  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Santa Clara",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "90840",
      "recalling_firm": "Abbott Vascular",
      "address_1": "3200 Lakeside Dr",
      "address_2": "N/A",
      "postal_code": "95054-2807",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide including in the states of PR, WA, NE, MO, IA, KS, TN, MN, ND, MD, VA, MA, PA, DC, LA, KY, CA, NV, IL, WI, TX, CT, VT, AL, AR, GA, MS, MI, OH, FL, NC, NJ, CO, NY, DE, OR, SC, IN, HI, AZ, NH, ME, WV, OK, ID, UT, MT, NM, RI, AK, SD and the countries of AE, AR, AT, AU, BE, BG, BR, BY, CA, CH, CL, CN, CO, CR, CY, CZ, DE, DK, EC, EE, EG, ES, FI, FR, GB, GR, HK, HU, ID, IE, IL, IN, IR, IT, JO, JP, KR, KW, LB, LU, MT, MX, MY, NL, NO, NZ, OM, PE, PH, PK, PL, PT, QA, RE, RO, RU, SA, SE, SG, SI, SK, TH, TN, TR, TW.",
      "recall_number": "Z-0115-2023",
      "product_description": "Product/Part: MitraClip XTR Clip Delivery System US/CDS0601-XTR,  MitraClip G4 Clip Delivery System NT US/CDS0701-NT,  MitraClip G4 Clip Delivery System NTW US/CDS0701-NTW,  MitraClip G4 Clip Delivery System XT US/CDS0701-XT,  MitraClip G4 Clip Delivery System XTW US/CDS0701-XTW.  For cardiac procedures.",
      "product_quantity": "18,042",
      "reason_for_recall": "An increase in clip locking malfunctions, Establish Final Arm Angle and Clip Opening While Locked, have been observed. Contributing causes may be clip locking component material properties and certain use conditions. Unsuccessful clip procedure can lead to recurrence of severe mitral regurgitation, device embolization, or tissue injury.",
      "recall_initiation_date": "20220908",
      "center_classification_date": "20221020",
      "report_date": "20221026",
      "code_info": "Part/UDI: CDS0601-XTR/08717648226366, CDS0701-NT/08717648230943, CDS0701-NTW/08717648230967, CDS0701-XT/08717648230981, CDS0701-XTW/08717648231001,  All lots.",
      "more_code_info": ""
    }
  ]
}