      "country": "United States",
      "city": "Indianapolis",
      "address_1": "9115 Hague Rd",
      "reason_for_recall": "Roche has received four reports of injury caused by the sharp edge of the center guide rail of an  analyzer\u0019s sample rack tray. The cleaning procedure for the sample rack tray has been revised to clarify the cleaning process.",
      "address_2": "",
      "product_quantity": "9014 in total",
      "code_info": "Not applicable",
      "center_classification_date": "20161025",
      "distribution_pattern": "US Nationwide Distribution",
      "state": "IN",
      "product_description": "cobas 6800/8800 system  Hepatitis Viral B DNA Detection    Product Usage:  The cobas p512 pre-analytical system is a computer controlled fully automatic system for sorting of open and closed barcoded and centrifuged sample tubes. It includes modules for registration and decapping of sample tubes, liquid level detection (optional) and sample quality detection (optional), as well as recapping of sample tubes (optional).",
      "report_date": "20161102",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Roche Diagnostics Operations, Inc.",
      "recall_number": "Z-0110-2017",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "75186",
      "termination_date": "20170605",
      "more_code_info": "",
      "recall_initiation_date": "20160715",
      "postal_code": "46256-1025",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}