{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Fremont",
      "state": "CA",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86583",
      "recalling_firm": "Stryker Neurovascular",
      "address_1": "47900 Bayside Pkwy",
      "address_2": "N/A",
      "postal_code": "94538-6515",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "N/A",
      "distribution_pattern": "U.S.: PA, OH, AK, NY, VA, IL, CA, NJ, AL, NV, FL, KY, AR, MA, TX, MI, NC, TN, WI, WA, AZ, OR, GA, VT, UT, DC, MO, IN, OK, NH, LA, HI, DE, SC, MS, RI, MD, SD, KS, ID, CO, CT, IA, NM, WV, MN, GU, NE.    Foreign: New Zealand, Argentina, Estonia, Lithuania, Slovenia, Australia, Finland, Luxembourg, South Africa, Austria, France, Mexico, South Korea, Belgium, Germany, Netherlands, Spain, Brazil, Greece, Norway, Sweden, Canada, Hong Kong, Philippines, Switzerland, Chile, Hungary, Poland, Taiwan, Colombia, India, Portugal, Thailand, Croatia, Ireland, Romania, Turkey, Czech Republic, Israel, Saudi Arabia, United Arab Emirates, Denmark, Italy, Singapore, United Kingdom, Egypt, Japan, Slovakia, Vietnam",
      "recall_number": "Z-0114-2021",
      "product_description": "Trevo XP ProVue Retriever, REF (Size):  90182 (4MM x 20MM); 90185 (4MM x 30MM); 90186 (6MM x 25MM). Procedure Pack REF: 80052, 93067, 93068, M0032PK42010001, M0033PK42022001, M0033PK43022001, M0033PK62523001, M0032PK42010002, M0032PK43010002, M0032PK62510002, M0033PK42022002, M0033PK43022002, M0033PK62523002",
      "product_quantity": "5365",
      "reason_for_recall": "Increase in fracture complaints of the flexible, tapered core wire, resulting in stent retriever separation from the core wire during use. The stent retriever can be left behind in the vasculature.",
      "recall_initiation_date": "20200921",
      "center_classification_date": "20201020",
      "termination_date": "20240401",
      "report_date": "20201028",
      "code_info": "REF/Lots: 90182/ 0000030024, 0000030025, 0000030026, 0000030251, 0000030252, 0000030253, 0000030408, 0000030409, 0000030410, 0000030921, 0000030922, 0000030931, 0000031336, 0000031585, 0000031761, 0000031762, 0000031763, 0000032042, 0000032043, 0000032309, 0000032310, 0000032400, 0000032401, 0000032402, 0000032403, 30024, 30251, 30922, 31336, 31761, 31762;  90185/ 0000030030, 0000030031, 0000030260, 0000030261, 0000030407, 0000030415, 0000030416, 0000030924, 0000030925, 0000030926, 0000031338, 0000031588, 0000031764, 0000031765, 0000032040, 0000032041, 0000032313;  90186/ 0000030032, 0000030262, 0000030406, 0000030417, 0000030418, 0000030419, 0000030927, 0000030928, 0000030929, 0000030930, 0000031339, 0000031340, 0000031341, 0000031587, 0000032046, 0000032047;  M0032PK42010002/ 38166585, 38191261, 38191264, 38449970, 38636576;  M0032PK43010002/ 36967533, 36967572, 36967573, 36967574, 36967575, 36967576, 36967577, 36967578, 36967579, 36967580;    93067/ 60200184892, 60200184893, 60200185332, 60200186591, 60200186592, 60200187741, 60200187742, 36836835, 36836838, 36891540, 36912340, 36912354, 36967536, 36984339, 36984362, 36984363, 36984364, 36984365, 36984366, 37003135, 37003141, 37003144, 37049334, 37049335, 37049336, 37049345, 37056132, 37077332, 37089641, 37108537, 37108540, 37125434, 37125468, 37158939, 37158940, 37174352, 37195439, 37195440, 37195441, 37195442, 37195444, 37195448, 37228358, 37228360, 37329459, 37495860, 37510394, 36448781, 36461581, 36461583, 36498382, 36498384, 36498680, 36498681, 36498682, 36529509, 36529545, 36529546, 36529547, 36529548, 36529549, 36529550, 36529551, 36529552, 36529553, 36529554, 36529555, 36529556, 36547189, 36547195, 36547197, 36547215, 36632788, 36632794, 36652186, 36691283, 36714887, 36735085, 36735086, 36735087, 36735088, 36735089, 36753486, 36753488, 36753497, 36799849, 36799850, 36799851, 36799859, 36799860, 36836842, 36836843, 36857851, 36857852, 36857853, 36857854, 36857855, 36857856, 36875438, 36875442, 36875443, 36875444, 36875453, 36912333, 36912353, 36930560, 36930561, 36930565, 36930569, 36930579, 36947233, 36947255, 36947256, 36947257, 36947258, 36967551, 36984352, 36984354, 36984355, 