  8526 ,  8527 ,  8528 ,  8529 ,  8530 ,  8531 ,  8532 ,  8533 ,  8534 ,  8535 ,  8536 ,  8537 ,  8538 ,  8539 ,  8540 ,  8541 ,  8542 ,  8543 ,  8544 ,  8545 ,  8546 ,  8547 ,  8548 ,  8549 ,  8550 ,  8551 ,  8552 ,  8553 ,  8554 ,  8555 ,  8556 ,  8557 ,  8558 ,  8560 ,  8561 ,  8562 ,  8563 ,  8564 ,  8565 ,  8566 ,  8567 ,  8568 ,  8569 ,  8570 ,  8571 ,  8572 ,  8573 ,  8574 ,  8575 ,  8576 ,  8577 ,  8578 ,  8579 ,  8580 ,  8581 ,  8582 ,  8583 ,  9001 ,  9002 ,  9003 ,  9004 ,  9005 ,  9006 ,  9007 ,  9008 ,  9009 ,  9010 ,  9011 ,  9012 ,  9013 ,  9014 ,  9015 ,  9016 ,  9017 ,  9018 ,  9019 ,  9020 ,  9021 ,  9022 ,  9023 ,  9024 ,  9025 ,  9026 ,  9027 ,  9028 ,  9029 ,  9030 ,  9031 ,  9032 ,  9033 ,  9034 ,  9035 ,  9036 ,  9037 ,  9038 ,  9039 ,  9040 ,  9041 ,  9042 ,  9043 ,  9044 ;     *** 21) Part Number: 7761161; Serial Numbers: 8501 ,  8502 ,  8503 ;     *** 22) Part Number: 7823920; Serial Numbers: 3097 ,  7036 ,  71101 ,  8502 ,  8505 ,  8506 ,  8507 ,  8508 ,  8509 ,  8510 ,  8511 ,  8513 ,  8514 ,  8515 ,  8516 ,  8517 ,  8518 ,  8519 ,  8520 ,  8521 ,  8522 ,  9821 ;     *** 23) Part Number: 7823946; Serial Numbers: 10000 ,  10001 ,  10002 ,  10003 ,  10004 ,  10005 ,  10006 ,  10007 ,  10008 ,  10009 ,  10010 ,  10011 ,  10012 ,  10013 ,  10014 ,  10015 ,  10016 ,  10017 ,  10018 ,  10019 ,  10020 ,  10021 ,  10022 ,  10023 ,  10024 ,  10025 ,  10026 ,  10027 ,  10028 ,  10029 ,  10030 ,  10031 ,  10032 ,  10033 ,  10034 ,  10035 ,  10036 ,  10037 ,  10038 ,  10039 ,  10040 ,  10041 ,  10042 ,  10043 ,  10044 ,  10045 ,  10046 ,  10047 ,  10048 ,  10049 ,  10050 ,  10051 ,  10052 ,  10053 ,  10054 ,  10055 ,  10056 ,  10057 ,  10058 ,  10059 ,  10060 ,  10061 ,  10062 ,  10063 ,  10064 ,  10065 ,  10066 ,  10067 ,  10068 ,  10069 ,  10070 ,  10071 ,  10072 ,  10073 ,  10074 ,  10075 ,  10076 ,  10077 ,  10078 ,  10079 ,  10080 ,  10082 ,  10083 ,  10084 ,  10085 ,  10086 ,  10087 ,  10088 ,  10089 ,  10090 ,  10091 ,  10092 ,  10093 ,  10094 ,  10095 ,  10096 ,  10097 ,  10098 ,  10099 ,  10100 ,  10101 ,  10102 ,  10103 ,  10104 ,  10105 ,  10106 ,  10107 ,  10108 ,  10109 ,  10110 ,  10111 ,  10112 ,  10113 ,  10114 ,  10115 ,  10116 ,  10117 ,  10118 ,  10119 ,  10120 ,  10121 ,  10122 ,  10123 ,  10124 ,  10125 ,  10126 ,  10127 ,  10128 ,  10129 ,  1013 ,  10130 ,  10131 ,  10132 ,  10133 ,  10134 ,  10135 ,  10136 ,  10137 ,  10138 ,  10139 ,  10140 ,  10141 ,  10142 ,  10143 ,  10144 ,  10145 ,  10146 ,  10147 ,  10148 ,  10149 ,  10150 ,  10151 ,  10152 ,  10153 ,  10154 ,  10155 ,  10156 ,  10157 ,  10158 ,  10159 ,  10160 ,  10161 ,  10162 ,  10163 ,  