      "status": "Terminated",
      "city": "West Chester",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "65997",
      "recalling_firm": "Synthes USA HQ, Inc.",
      "address_1": "1302 Wrights Ln E",
      "address_2": "N/A",
      "postal_code": "19380-3417",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US (nationwide) and country of Mexico.",
      "recall_number": "Z-0114-2014",
      "product_description": "Synthes Application Instrument Sternal Zip Flex System    intended for use in primary or secondary closure/repair of the sternum following sternotomy or fracture of the sternum to stabilize the sternum and promote fusion",
      "product_quantity": "319",
      "reason_for_recall": "When using the 1st generation Zip Fix Application Instrument it is possible to cut the implant while the tensioning trigger is being squeezed which could lead to loosening of the implant.",
      "recall_initiation_date": "20130813",
      "center_classification_date": "20131030",
      "termination_date": "20160818",
      "report_date": "20131106",
      "code_info": "Part No.: 03.501.080 with multiple lot numbers: 3783492  3783913  3788496  3822332  7516728  7606881  7635218  7653178  7659168  7666085  7671934  7679825  7689244  7694377  7700691  7705157  7720599  7738572  7738573  7740498  7742713  7767497  7797648  7803768  7806881  7818677  7818682  7821672  7827088  7831855  7833606  7858407  8068078"
    }
  ]
}