{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-06-17",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80323",
      "recalling_firm": "GE Medical Systems, LLC",
      "address_1": "3200 N Grandview Blvd",
      "address_2": "N/A",
      "postal_code": "53188-1693",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution: US (nationwide) including states of:: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT,  WA, WI, WV; and countries of: Algeria, Argentina, Australia, Austria, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, CHINA, Colombia, Costa Rica, Croatia, CYPRUS, Czech Republic, DENMARK, Dominican Republic, Ecuador, Egypt, Estonia, Finland, France, Georgia, Greece, Guadeloupe, Guatemala, Guyana, Honduras, Hong Kong, Iceland, India, Indonesia, Iran, IRAQ, Ireland, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea (Republic of), Kuwait, Latvia, Lebanon, Libyan Arab Jamahiriya, Lithuania, Luxembourg, Macedonia, Malaysia, Mexico, Netherlands, New Zealand, Norway, Oman, Pakistan, Paraguay, Philippines, Portugal, Puerto Rico, Russia, Saudi Arabia, SINGAPORE, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Syria, Taiwan, Thailand, Turkey, United Arab Emirates, Uruguay, Uzbekistan, Venezuela, and Vietnam.",
      "recall_number": "Z-0112-2019",
      "product_description": "GE Healthcare MRI Systems, Software Version 15.x or HD 16 sold under the following product names:      *ER*15T HDXT ECHOSPEED 16 CHANNEL, *OC*1.5T HDX ECHOSPEED 8 CHANNEL, *SC*1.5T HD TO HDX UPGRADE, *TR*15T HDXT 16CH, *TR*3.0T HDX TWINSPEED 16 CHANNEL, 1,5T HDXT, 1.5 HD TO HDX UPGRADE, 1.5 HDX, 1.5 HDX ECHOSPEED 16 CH, 1.5 HDX ECHOSPEED 8 CHANNEL, 1.5 HDXT, 1.5 HDXT-16 , 1.5 LX TO HDX 16C FORKLIFT UPG, 1.5 LX TO HDX 8C FORKLIFT UPG, 1.5 T UPGRADE TO HDX, 1.5EXCIT TO HDXT UPG, 1.5T  HDX  16 CHANNEL, 1.5T 16CH UPG, 1.5T 6 CH HDX , UPGRADE, 1.5T 8CH HDXT TO 16CH HDXT UPG, 1.5T CRM LX TO HDX UPG, 1.5T ES HDX W/ONCOLOGY PACK, 1.5T EXC.TO HDX UPG, 1.5T EXC.TO HDXT UPG, 1.5T EXCITE HD TO HDX UPG, 1.5T EXCITE TO HDX UPG, 1.5T EXCITE TO HDX UPG, 1.5T EXCITE TO HDX UPG. W/ACGD, 1.5T EXCITE TO HDXT UPG, 1.5T EXCITE TO HDXT UPGRADE, 1.5T EX-HDX, 1.5T EX-HDXTWIN, 1.5T HD ES/HS MOBILE, 1.5T HD TO HDX UPG. 1.5T HD TO HDX UPGRADE, 1.5T HD TO HDXT UPG, 1.5T HD TO HDXT UPGRADE  1.5T HDI,1.5T HDI 16CH FIXED, 1.5T HDI 8CH FIXED, 1.5T HDI ECHOSPEED  1.5T HDI ECHOSPEED 8 CHANNEL, 1.5T HDI ECHOSPEED 8 CHANNEL  1.5T HDI ECHOSPEED 8 CHANNEL., 1.5T HDI ECHOSPEED 8 CHANNEL,   1.5T HDI ECHOSPEED 8 CHAN¿SO2762285, 1.5T HDI ECHOSPEED 8CH,   1.5T HDI ECHOSPEED MOBILE, 1.5T HDI ECSHOSPEED 8CH, 1.5T HDI TO HDXT UPGRADE, 1.5T HDX, 1.5T HDX - MEIRAV (BREAST), 1.5T HDX 8CM  1.5T HDX ECHOSPD 16, 1.5T HDX ECHOSPEED, 1.5T HDX ECHOSPEED   1.5T HDX ECHOSPEED 16, 1.5T