10164 ,  10165 ,  10166 ,  10167 ,  10168 ,  10169 ,  10170 ,  10171 ,  10172 ,  10173 ,  10174 ,  10175 ,  10176 ,  10177 ,  10178 ,  10179 ,  10180 ,  10181 ,  10182 ,  10183 ,  10184 ,  10185 ,  10186 ,  10187 ,  10188 ,  10190 ,  10191 ,  10192 ,  10193 ,  10194 ,  10195 ,  10196 ,  10197 ,  10198 ,  10199 ,  10200 ,  10201 ,  10202 ,  10203 ,  10204 ,  10205 ,  10206 ,  10207 ,  10208 ,  10209 ,  10210 ,  10211 ,  10212 ,  10213 ,  10214 ,  10215 ,  10216 ,  10217 ,  10218 ,  10219 ,  10220 ,  10221 ,  10222 ,  10223 ,  10224 ,  10225 ,  10226 ,  10227 ,  10228 ,  10229 ,  10230 ,  10231 ,  10232 ,  10233 ,  10234 ,  10235 ,  10236 ,  10237 ,  10238 ,  10239 ,  10240 ,  10241 ,  10242 ,  10243 ,  10244 ,  10245 ,  10246 ,  10247 ,  10248 ,  10249 ,  10250 ,  10251 ,  10252 ,  10253 ,  10254 ,  10255 ,  10256 ,  10257 ,  10258 ,  10259 ,  10260 ,  10261 ,  10262 ,  10263 ,  10264 ,  10265 ,  10266 ,  10267 ,  10268 ,  10269 ,  10270 ,  10271 ,  10272 ,  10273 ,  10274 ,  10275 ,  10276 ,  10277 ,  10278 ,  10279 ,  10280 ,  10281 ,  10282 ,  10283 ,  10284 ,  10285 ,  10286 ,  10287 ,  10288 ,  10289 ,  10290 ,  10291 ,  10292 ,  10293 ,  10294 ,  10295 ,  10296 ,  10297 ,  10298 ,  10299 ,  10300 ,  10301 ,  10302 ,  10303 ,  10304 ,  10305 ,  10306 ,  10307 ,  10308 ,  10309 ,  10310 ,  10311 ,  10312 ,  10313 ,  10314 ,  10315 ,  10316 ,  10317 ,  10318 ,  10319 ,  10320 ,  10321 ,  10322 ,  10323 ,  10324 ,  10325 ,  10326 ,  10327 ,  10328 ,  10329 ,  10330 ,  10331 ,  10332 ,  10333 ,  10334 ,  10335 ,  10336 ,  10337 ,  10338 ,  10339 ,  10340 ,  10341 ,  10342 ,  10343 ,  10344 ,  10345 ,  10346 ,  10347 ,  10348 ,  10349 ,  10350 ,  10351 ,  10352 ,  10353 ,  10354 ,  10355 ,  10356 ,  10357 ,  10358 ,  10359 ,  10360 ,  10361 ,  10362 ,  11001 ,  11002 ,  11003 ,  11004 ,  11005 ,  11006 ,  11007 ,  11008 ,  11009 ,  11010 ,  11011 ,  11012 ,  11013 ,  11014 ,  11015 ,  11016 ,  11017 ,  11018 ,  11019 ,  11020 ,  11021 ,  11022 ,  11023 ,  11024 ,  11025 ,  11026 ,  11027 ,  11028 ,  11029 ,  11030 ,  11031 ,  11032 ,  11033 ,  11034 ,  11035 ,  11036 ,  11037 ,  11038 ,  11039 ,  11040 ,  11041 ,  11042 ,  11043 ,  11044 ,  11045 ,  11046 ,  11047 ,  11048 ,  11049 ,  11050 ,  11051 ,  11052 ,  11053 ,  11054 ,  11055 ,  11056 ,  11057 ,  11058 ,  11059 ,  11060 ,  11061 ,  11062 ,  11063 ,  11064 ,  11065 ,  11066 ,  11067 ,  11068 ,  11069 ,  11070 ,  11071 ,  11072 ,  11073 ,  11074 ,  11075 ,  11076 ,  11077 ,  11078 ,  11079 ,  11080 ,  11081 ,  11082 ,  11083 ,  11084 ,  11085 ,  11086 ,  11087 ,  11088 ,  11089 ,  11090 ,  11091 ,  11092 ,  11093 ,  11094 ,  11095 ,  