{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "63027",
      "recalling_firm": "GE Healthcare, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "N/A",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "N/A",
      "distribution_pattern": "Nationwide distribution",
      "recall_number": "Z-0113-2013",
      "product_description": "GE Healthcare innova 3100, Mobile X-ray system.     The innova systems are indicated for use in generating fluoroscopic images of human anatomy for vascular angiography, diagnostic and interventional procedures, and optionally , rotational angiography procedures.  They are also indicated for generating fluoroscopic images of human anatomy for cardiology, diagnostic, and interventional procedures.",
      "product_quantity": "199 installed in the US",
      "reason_for_recall": "It was discovered of a potential failure of the spectral filter mechanism inside the collimator of the innova 3100, innova 3100-IQ and innova 3131-IQ systems.",
      "recall_initiation_date": "20111220",
      "center_classification_date": "20121022",
      "termination_date": "20130918",
      "report_date": "20121031",
      "code_info": "Model number 5892935G052G"
    }
  ]
}