HDX ECHOSPEED 16 CHANNEL, 1.5T HDX ECHOSPEED 16 CHANNEL, 1.5T HDX ECHOSPEED 16 CHANNELS, 1.5T HDX ECHOSPEED 16CH¿SO2659508, 1.5T HDX ECHOSPEED 16CH-SO2718893, 1.5T HDX ECHOSPEED 32 CHANNEL, 1.5T HDX ECHOSPEED 8 CH, 1.5T HDX ECHOSPEED 8 CHANNEL, 1.5T HDX ECHOSPEED MOBILE  1.5T HDX ECHOSPEED MRI, 1.5T HDX ES, 1.5T HDX ES 16, 1.5T HDX TWINSPEED, 1.5T HDX TWINSPEED 16 CHANNEL, 1.5T HDXT  1.5T HDXT 16CH,1.5T HDXT 16CH CXK4, 1.5T HDXT 16CH ES FIXED  1.5T HDXT 8 CH, 1.5T HDXT 8CH ES FIXED, 1.5T HDXT 8CH UP, 1.5T HDXT ECHOSPEE, 1.5T HDXT ECHOSPEED, 1.5T HDXT ECHOSPEED 16 CH  1.5T HDXT ECHOSPEED 16 CHAN, 1.5T HDXT ECHOSPEED 16 CHANNEL  1.5T HDXT ECHOSPEED 8, 1.5T HDXT ECHOSPEED 8 CHANN, 1.5T HDXT ECHOSPEED 8 CHANNEL, 1.5T HDXT ECHOSPEED 8 CHANNEL , 1.5T HDXT ECHOSPEED 8_CHANNEL, 1.5T HDXT ECHOSPEED MOBILE, 1.5T HDXT ES 8 CH, 1.5T HDXT GS, 1.5T HDXT TWIN, 1.5T HDXT TWINSPEE, 1.5T HDXT TWINSPEED 16 CHAN,1.5T HDXT UPG, 1.5T HDXTID, 1.5T LX HDX, 1.5T LX TO 16 CH HDXT UPG, 1.5T LX TO 16 CHAN, 1.5T LX TO 16 CHANNEL HDX UPG, 1.5T LX TO 16 CHANNEL HDX UPG, 1.5T LX TO 16 CHANNEL HDX UPGRADE, 1.5T LX TO 16 CHANNEL HDXT, 1.5T LX TO 16 CHANNEL HDXT UPG, 1.5T LX TO 16CHANNEL HDX U/G2777896, 1.5T LX TO 8 CH HDX UPG, 1.5T LX TO 8 CH HDXT UPG, 1.5T LX TO 8 CHANNEL HDX UP, 1.5T LX TO 8 CHANNEL HDX UPG, 1.5T LX TO 8 CHANNEL HDX UPG, 1.5T LX TO 8 CHANNEL HDX UPG.827015, 1.5T LX TO 8 CHANNEL HDXT UPG, 1.5T LX TO 8 CHANNEL HDXT UPGRADE, 1.5T LX TO HDXT  1.5T NON LCC4X5X TO HDXT UPGD, 1.5T NON-LCC LX/HRZ 5.5 TO 1.5T HDX, 1.5T SIGNA HDX UPG, 1.5T SIGNA HDXT, 1.5TEXCITE TO HDX UPG W/O A, 1.5TEXCITE TO HDX UPG W/O ACGD, 1.5TNONLCC LXHZ TO HDXT UPG, 1.5TNONLCC LXHZ TO HDXUPGW/ACD, 1.5TNONLCC4X5X TO HDXUPGW/OACD, 1.5TNONLCCLXHZ TO HDXUPGW/O ACD, 1¿5T 8CH HDXT MRI, 1¿5T HDXT, 1¿5T HDXT MRI, 1-5T CRM LX TO HDX , G.W/OACGD, 1-5T CRM LX TO HDXT UPG, 1-5T EXCITE TO  15T EXCITE TO HDX UPG, 1-5T HDI ECHOSPEED 16 CHANNEL, 15T HDI ECHOSPEED 8 CHANNEL, 1-5T HDX ECHOSPEED 16 CHANN, 15T HDX ECHOSPEED 16 CHANNEL, 15T HDX ECHOSPEED 8 CHANNEL, 15T HDX ECHOSPEED MOBILE, 1-5T HDX ECHOSPEED MOBILE, 1-5T HDX ECHOSPEED TRANSPORT, 1-5T HDX TWINSPEED 8 CHANNEL, 1-5T HDXT ECHOSPEED 16, 1-5T HDXT ECHOSPEED 16 CHAN, 1-5T HDXT ECHOSPEED 16 CHAN-HD16.0, 1-5T HDXT ECHOSPEED 16 CHANNEL,   1-5T HDXT ECHOSPEED 16 CHANNEL, 1-5T HDXT ECHOSPEED 16 UPGRADE, 1-5T HDXT ECHOSPEED 8 CHANN, 15T HDXT ECHOSPEED 8 CHANNEL, 1-5T HDXT ECHOSPEED MOBILE, 1-5T HDXT ECHOSPEED RELOCATABLE, 1-5T LX TO 16 CHANNEL HDX UPG, 1-5T LX TO 16 CHANNEL HDXT, 1-5T LX TO 8 CHANNEL HDXT U, 1-5T LX TO",
      "product_quantity": "5,200 total units",
      "reason_for_recall": "The latest on-site software version may not have been reinstalled at some sites after service activities were performed resulting in images possibly  flipped left/right  and/or there can be patient data mismatch.",