11096 ,  11097 ,  11098 ,  11099 ,  11100 ,  11101 ,  11102 ,  11103 ,  11104 ,  11105 ,  11106 ,  11107 ,  11108 ,  11109 ,  11110 ,  11111 ,  11112 ,  11113 ,  11114 ,  11115 ,  11116 ,  11117 ,  11118 ,  11119 ,  11120 ,  11121 ,  11122 ,  11123 ,  11124 ,  11125 ,  11126 ,  11127 ,  11128 ,  11129 ,  11130 ,  11131 ,  11132 ,  11133 ,  11134 ,  11135 ,  11136 ,  11137 ,  11138 ,  11139 ,  11140 ,  11141 ,  11142 ,  11143 ,  11144 ,  11145 ,  11146 ,  11147 ,  11148 ,  11149 ,  11150 ,  11151 ,  11152 ,  11153 ,  11154 ,  11155 ,  11156 ,  11157 ,  11158 ,  11159 ,  11160 ,  11161 ,  11162 ,  11163 ,  11164 ,  11165 ,  11166 ,  11167 ,  11168 ,  11169 ,  11171 ,  11172 ,  11173 ,  11174 ,  11175 ,  11176 ,  11177 ,  11178 ,  11179 ,  11180 ,  11181 ,  11182 ,  11183 ,  11184 ,  11185 ,  11186 ,  11187 ,  11188 ,  11189 ,  11190 ,  11191 ,  11192 ,  11193 ,  11194 ,  11195 ,  11196 ,  11197 ,  11198 ,  11199 ,  11200 ,  11201 ,  11202 ,  11203 ,  11204 ,  11205 ,  11206 ,  11207 ,  11208 ,  11209 ,  11210 ,  11211 ,  11212 ,  11213 ,  11214 ,  11215 ,  11216 ,  11217 ,  11218 ,  11219 ,  11220 ,  11221 ,  11222 ,  11223 ,  11224 ,  11225 ,  11226 ,  11227 ,  11228 ,  11229 ,  11230 ,  11231 ,  11232 ,  11233 ,  11234 ,  11235 ,  11236 ,  11237 ,  11238 ,  11239 ,  11240 ,  11241 ,  11242 ,  11243 ,  11244 ,  11245 ,  11246 ,  11247 ,  11248 ,  11249 ,  11250 ,  11251 ,  11252 ,  11253 ,  11254 ,  11255 ,  11256 ,  11257 ,  11258 ,  11259 ,  11260 ,  11261 ,  11262 ,  11263 ,  11264 ,  11265 ,  11266 ,  11267 ,  11268 ,  11269 ,  11270 ,  11271 ,  11272 ,  11273 ,  11274 ,  11275 ,  11276 ,  11277 ,  11278 ,  11279 ,  11280 ,  11281 ,  11282 ,  11283 ,  11284 ,  11285 ,  11286 ,  11287 ,  11288 ,  11289 ,  11290 ,  11291 ,  11292 ,  11293 ,  11294 ,  11295 ,  11296 ,  11297 ,  11298 ,  11299 ,  11300 ,  11301 ,  11302 ,  11303 ,  11304 ,  11305 ,  11306 ,  11307 ,  11308 ,  11309 ,  11310 ,  11311 ,  11312 ,  11313 ,  11314 ,  11315 ,  11316 ,  11317 ,  11318 ,  11319 ,  11320 ,  11321 ,  11322 ,  11323 ,  11324 ,  11325 ,  11326 ,  11327 ,  11328 ,  11329 ,  11330 ,  11331 ,  11332 ,  11333 ,  11334 ,  11335 ,  11336 ,  11337 ,  11338 ,  11339 ,  11340 ,  11341 ,  11342 ,  11343 ,  11344 ,  11345 ,  11346 ,  11347 ,  11348 ,  11349 ,  11350 ,  11351 ,  11352 ,  11353 ,  11354 ,  11355 ,  11356 ,  11357 ,  11358 ,  11359 ,  11360 ,  11361 ,  11362 ,  11363 ,  11365 ,  11366 ,  11367 ,  11368 ,  11369 ,  11370 ,  11371 ,  11372 ,  11373 ,  11374 ,  11375 ,  11376 ,  11377 ,  11378 ,  11379 ,  11380 ,  11381 ,  11382 ,  11383",
      "center_classification_date": "20141022",
