{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Bolton",
      "state": "N/A",
      "country": "United Kingdom",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "90849",
      "recalling_firm": "Medicina Uk Ltd",
      "address_1": "Unit 2",
      "address_2": "Rivington View Business Park",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Nationwide",
      "recall_number": "Z-0112-2023",
      "product_description": "Medicina Reusable Oral syringe 1ml  Code: OTH01",
      "product_quantity": "111 boxes (11,100 syringes)",
      "reason_for_recall": "Plunger component has been manufactured with an unapproved material (Polyethylene not Polypropylene) and does not meet raw material specification",
      "recall_initiation_date": "20220704",
      "center_classification_date": "20221020",
      "report_date": "20221026",
      "code_info": "UDI-DI:  (01)05056115402869 (01)05056115402869(17)260700(10)HM21053   Batch number: HM18041, HM21053",
      "more_code_info": ""
    }
  ]
}