      "recall_initiation_date": "20180607",
      "center_classification_date": "20181015",
      "termination_date": "20200618",
      "report_date": "20181024",
      "code_info": "Model Number, Serial Number (System ID) :    N/A, 000000000R7742 (MP697620); N/A, 00000834BD4B76 (M2865810);   46-294231G1, 0002005-0607-9 (80040MRS02); N/A, 00000293242MR4 (GOLD00EU79); 5127452, 00000274388MR8 (M4160077); 2226300, 000000000R7586 (PL2130MR01); 2226300, 000000000R2785 (A5106403);   N/A, N/A (083027060000499); 2226300, 000000000R5870 (A5177224);   N/A, 000000000R4933 (C4194430); 2226300, 000000000R7628 (PL1807MR01); 5911000-4 , 00001201FMM0FJ (083027829345912);   2226300, 000000000R5979 (818713MR1); 5308573, 00000292202MR9 (706378MR2); 46-294231G1, 002003-1211-30 (KW1005MR02); 2226300, 000000000R5829 (PL1677MR01); 46-294231G1, 000099-1119-10 (A5401507); N/A, N/A (083027040099711); 46-294231G1, 002001-0530-21 (0920279017); 2377062-61, 00000010633YR6 (808433TMR2);   46-294231G1, 002001-1001-91 (PC1961MR01); 5308573, 00000300778MR8 (0856270046); 2226300, 000000000R5217 (PL1649MR01); N/A, N/A (PL0195MR01); 2226300, 000000000R4456 (70772MRS90); 5127452, 000000HDMR6307 (DK1297MR01); 2226300, 000000000R1970 (M4018076);   46-294231G1, 0000094-1209-4 (178SIGNA15); 2377062-2, 00000963500YM4 (0910272057); 46-294231G1, 002004-0211-05 (PC2703MR02); 46-294231G1, 002001-0301-23 (YM0731); 46-294231G1, 002003-0128-22 (YM2020); 2226300, 000000000R5133 (KAISERVJOMR); 5127452-3, 00001011BD6645 (082427040110); 46-294231G1, 0000098-0617-2 (414SIGNAHOR); 5308573, 00000294505MR3 (34407MRS01); 5107849, 00000275094MR1 (DZ1004MR02); 2226300, 000000000R6384 (0850270266); N/A, 00000000R7803A (2900429); 2214149-2, 00000003388M26 (MRR8311); N/A, 000000000R6449 (110040MR01); 2214149-2, 00000001634M25 (MRR8049); /A, N/A (0910272067); 2226300, 000000000R6691 (A200581902);   2226300, 000000000R6343 (787831SIGNAHDX); 2226300, 000000000R6461 (SA2161MR01); 2226300, 000000000R7934 (78003MRS01); 5308573, 00000291238MR4 (MA1135MR01); 5308573, 00000299348MR3 (07025MRS01); 2226300, 000000000R7004 (PL1426MR02); 2226300, 000000000R7487 (1306777); 2226300, 000000000R6769 (IL1041MR08); 395001, 00000280692MR5 (CC0820MR09); 2226300, 000000000R6889 (PL1735MR01); N/A, 000000000R6609 (M4464817); 2226300, 000000000R6707 (083027207801078); N/A, 00000014680GQ3 (M4033099);   5127452, 000000HDMR3604 (01203MRS01); N/A, N/A (0910274028);   2226300, 000000000R6050 (10563MRS01); 2226300, 000000000R5325 (SA1056MR01); 2226300, 000000000R6042 (SA1047MR01);   46-294231G1, 0002001-0219-8 (281477MWMR); 2226300, 000000000R5823 (A5718006); 2226300, 000000000R6111 (DZ1250MR01); N/A, N/A (DK1228MR07); 46-294231G1, 0000098-1207-9 (IE1028MR02); 