      "distribution_pattern": "Worldwide Distribution -- USA, including the states of AK, AL, AR, AZ, CA, CO, CT,  DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY; and, the District of Columbia, Guam, and Puerto Rico; and, the countries of Albania, Algeria, Angola, Argentina, Austria, Australia, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Bulgaria, Cameroon, Canada, Chile, China, Columbia, Congo, Costa Rica, Cuba, Croatia, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Georgia, Germany, Greece, Honduras, Hong Kong, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kuwait, Latvia, Lebanon, Luxembourg, Macedonia, Madagascar, Malaysia, Mexico, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Niger, Nigeria, Norway, Oman, Pakistan, Palestine, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Sri Lanka, Sweden, Switzerland, Syria, Taiwan, Thailand, Trinidad and Tobago, Tunisia, Turkey, Uganda, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Vietnam, and Yemen.",
      "state": "IL",
      "product_description": "e.cam line of gamma camera systems     Used to detect or image the distribution of radionuclides in the body or organ.",
      "report_date": "20141029",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Siemens Medical Solutions USA, Inc.",
      "recall_number": "Z-0116-2015",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "66778",
      "termination_date": "20161223",
      "more_code_info": ",  11384 ,  11385 ,  11386 ,  11387 ,  11388 ,  11389 ,  11390 ,  11391 ,  11392 ,  11393 ,  11394 ,  11395 ,  11396 ,  11397 ,  11398 ,  11399 ,  11400 ,  11401 ,  11402 ,  11403 ,  11404 ,  11405 ,  11406 ,  11407 ,  11408 ,  11409 ,  11410 ,  11411 ,  11412 ,  11413 ,  11414 ,  11415 ,  11417 ,  11418 ,  11419 ,  11420 ,  11421 ,  11422 ,  11423 ,  11424 ,  11425 ,  11426 ,  11427 ,  11428 ,  11429 ,  11430 ,  11431 ,  11432 ,  11433 ,  11434 ,  11435 ,  11436 ,  11437 ,  11438 ,  11439 ,  11440 ,  11441 ,  11442 ,  11443 ,  11444 ,  11445 ,  11446 ,  11447 ,  11448 ,  11449 ,  11450 ,  11451 ,  11452 ,  11453 ,  11454 ,  11455 ,  11456 ,  11457 ,  11458 ,  11459 ,  11460 ,  11461 ,  11462 ,  11463 ,  11464 ,  11465 ,  11466 ,  11467 ,  11468 ,  11469 ,  11470 ,  11471 ,  11472 ,  11473 ,  11474 ,  11475 ,  11476 ,  11477 ,  11478 ,  11479 ,  11480 ,  11481 ,  11482 ,  11483 ,  11484 ,  11485 ,  11486 ,  11487 ,  11488 ,  11489 ,  11490 ,  11492 ,  11493 ,  11494 ,  11495 ,  11496 ,  11497 ,  11498 ,  11499 ,  11500 ,  11501 ,  11502 ,  11503 ,  11504 ,  11505 ,  11506 ,  11507 ,  11508 ,  11509 ,  11510 ,  11511 ,  11512 ,  11513 ,  11514 ,  11515 ,  11516 ,  11517 ,  