37003140, 37011643, 37032438, 37049357, 37108533, 37108544, 37108546, 37108547, 37108549, 37108550, 37108551, 37108565, 37108566, 37108567, 37108568, 37108569, 37108570, 37108571, 37108572, 37108573, 37125435, 37125448, 37125449, 37125456, 37125457, 37125458, 37141942, 37158947;    93068/ 60200189251, 60200189252, 60200189253, 60200190221, 60200190222, 60200190223, 60200190224, 60200191041, 60200191042, 60200191043, 60200191044, 37032432, 37032434, 37032435, 37077333, 37108538, 37141980, 37195438, 37212242, 37228359, 37263852, 37281945, 37313953, 37345011, 37345012, 37345013, 37384855, 37424965, 37479670, 37479734, 37510391, 37510392, 37539290, 37539291, 37539301, 37610409, 37659707, 37676912, 37693604, 37693635, 37713603, 37713604, 37754601, 37795612, 37814519, 37814520, 37814521, 37814522, 37849537, 37849538, 37849539, 37863831, 37969756, 37969757, 37969758, 37984459, 37984460, 37984468, 38013747, 38013948, 38013951, 38013952, 38013962, 38028310, 38045557, 38045558, 38045566, 38045567, 38045570, 37049353, 37049354, 37065739, 37077336, 37108545, 37108548, 37125460, 37125461, 37125467, 37158943, 37195443, 37212245, 37228338, 37228344, 37228345, 37228346, 37228357, 37228361, 37244247, 37244252, 37244253, 37263862, 37263875, 37297752, 37313955, 37329458, 37329460, 37329464, 37329465, 37344990, 37344996, 37364784, 37364785, 37364786, 37364787, 37384864, 37402863, 37402864, 37479669, 37479711, 37479714, 37479715, 37479752, 37480074, 37480075, 37480076, 37480078, 37480081, 37480084, 37480086, 37480089, 37480090, 37480091, 37480092, 37480093, 37480094, 37480095, 37495864, 37495867, 37495868, 37543891, 37543895, 37575302, 37575304, 37575315, 37575317, 37610402, 37628108, 37628109, 37644715, 37676901, 37676904, 37676905, 37676909, 37693620, 37693621, 37693622, 37693623, 37693624, 37693625, 37693630, 37693634, 37713620, 37733704, 37733712, 37733716, 37754602, 37754607, 37754608, 37754615, 37754616, 37754617, 37754618, 37754621, 37777326, 37777327, 37777335, 37795619, 37795620, 37795621, 37795622, 37814544, 37814545, 37833331, 37833332, 37849512, 37849513, 37849517, 37849518, 37849520, 37849521, 37849525, 37849526, 37849527, 37849543, 37849552, 37850811, 37850812, 37863830, 37863838, 37863842, 37863850, 37882547, 37882548, 37882554, 37919235, 37923331, 37954149, 37969749, 37969775, 37969776, 37969777, 37969778, 37984471, 38000241;    M0032PK62510002/ 36967569, 37011639, 37882543, 37938251;  M0033PK42022002/ 38146882, 38146883, 38166603, 38166604, 38174955, 38174975, 38185258, 38185259, 38185260, 38185263, 38185267, 38195456, 38281762, 38391766, 38391767, 38391769, 38391770, 38391771, 38391772, 38392064, 38392960, 38395862, 38476062, 38476065, 38476066, 38476067, 38516767, 38599024, 38166612, 38229563, 38261557, 38311966, 38392961, 38395871, 38408188, 38453760, 38453765, 38492462, 38572772, 38586764, 38636571, 38649867;  M0033PK43022002/ 36431078, 36431081, 36431082, 36448779, 36448780, 36498687, 36498688, 36498689, 36498690, 36498691, 36514194, 36529540, 36529541, 36547182, 36547187, 36547204, 36614192, 36632790, 36632797, 36672623, 36714883, 36714884, 36714886, 36735105, 36735106, 36753494, 36753495, 36753496, 36769792, 36769793, 36787186, 36787189, 36799835, 36799840, 36836844, 36836845, 36875448, 36875449, 36875450, 36875451, 36912346, 36912351, 36912352, 36984348, 36984349, 36984350, 36984351, 36984353, 37011640, 37032440, 37049350, 37049355, 37077337, 37077339, 37125462, 37125463, 37125464, 37158944, 37158946, 37174350, 36672623, 36984337, 37017833, 37174338, 37174353, 37174355, 37228363;    M0033PK62523002/ 37032443, 37032444, 37077341, 37228342, 37228343, 37281954, 37329456, 37329457, 37364754, 37364788, 37610411, 37777339, 37882544, 37938250, 38028262, 38028271, 38028303, 38045556, 37011644, 37017836, 37174339, 37174354, 37263855, 37263856, 37263858, 37313949, 37364768, 37479672, 37479677, 37479679, 37575306, 37644701, 37676913, 37693605, 37713605, 38000243, 38057455;    M0032PK42010001/ 60200193791;  M0033PK42022001/ 00193801, 00193802, 00193811, 00199272, 00199274, 60200189261;  M0033PK43022001/ 60200184921, 