2226300, 000000000R5149 (00398MRS01); 2226300, 000000000R5639 (00441MRS01);   2226300, 000000000R7809 (CBSMR1); 2226300, 000000000R5836 (083027862100837); 46-294231G1, 000004-0528-17 (083027602200104);   2226300, 000000000759 (083027603014911); N/A, 002004-0323-16 (GOLD0EU131); 2226300, 000000000R7996 (MRR7977); 2226300, 000000000R7781 (PL1639MR02); N/A, 000000000R7452 (2869229);   N/A, N/A (HC2821MR01); 5308573, 00000294659MR8 (083027848131849);   46-294231G1, 0002000-1003-3 (0835270003); 2226300, 000000000R7237 (M4212028); 2395001, 00000276676MR4 (083027842100686); N/A, N/A (0910272088); N/A, N/A (0859270005); 2226300, 000000000R7837 (2913994HDXT); 5127452-3, 00001011BD667C (DZ1005MR01); 2226300, 000000000R7563 (IR1906MR01); 5911000-4 , 00001239FMM01H (083027800039612); 5127452-3, 00000919BD5B8F (FI1024MR02);   2226300, 000000000R7940 (SA2214MR01); 5127452-3, 00000837BD4ESC (0850270435); 46-294231G1, 000006-0523-15 (0850270200); 5127452-3, 00001003BD6435 (083027601000510); 5308573, 00000302567MR3 (083027243783711); 2377062, 00000941265YM1 (600118MR02); N/A, N/A (0850270291); 2395001-3, 00000289020MR0 (083027202100488); 2226300, 000000000R2404 (PL1169MR01); N/A, N/A (083027601013710); N/A, N/A (0910274011); 46-294231G1, 0000099-0506-1 (0910272025); 5127452, 000000HDMR5710 (IE1038MR01); N/A, N/A (083027001018110); N/A, N/A (083027592052828); 2226300, 000000000R4321 (SA1220MR01); 46-294231G1, 002002-1015-30 (PC0114MR04); 46-294231G1, 002003-0820-07 (PL1394MR01); 46-294231G1, 002001-0619-24 (IE1006MR01); 2226300, 0000000000R891 (0920279013); N/A, N/A (0910272038); 46-294231G1, 002000-0418-17 (SA1001MR02); 46-294231G1, 0`2001-0219-13 (259170MR05); N/A, 0000099-1217-4 (10140MRM44); N/A, 00000303865MR0 (10331MRS01); 46-294231G1, 0000096-0614-8 (801224MR); N/A, 0000096-0520-5 (AE1025MR01); 46-294231G1, 002001-0709-20 (0856270038);   5308573, 00000301170MR7 (MRR8186); 5127452, 000000HDMR6250 (00213MRS01); 46-294231G1, 002004-0917-09 (600092MR03); N/A, N/A (0856270050); 46-294231G1, 0000097-0922-9 (213340MR); 46-294231G1, 0000094-0928-7 (4224SIGN15); 5127452, 000000HDMR5751 (70383MRS01);   N/A, N/A (083027592002609); 5127452-3, 00000903BD55FD (083027002100119); 2226300, 000000000R7217 (083027848480709);   5127452, 00000HDMR10911 (801262MR2); 46-294231G1, 002001-0425-10 (765INNERMR); 2377062, 00000962578YM1 (519685MR); 2377062-61, 00000009061YR3 (DK1401MR01); 2226300, 000000000R6839 (501268MR);   5308573, 00000301743MR1 (45019MRS01); 5127452, 000000HDMR6310 (0834270012); 2226300, 000000000R6802 (H10649MR06); 5107849-2, 00000288195MR1 (H10652MR08); 2395001, 00000286008MR8 (21234MRM02); 2108800-2, 535651 (SHIMR02604); 2226300, 000000000R6903 (503763MR1); 46-294231G1, 002002-0627-18 (RADNETMR611); N/A, N/A (083027820006912); N/A, N/A (4954MR15T);   