11518 ,  11519 ,  11520 ,  11521 ,  11522 ,  11523 ,  11524 ,  11525 ,  11526 ,  11527 ,  11528 ,  11529 ,  11530 ,  11531 ,  11532 ,  11533 ,  11534 ,  11535 ,  11536 ,  11537 ,  11538 ,  11539 ,  11540 ,  11541 ,  11542 ,  11543 ,  11544 ,  11545 ,  11546 ,  11547 ,  11548 ,  11549 ,  11550 ,  11551 ,  11552 ,  11553 ,  11554 ,  11555 ,  11556 ,  11557 ,  11558 ,  11559 ,  11560 ,  11562 ,  11563 ,  11564 ,  11565 ,  11566 ,  11567 ,  11568 ,  11569 ,  11570 ,  11571 ,  11572 ,  11573 ,  11574 ,  11575 ,  11576 ,  11577 ,  11578 ,  11579 ,  11580 ,  11581 ,  11582 ,  11583 ,  11584 ,  11585 ,  11586 ,  11587 ,  11588 ,  11589 ,  11590 ,  11591 ,  11592 ,  11593 ,  115930 ,  11594 ,  11595 ,  11596 ,  11597 ,  11598 ,  11599 ,  11600 ,  11601 ,  11602 ,  11603 ,  11604 ,  11605 ,  11606 ,  11607 ,  11608 ,  11609 ,  11610 ,  11611 ,  11612 ,  11613 ,  11614 ,  11615 ,  11616 ,  11617 ,  11618 ,  11619 ,  11620 ,  11621 ,  11622 ,  11623 ,  11624 ,  11625 ,  11626 ,  11627 ,  11628 ,  11629 ,  11630 ,  11631 ,  11632 ,  11633 ,  11634 ,  11635 ,  11636 ,  11637 ,  11638 ,  11639 ,  11640 ,  11641 ,  11642 ,  11643 ,  11644 ,  11645 ,  11646 ,  11647 ,  11648 ,  11649 ,  11650 ,  11651 ,  11652 ,  11653 ,  11654 ,  11655 ,  11656 ,  11657 ,  11658 ,  11659 ,  11660 ,  11661 ,  11662 ,  11663 ,  11664 ,  11665 ,  11666 ,  11667 ,  11668 ,  11669 ,  11670 ,  11671 ,  11672 ,  11673 ,  11674 ,  11675 ,  11676 ,  11677 ,  11678 ,  11679 ,  11680 ,  11681 ,  11682 ,  11683 ,  11684 ,  11685 ,  11686 ,  11687 ,  11688 ,  11689 ,  11690 ,  11691 ,  11692 ,  11693 ,  11694 ,  11695 ,  11696 ,  11697 ,  11698 ,  11699 ,  11700 ,  11701 ,  11702 ,  11703 ,  11704 ,  11705 ,  11706 ,  11707 ,  11708 ,  11709 ,  11710 ,  11711 ,  11712 ,  11713 ,  11714 ,  11715 ,  11716 ,  11717 ,  11718 ,  11719 ,  11720 ,  11721 ,  11722 ,  11723 ,  11724 ,  11725 ,  11726 ,  11727 ,  11728 ,  11729 ,  11730 ,  11731 ,  11732 ,  11733 ,  11734 ,  11735 ,  11736 ,  11737 ,  11738 ,  11739 ,  11740 ,  11741 ,  11742 ,  11743 ,  11744 ,  11745 ,  11746 ,  11747 ,  11748 ,  11749 ,  11750 ,  11751 ,  11752 ,  11753 ,  11754 ,  11755 ,  11756 ,  11757 ,  11758 ,  11759 ,  11760 ,  11761 ,  11762 ,  11763 ,  11764 ,  11765 ,  11766 ,  11767 ,  11768 ,  11769 ,  11770 ,  11771 ,  11772 ,  11773 ,  11774 ,  11775 ,  11776 ,  11777 ,  11778 ,  11779 ,  11780 ,  11781 ,  11782 ,  11783 ,  11784 ,  11785 ,  11786 ,  11787 ,  11788 ,  11789 ,  11790 ,  11791 ,  11792 ,  11793 ,  11794 ,  12345 ,  1278 ,  5382 ,  5389 ,  5517 ,  5764 ,  6481 ,  7778 ,  7853 ,  7883 ,  7990 ,  8484 ,  8490 ,  9001 ,  9002 ,  9003 ,  9004 ,  9005 ,  9006 ,  9007 ,  9008 ,  9009 ,  9010 ,  9011 ,  9012 ,  9013 ,  9014 ,  9015 ,  9016 ,  9017 ,  9018 ,  9019 ,  9020 ,  9021 ,  9022 ,  9023 ,  9024 ,  9025 ,  9026 ,  9027 ,  9028 ,  9029 ,  9030 ,  9031 ,  9032 ,  9033 ,  9034 ,  9035 ,  9036 ,  9037 ,  9038 ,  9039 ,  9040 ,  9041 ,  9042 ,  9043 ,  9044 ,  9045 ,  9046 ,  9047 ,  9048 ,  9049 ,  9050 ,  9051 ,  9052 ,  9053 ,  9054 ,  9055 ,  9056 ,  9057 ,  9058 ,  9059 ,  9060 ,  9061 ,  9062 ,  9063 ,  9064 ,  9065 ,  9066 ,  9067 ,  9068 ,  9069 ,  9070 ,  9071 ,  9072 ,  9073 ,  9074 ,  9075 ,  9076 ,  9077 ,  9078 ,  9079 ,  9080 ,  9081 ,  9082 ,  9083 ,  9084 ,  9085 ,  9086 ,  9087 ,  9088 ,  9089 ,  9090 ,  9091 ,  9092 ,  9093 ,  9094 ,  9095 ,  9096 ,  9097 ,  9098 ,  9099 ,  9100 ,  9101 ,  9102 ,  9103 ,  9104 ,  9105 ,  9106 ,  9107 ,  9108 ,  9109 ,  9110 ,  9111 ,  9112 ,  9113 ,  9114 ,  9115 ,  9116 ,  9117 ,  9118 ,  9119 ,  9120 ,  9121 ,  9122 ,  9123 ,  9124 ,  9125 ,  9126 ,  9127 ,  9128 ,  9129 ,  9130 ,  9131 ,  9132 ,  9133 ,  9134 ,  9135 ,  9136 ,  9137 ,  9138 ,  9139 ,  9140 ,  9141 ,  9142 ,  9143 ,  9144 ,  9145 ,  9146 ,  9147 ,  9148 ,  9149 ,  9150 ,  9151 ,  9152 ,  9153 ,  9154 ,  9155 ,  9156 ,  9157 ,  9158 ,  9159 ,  9160 ,  9161 ,  9162 ,  9163 ,  9164 ,  9165 ,  9166 ,  9167 ,  9168 ,  9169 ,  9170 ,  9171 ,  9172 ,  9173 ,  9174 ,  9175 ,  9176 ,  9177 ,  9178 ,  9179 ,  9180 ,  9181 ,  9182 ,  9183 ,  9184 ,  9185 ,  9186 ,  9187 ,  9188 ,  9189 ,  9190 ,  9191 ,  9192 ,  9193 ,  9194 ,  9195 ,  9196 ,  9197 ,  9198 ,  9199 ,  9200 ,  9201 ,  9202 ,  9203 ,  9204 ,  9205 ,  9206 ,  9207 ,  9208 ,  9209 ,  9210 ,  9211 ,  9212 ,  9213 ,  9214 ,  9215 ,  9216 ,  9217 ,  9218 ,  9219 ,  9220 ,  9221 ,  9222 ,  9223 ,  9224 ,  9225 ,  9226 ,  9227 ,  9228 ,  9229 ,  9230 ,  9231 ,  9232 ,  9233 ,  9234 ,  9235 ,  9236 ,  9237 ,  9238 ,  9239 ,  9240 ,  9241 ,  9242 ,  9243 ,  9244 ,  9245 ,  9246 ,  9247 ,  9248 ,  9249 ,  9250 ,  9251 ,  9252 ,  9253 ,  9254 ,  9255 ,  9256 ,  9257 ,  9259 ,  9260 ,  9261 ,  9262 ,  9263 ,  9264 ,  9265 ,  9266 ,  9267 ,  9268 ,  9269 ,  9270 ,  9271 ,  9272 ,  9273 ,  9274 ,  9275 ,  9276 ,  9277 ,  9278 ,  9279 ,  9280 ,  9281 ,  9282 ,  9283 ,  9284 ,  9285 ,  9287 ,  9288 ,  9289 ,  9290 ,  9291 ,  9292 ,  9293 ,  9294 ,  9295 ,  9296 ,  9297 ,  9298 ,  9299 ,  9300 ,  9301 ,  9302 ,  9303 ,  9304 ,  9305 ,  9306 ,  9307 ,  9308 ,  9309 ,  9310 ,  9311 ,  9313 ,  9724 ,  ECGS1 ;     *** 24) Part Number: 7823953; Serial Numbers: 10000 ,  10001 ,  10002 ,  10003 ,  10004 ,  10005 ,  10006 ,  10007 ,  10008 ,  10009 ,  10010 ,  10011 ,  10012 ,  10013 ,  