60200184931, 60200185352, 60200186011, 60200186831, 60200191921;  M0033PK62523001/ 60200191051, 60200191061;   80052/ 38134761, 38157260, 38157264, 38157268, 38157270, 38166555, 38166559, 38174961, 38185261, 38191257, 38205655, 38210562, 38221557, 38250460, 38250462, 38250463, 38290862, 38290863, 38290868, 38309361, 38311965, 38334860, 38343560, 38343561, 38363960, 38363962, 38363969, 38363970, 38371260, 38394060, 38395875, 38395880, 38408169, 38408179, 38408180, 38408185, 38408190, 38418762, 38429961, 38429971, 38439661,38453562, 38461768, 38461791, 38461792, 38476060, 38492479, 38492480, 38503762, 38503765, 38531178, 38544363, 38544364, 38557054, 38557055, 38557056, 38572773, 38572776, 38586760, 38586765, 38586767, 38586773, 38610462, 38623662, 38623663, 38623664, 38636560, 38636564, 38649866, 38660362, 38660363, 38660364, 38660365, 38660366, 38660367, 38660369, 38660370, 38671367, 38682671, 38682672, 38763862, 38131857, 38146859, 38146881, 38157255, 38157256, 38157276, 38157278, 38157279, 38166556, 38166562, 38166563, 38166564, 38166566, 38166567, 38166571, 38166572, 38166573, 38166574, 38166575, 38166576, 38166577, 38166578, 38166579, 38166580, 38166583, 38166584, 38166586, 38166587, 38166588, 38166589, 38166590, 38166591, 38166592, 38166593, 38166594, 38166595, 38166596, 38166597, 38166598, 38166599, 38166600, 38166601, 38166602, 38166605, 38166606, 38166607, 38166608, 38166609, 38166610, 38166611, 38174956, 38174966, 38174967, 38174972, 38174974, 38191271, 38195455, 38195458, 38210561, 38210565, 38210567, 38210569, 38210570, 38210572, 38210574, 38221560, 38221563, 38226159, 38226160, 38226161, 38226162, 38226163, 38226164, 38226165, 38226166, 38226167, 38226176, 38229556, 38229558, 38229576, 38239160, 38250456, 38250465, 38250470, 38250471, 38250472, 38250473, 38250474, 38250475, 38250476, 38250477, 38250478, 38250479, 38253255, 38261563, 38261564, 38261565, 38261566, 38261576, 38261577, 38271765, 38271766, 38271767, 38271768, 38271769, 38271770, 38271771, 38271772, 38271773, 38271774, 38271775, 38271776, 38271777, 38271778, 38271779, 38271780, 38271781, 38271782, 38271783, 38271784, 38271785, 38271786, 38271787, 38271788, 38271789, 38271790, 38271791, 38271792, 38271793, 38271794, 38271795, 38271796, 38271797, 38271798, 38271799, 38271800, 38271801, 38271802, 38271803, 38271804, 38271805, 38271806, 38271807, 38271808, 38271809, 38271810, 38271811, 38271812, 38281756, 38281759, 38281761, 38281766, 38311968, 38311970, 38311973, 38322364, 38322365, 38322366, 38322367, 38322368, 38322383, 38334862, 38343567, 38343568, 38343572, 38343581, 38352169, 38352171, 38363966, 38363967, 38363971, 38363972, 38363973, 38363974, 38363975, 38363976, 38363977, 38363978, 38363979, 38363980, 38377965, 38377966, 38377967, 38377968, 38377983, 38377984, 38377985, 38377986, 38395864, 38395866, 38395867, 38395886, 38408161, 38408162, 38408168, 38408183, 38408184, 38408191, 38418558, 38429960, 38429966, 38429967, 38439665, 38449963, 38449964, 38449971, 38449972, 38461767, 38461769, 38461777, 38461779, 38461780, 38461787, 38461788, 38492465, 38492467, 38492469, 38492474, 38503774, 38503775, 38516766, 38525960, 38525961, 38531162, 38531165, 38531166, 38531167, 38531168, 38531169, 38531170, 38531171, 38531172, 38531174, 38531175, 38539060, 38539061, 38544361, 38572761, 38572762, 38572763, 38572764, 38572765, 38572767, 38572770, 38572777, 38572778, 38586768, 38610460, 38610461, 38610465, 38619160, 38623660, 38623665, 38623666, 38623667, 38623668, 38623669, 38623674, 38623675, 38623676, 38623677, 38623678, 38623679, 38623680, 38623681, 38623682, 38623683, 38623684, 38623685, 38623686, 38623687, 38623688, 38636574, 38636575, 38636578, 38636579, 38636580, 38636581, 38636582, 38636585, 38649862, 38649863, 38649865, 38660361, 38660371, 38660373, 38660381, 38660382, 38671361, 38671370, 38671371, 38671372, 38671373, 38682661, 38682663, 38682664, 38682665, 38682667, 38682668, 38682670, 38682679, 38763863, 38813063, 39030961, 39030962, 39030963, 00197071, 00197072, 00199091, 00199092, 00199101, 00199102, 60200193751, 60200195801, 60200195802;"
    }
  ]
}