N/A, 000000000R6911 (2805953); 46-294231G1, 0000091-1022-5 (205431MR2); 5911000-4 ,00001227FMM06G (083027200018712);   N/A, N/A (083027590475005); N/A, N/A (1111395); 2214149-2, 00000005560M28 (IQ1394MR01); N/A, N/A (RADNETMRM01); 46-294231G1, 002005-0627-18 (ALLIANCESIG421); 46-294231G1, 0000099-0125-9 (907276ARMR); N/A, N/A (6444MRDICO); N/A, N/A (2809956); 46-294231G1, 0000001-0215-1 (INSIGHTMR1915); 2226300, 000000000R6838 (623445MR1); 2377062-2, 00000943572YM8 (201632MR); 46-294231G1, 0002002-0510-7 (ALLIANCESIG474); 46-294231G1, 002002-0125-22 (559585CCICMR2); 46-294231G1, 002002-0405-12 (205783TSMR);   5127452, 000000HDMR5415 (850484OMRI); 5308573, 00000300970MR1 (RADNETMR507); 5127452, 00000HDMR10886 (916423MR); 5127452, 000000HDMR3674 (501370CHMR); 5107849, 00000269580MR7 (PL3040MR01); 46-294231G1, 002004-1209-10 (YM1260); 2395001-2, 00000284023MR9 (083027002101567); 46-294231G1, 0000007-403-15 (083027802000517); 5127452, 000000HDMR5483 (1063THDX); 5127452, 000000HDMR3964 (SA1001MR03); 2395001-2, 00000280672MR7 (00440MRS02); N/A, 000000HDMR5233 (110037MR02); 5127452-3, 00001012BD66E0 (083027000143110); 2358431, 00000000WB0607 (AZ4013MR01); 5127452-3, 00001009BD6557 (BA4080MR01); 5308573-2, 00000299495MR2 (MRWB0540); N/A, N/A (5606MRS3T); 5127452-3, 00001009BD65F2 (083027009445910); N/A, N/A (773665MR3T); 5127452-3, 00001009BD6579 (2887688HDXT30T); 5127452, 00000275305MR1 (831423MR); N/A, N/A (2558948); 46-294231G1, 0002005-0607-4 (7067213TMR); 5127452-3, 000S0933BD5DFD (719542MR); N/A, 002003-0912-12 (DK1047MR04); 5127452, 000000HDMR5406 (0910273033); 5127452, 000000HDMR2642 (0850270218); 5421824-2, 00000310388MR4 (083027240009214); 2377062-2, 00000963651YM5 (IL1194MR01); 46-294231G1, 002005-1020-29 (HBLUCOMR01); N/A, 002004-0917-02 (00473MRS01); N/A, N/A (MR4013522); N/A, N/A (2679752); N/A, N/A (2541440); 2226300, 000000000R1733 (MRR1733); 2226300, 0000000000R273 (MRR273); 46-294231G1, 000099-0205-11 (0910272023);   N/A, N/A (PL1681MR01); N/A, N/A (H4443MR04); 5308573, 00000301741MR5 (MRR8267); 5308573, 00000301223MR4 (2918179MR); 2214149-2, 00000002339M20 (MRR8325); 2214149-2, 00000003155M29 (MRR8701); N/A, 000096-0126-27 (A020MR02); N/A, 00000175300MR3 (10079MRM42);   N/A, N/A (5272MRSIGNA4X); 2226300, 000000000R5436 (DEN00700);   2226300, 000000000R6601 (CE165329); N/A, N/A (0910273026);   46-294231G1, 002002-1126-23 (082427100148); 5308573, 00000300700MR2 (DEN00933); 46-294231G1, 000099-0122-13 (YE600210); 5127452, 000000HDMR5250 (DZ1009MR02); 2226300, 000000000R3823 (CS1275MR02); 2226300, 000000000R2680 (ZA2159MR02); 2214149-2, 00000014057GQ4 (082427030054); 2226300, 000000000R5680 (082427130073); 46-294231G1, 002005-1201-30 (HU1039MR03); 2226300, 000000000R6212 (DK1306MR02); 2214149-2 ,00000007668M27 (082427280013); 5308573, 00000301498MR2 (082427240016); 2226300, 000000000R7667 (082427050024); 5308573, 00000301500MR5 (MRR8257);   2226300, 000000000R6024 (CS1020MR01); 5308573, 00000291246MR7 (2761886); N/A, 000000000R6222 (2726027); 5308573, 00000301383MR6 (082427030063); 2226300, 000000000R7426 (082427100239); 46-294231G1, 002001-0718-35 (0910273020); 5308573, 00000301264MR8 (082427030061);   2226300, 000000000R6289 (082427160057); 2226300, 000000000R6200 (MPX19700); 2226300, 000000000R7028 (082427210051); 2226300, 000000000R6582 (DZ1259MR01); 2226300, 000000000R5895 (SA1129MR02); 2226300, 000000000R6955 (2807539); 2226300, 000000000R5724 (PPS18202); 2226300, 000000000R5733 (RU1550MR01);   2226300, 000000000R6139 (BG4538MR01); 2226300, 000000000R6439 (RU1279MR01); 2226300, 000000000R5352 (GE1004MR01); N/A, 000000000R6511 (2767032); 2214149-2, 00000004486M27 (082427080033);   2226300, 000000000R6806 (082427040088); 2226300, 000000000R5120 (0910273030); 2226300, 000000000R7718 (082427250022); 2226300, 000000000R5579 (RU1067MR02); N/A, N/A (082427040098);   5421824-2, 00000310944MR4 (082427080047); 2226300, 000000000R7870 (MRR7870); 46-294231G1, 000099-0817-11 (MPX47002); 5308573, 00000298557MR0 (0853270042IB1); 5421824, 00000311257MR0 (082427190081); 2214149-2, 00000002805M20 (082427040129); 2226300, 000000000R7050 (UPW00400); 2226300, 000000000R7884 (NPV85600); 2214149-2, 00000002750M28 (CS1338MR01); 2226300, 000000000R7342 (082427030049); 2214149-2, 00000003393M26 (RU1031MR01); 2226300, 000000000R7130 (CS1319MR01); 5127452-3, 00001012BD66A2 (0853270074IB1); 46-294231G1, 0002005-1201-2 (082427120126);   N/A, N/A (082427160098); 46-294231G1, 0000099-0414-7 (CP223906);   N/A, 000000000R3618 (BG4641MR01); 46-294231G1, 002002-1030-14 (H3943MR01); 46-294231G1, 002001-1018-36 (ZA2103MR01); 2214149-2, 00000005509M25 (082427230084); 5421824, 00000311572MR2 (082427140132); 5269997-9, 00000015025YR0 (0828270033); 2226300, 000000000R5260 (082427030103); 46-294231G1, 0000095-1120-5 (RU1089MR01); 5421824, 00000311574MR8 (082427230107); 2377062-61, 00000011084YR1 (27008MRS01); 2214149-2, 00000003870M23 (RU1457MR01); 5127452-3 ,00000838BD4ECB (35062MRS02); 7849-2, 00000283036MR2 (55027MRS01); 2395001, 00000285169MR9 (JO4109MR01); 5421824-2, 00000311280MR2 (082427230097); 507849-2, 00000288588MR7 (DK1393MR01); 2214149-2 , 00000007860M20 (082427210142MM3); 2214149-2, 00000004304M22 (H2294MR07); 2226300, 000000000R7091 (48021MRS01); 2214149-2, 00000004492M25 (KA2839MR01); 2214149-2, 00000003509M27 (RU1173MR01);   2226300, 000000000R3065 (082427160161); N/A, N/A (570844HCM3T);   5127452, 000000HDMR3447 (DPM32124); 5421834-2, 00000311145MR7 (082427130084); 5127452-3, 00000919BD5BB7 (082427220041); N/A, N/A (082427120070); 5127452, 000000HDMR1125 (082427040073); 2395001-2, 00000275466MR1 (DEN51500); 5127452, 000000HDMR5758 (082427040092); N/A, 0000000HDMR357 (CS1006MR03); 5127452, 000000HDMR1900 (0910274020); 5127452-3, 00001003BD640C (082427080026); 5127452-3, 00000938BD5F17 (082427190069); 5127452, 00000275290MR5 (DPC79014); N/A, N/A (082427100192); 5308573-2, 00000299275MR8 (082427200036); 5127452-3, 00001012BD66E1 (PER08626); 5127452-3, 00001003BD640D (PER28007); 5127452-3, 00000938BD5F14 (RO1189MR01); 5127452, 000000HDMR6035 (082427140120); 5107849-3, 00000302786MR9 (MRJ061); N/A, 00000293706MR8 (NPV58000); N/A, N/A (0910276003); N/A, N/A (626570MR); N/A, N/A (907561AIMR); N/A, N/A (309451MR); 46-294231G1, 000002-1003-15 (201967MR); 2226300, 000000000R5771 (0853270133GS);   46-294231G1, 0000099-0521-6 (ALLIANCESIG85); 2377062-61, 00000011265YR6 (MRR6469); N/A, N/A (083027067005818); N/A, N/A (520GVHMR); N/A, N/A (2939576HDXT); N/A, N/A (520872MR); 2226300, 000000000R1970 (H10650MR08); 2226300, 000000000R6733 (A197998701);   2226300, 000000000R6448 (0828270030); 2214149-2 , 00000015124GQ1 (0834270011); N/A, 002000-0907-10 (CPM80615); 5127452, 000000HDMR5960 (0834270010); 5127452-3 , 00001009BD6578 (0834270015); 46-294231G1, 000096-0126-15 (YP630007); 2226300, 000000000R6352 (YP600200); 2226300, 000000000R2306 (RU1086MR01);   46-294231G1, 0002002-0506-3 (FEC80061); 2377062, 4337424 (MEX34410);   5127452, 000000HDMR4309 (UPW28900); 2395001, 00000280327MR8 (UPW21605); 2226300, 000000000R5291 (YP601000); 46-294231G1, 002000-0426-18 (0856270029); 46-294231G1, 0000099-0722-1 (ME208331);   2226300, 000000000R4276 (FPF35809); 46-294231G1, 002004-1117-12 (0828270020IB01); N/A, 00000001400M21 (NPV84508); 2377062-61         , 00000011251YR6 (MEX71105); 2395001, 00000283516MR3 (MP118825);   46-294231G1, 002003-1027-20 (A5103135); 5396469-2, 00000310306MR6 (NO1005MR02); 46-294231G1, 0000095-0810-3 (A5278401); 46-294231G1, 002003-0903-12 (A5111732); 2226300, 000000000R4164 (A5684809);   2226300, 000000000R6773 (PER08305); 2226300, 000000000R6502 (2766544); 2226300, 000000000R5403 (H1063MR18); 2226300, 000000000R5912 (H1420MR14); 46-294231G1, 002002-0809-08 (CPZ09609);   46-294231G1, 002003-0708-01 (PER00417); 46-294231G1, 002000-1013-15 (A5632405); 2214149-2, 00000002229M23 (MRR8295); 2226300, 000000000R2375 (A5745509); N/A, N/A (0853270100); N/A, N/A (083027100973317); N/A, 002002-0814-05 (H1066MR16); 46-294231G1, 000093-0528-15 (DE528434); 2226300, 000000000R7136 (0910272086);   2395001, 00000280778MR2 (0843270015); N/A, N/A (0910274013);   46-294231G1, 0000098-0608-5 (0910272020); 2395001, 00002835171MR1 (083027040006015); 2226300, 000000000R6574 (2767673); 2226300, 000000000R5159 (3882647967MROCC); 46-294231G1, 0002005-0331-3 (H12041MR02); 46-294231G1, 0000000MRU0407 (IE1025MR01);   