10014 ,  10015 ,  10016 ,  10017 ,  10018 ,  10019 ,  10020 ,  10021 ,  10022 ,  10023 ,  10024 ,  10025 ,  10026 ,  10027 ,  10028 ,  10029 ,  10030 ,  10031 ,  10032 ,  10033 ,  10034 ,  10035 ,  10036 ,  10037 ,  10038 ,  10039 ,  10040 ,  10041 ,  10042 ,  10043 ,  10044 ,  10045 ,  10046 ,  10047 ,  10048 ,  10049 ,  10050 ,  10051 ,  10052 ,  10053 ,  10054 ,  10055 ,  10056 ,  10057 ,  10058 ,  10059 ,  10060 ,  10061 ,  10062 ,  10063 ,  10064 ,  11001 ,  11002 ,  11003 ,  11004 ,  11005 ,  11006 ,  11007 ,  11008 ,  11009 ,  11010 ,  11011 ,  11012 ,  11013 ,  11014 ,  11015 ,  11016 ,  11017 ,  11018 ,  11019 ,  11020 ,  11021 ,  11022 ,  11023 ,  11024 ,  11025 ,  11026 ,  11027 ,  11028 ,  11029 ,  11030 ,  11031 ,  11032 ,  11033 ,  11034 ,  11035 ,  11036 ,  11037 ,  11038 ,  11039 ,  11040 ,  11041 ,  11042 ,  11043 ,  11044 ,  11045 ,  11046 ,  11047 ,  11048 ,  11049 ,  11050 ,  11051 ,  11052 ,  11053 ,  11054 ,  11055 ,  11056 ,  11057 ,  11058 ,  11059 ,  11060 ,  11061 ,  11063 ,  11064 ,  11065 ,  11066 ,  11067 ,  11068 ,  11069 ,  11070 ,  11071 ,  11072 ,  11073 ,  11074 ,  11075 ,  11076 ,  11077 ,  11078 ,  11079 ,  11080 ,  11081 ,  11082 ,  11083 ,  11084 ,  11085 ,  11086 ,  11087 ,  11088 ,  11089 ,  11090 ,  11091 ,  11092 ,  11093 ,  11094 ,  11095 ,  11096 ,  11097 ,  11098 ,  11099 ,  11100 ,  11101 ,  11102 ,  11103 ,  11104 ,  11105 ,  11106 ,  11107 ,  11108 ,  11109 ,  11110 ,  11111 ,  11112 ,  11113 ,  11114 ,  11115 ,  11116 ,  11117 ,  11118 ,  11119 ,  11120 ,  11121 ,  11122 ,  11123 ,  11124 ,  11125 ,  11126 ,  11128 ,  11129 ,  11130 ,  11132 ,  11133 ,  11134 ,  11136 ,  11137 ,  11138 ,  11139 ,  11140 ,  11141 ,  11142 ,  11143 ,  11144 ,  11145 ,  11146 ,  4626 ,  6116 ,  8501 ,  8502 ,  8503 ,  8504 ,  8505 ,  8506 ,  8507 ,  8509 ,  9001 ,  9002 ,  9003 ,  9004 ,  9005 ,  9006 ,  9007 ,  9008 ,  9009 ,  9010 ,  9011 ,  9012 ,  9013 ,  9014 ,  9015 ,  9016 ,  9017 ,  9018 ,  9019 ,  9020 ,  9021 ,  9022 ,  9023 ,  9024 ,  9025 ,  9026 ,  9027 ,  9028 ,  9029 ,  9030 ,  9031 ,  9032 ,  9033 ,  9034 ,  9035 ,  9036 ,  9037 ,  9038 ,  9039 ,  9040 ,  9041 ,  9042 ,  9043 ,  9044 ,  9045 ,  9658 ,  T1039 ;     *** 25) Part Number: 7823979; Serial Numbers: \"Serial   Number\"   6324 ,  6683 ,  9001 ,  9002 ,  9004 ,  9005 ,  9006 ,  9007 ,  9008 ,  9009 ,  9010 ,  9011 ,  9012 ,  9013 ,  9014 ,  9015 ,  9016 ,  9017 ,  9018 ,  9019 ,  9020 ,  9021 ,  9022 ,  9023 ,  9025 ,  9026 ,  9028 ,  9029 ,  9031 ,  9304",
      "recall_initiation_date": "20131028",
      "postal_code": "60195-2061",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}