  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "83811",
      "recalling_firm": "GE Healthcare, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "N/A",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide distribution to AZ, GA, MT, NC, NJ, NY, OH, WI.    International distribution to Australia, Bahrain, Bangladesh, Brazil, Cambodia, Chile, China, Colombia, Cyprus, Hong Kong, India, Iraq, Ireland, Italy, Kuwait, Lebanon, Malaysia, Mauritius, Mexico, MEXICO, Myanmar, Nepal, Oman, Panama, Peru, Philippines, Saudi Arabia, Singapore, Sri Lanka, Switzerland, Taiwan, Thailand, United Arab Emirates, United Kingdom, Uruguay, Vietnam",
      "recall_number": "Z-0113-2020",
      "product_description": "Achilles Express Bone Sonometer",
      "product_quantity": "241 devices",
      "reason_for_recall": "Certain Achilles Express and Achilles Insight systems were shipped with European CEE-7/7 Type Power Cords to countries where three protruding pin plugs are required for protective earth connection.",
      "recall_initiation_date": "20190903",
      "center_classification_date": "20191010",
      "termination_date": "20210216",
      "report_date": "20191016",
      "code_info": "System ID: ACE34924, AWE11358, ACE34606, ACE34652, ACE34651, AWE11894, AWE1196, ACE34927, ACE886H, AWE10447, AWE11047, AWE11098, AWE11304, AWE11312, AWE11317, AWE11381, AWE11285, ACH+12937, ACH+12932, ACH+12939, ACH+12935, ACH+12942, ACH+12936, ACH+12943, ACH+12938, ACH+12945, ACH+12940, ACH+12947, ACH+12952, ACH+12948, ACH+12871, ACH+12933, ACH+12872, ACH+12931, ACH+12290, ACH+12955, ACH+12975, ACH+12978, ACH+12983GA, ACH+12986, 2798401, 2815867, 2858452, 2876988, 2693419, 0002722584A, ACH10439, ACH31114, ACH10346, ACH10364, ACH10393, 828330086, 828330087,828330097, 828330098, 828330115, 828330091, 828330099, 828330119, 828330117, 83033100400411, 83033100024611, 83033000014510, 83033005000787, 83033004563599, 83033005000579, 83033005002887, 83033005003917, 83033000410110, 83033000415311, 83033004190909, 83033005000938, 83033005000978, 83033005002399, 83033009010429, 0002765007,   0002549803A, 83033000000018, 83033001234568, 83033009010046, 83033206010165, 83033209010165, 83033240415211, 83033245000577, 83033000011810, 0002830095, 0002960988, 0004225139, 0004241217, 83033820408610, 83033825000089, 83033825000098, 83033825000359, 83033825100325, 83033825101736, 83033826101736, 83033860068211, 83033200414211, 83033200012210, 83033200407710, 83033200407711, 83033200407712, 83033200413211, 83033205001468, 83033205105099, 83033205355099, 83033200012010, 83033200024711, 83033205104326, 83037200413111, 83033205002496, 83033205000068, 83033205004238, 83033605001326, 83033205102295, 83033209026036, 83033209026076, 0002540710, 0002767636, 0002767636A, 0002767636B, 0002767636C, 0002767636D, 83033005102305, 83033825000156, 83033205101507, 83033205101456, 83033205007588, 83033805004289, 83033205004228, 83033065000199, 83033205002037, 83033005105137, 83033806695010, 83033805001126, 0002790424, 83033600000008, 83033600308711, 83033605105157, IE1574LU01, A179802601, A258022402, 834334061, 834334010, 834334026, 834334028, 834334029, 834334004, 834334007, 834334006, 834334054, 834334055, 083433NZM30581, 083433NZM30585, 083433NZM30662, 83033008881210, 83033008881211, 0002501112B, 0002508220A, 0002791713, 0002818827, 0002732489, 0002739062, ACH+11056, 843334009, 084333NZM31123, 843334218, SA2155XR01, 847334059, 6073UL0164, 83033805200666, 853330027, 853330028, 853330029, 853330030, 853330031, 853330032, 853330033, 853330034, 853330057, 853330059, 853330060, 853330061, 853330062, 853330065, 853330072, 853330074, 853330080, 853330081, 853330082, 853330083, 853330084, 853330085, 856330003, 856330029, 856330002, 856330013, 856330019, 856330014, 856330016, 085633NZM10512, 085633NZM10524, 085633NZM11414, 085633NZM11693, 085633NZM12132, 085633NZM12134, 085633NZM12754, 085633NZM12755, 085633NZM12756, 856330009, 856330034, 856330007, 856330008, 856330004, 20181RAD02, 20181RAD03, 10423RAD02, 10423RAD03, 10423RAD04, 10423RAD05, 10423RAD06, 10423RAD07, 10423RAD08, 10423RAD09, 10423RAD10, 10423RAD11, 10423RAD12, 10423RAD13, 20356RAD03, 20357RAD01, 10422RAD02, 20355RAD02, 20109RAD02, 21001RAD01, 21008RAD01, 21008RAD02, 21008RAD03, 20358RAD01, 0002593515X, 0002638962, 859334009"
    }
  ]
}