2226300, 000000000R3756 (PVIL01MR01); 46-294231G1, 000097-1107-10 (PNUE01MR03); 46-294231G1, 002005-0823-17 (HPAD01MR01); N/A, N/A (PL1774MR02); 2226300, 000000000R6092 (0859270002); 2226300, 000000000R6021 (0859270001); 5127452, 000000HDMR1019 (RU3023MR01); N/A, 000000HDMR5374 (PM1633MR03);   2226300, 000000000R4692 (PL1704MR01); N/A, N/A (083027242000437);   2226300, 000000000R1663 (PC0976MR01); 46-294231G1, 0002001-0323-1 (MA1152MR01); 46-294231G1, 0000095-0912-5 (PX2261MR02); 5127452, 1529759 (MC6946MR07); 2395001-2, 00000278549MR1 (RU1127MR02);   46-294231G1, 000006-0523-10 (RU1834MR01); 5127452, 3877 (00240MRS01); 46-294231G1, 002000-0816-10 (PC0396MR03); 46-294231G1, 0000098-0528-5 (A5566913); 5127452, 000000HDMR2483 (A171995610); 2226300, 000000000R2753 (DK1343MR01); 2226300, 000000000R6499 (A5192410); N/A, 000000000R7233 (1324655); 5308573, 00000302145MR8 (MRR8303); 2226300, 000000000R6460 (22MRHDX2762068); 2226300, 000000000R5214 (H17641MR12);   N/A, 02001-12210-13 (A042MR01); 2226300, 000000000R7197 (2835341HDX); 2395001-2, 00000286650MR7 (H2193MR05); 2226300, 000000000R6520 (A51273121); 2226300, 000000000R2396 (MRR2396);   46-294231G1, 000007-0530-16 (UK28); 46-294231G1, 000097-0331-10 (DK1029MR01); 2214149-2 , 00000009304M27 (MRR8592); 2320288, 00000096546WH7 (10695MRM01); 2226300, 000000000R2753 (DK1394MR01); 2357500, 00000249262MR7 (16683THDX); 2226300, 000000000R7564 (A5146405); N/A, N/A (650742KSSFMR); 46-294231G1, 0000096-0419-4 (0920279006); N/A, N/A (916SRFMR); N/A, N/A (7366);   46-294231G1, 002002-0517-19 (IR1026MR01); N/A, N/A (0847270018);   46-294231G1, 0000099-0329-4 (A5127353); N/A, N/A (0847270025);   46-294231G1, 002001-1018-32 (PC7425MR01); N/A, 000000000R1116 (AH4511MR02); N/A, N/A (843549MR); 2226300, 000000000R1533 (LLS0120); 5308573, 00000293870MR2 (CC0820MR10); 46-294231G1, 002004-0713-22 (393115MR01); N/A, N/A (00200MRS03); N/A, N/A (ALLIANCESIG270); N/A, 0002000-1102-1 (DK1042MR01); 46-294231G1, 002004-0206-15 (397480MR01); N/A, N/A (707571KSRMR); N/A, N/A (8183MRLX); 2377062-2, 00000951076YM9 (805687MR2); 2226300, 0000000000R575 (HC1770MR01 ); N/A, N/A (A5183710); N/A, N/A (585786MRI); 2226300, 000000000R6712 (0847270023); 2226300, 000000000R5610 (0843270017); 46-294231G1, 002002-0315-13 0847270001); 46-294231G1, 0002000-0301-4 (SA1012MR01); N/A, N/A (570844TWINSPEED); N/A, N/A (422SIGNA15); N/A, 002000-1221-14 (DK1047MR03); 46-294231G1, 002003-0929-03 (N4192928); N/A, N/A (0910274008); 46-294231G1, 002003-0818-13 (34026MRS03); N/A, 00000275019MR8 (35045MRS03); 2226300, 000000000R5050 (082427130074); N/A, 00000310642MR4 (082427130082); 5421824, 00000311573MR0 (082427180036); 5453862-2, 308735MR0 (082427240048);   N/A, N/A (PL3331MR01); N/A, N/A (409637LX1); N/A, N/A (618798SMR)."
    }
  ]
}