      "status": "Terminated",
      "city": "Wayne",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "66144",
      "recalling_firm": "Fujinon Inc",
      "address_1": "10 Highpoint Dr",
      "address_2": "N/A",
      "postal_code": "07470",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution: US (nationwide) including states of: NJ, PA and VA; and country of Canada.",
      "recall_number": "Z-0112-2014",
      "product_description": "Fujinon/Fujifilm Ultrasonic Endoscope    Intended to provide ultrasonic images of submucosal and peripheral organs of the upper gastrointestinal tract for observation, diagnosis and endoscopic treatment.",
      "product_quantity": "12 Units",
      "reason_for_recall": "A limited number of endoscopes were shipped to FMSU-ESD with a manufacturing defect which included missing set screws used to more securely attach the Forceps Inlet (biopsy inlet port) to the endoscope.",
      "recall_initiation_date": "20130731",
      "center_classification_date": "20131029",
      "termination_date": "20140304",
      "report_date": "20131106",
      "code_info": "Model Number EG-530UT2.  Serial Numbers: NU038A018, NU038A019, NU038A022, NU038A027, NU038A028 and NU038A029"
    }
  ]
}