{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Brea",
      "state": "CA",
      "country": "United States",
      "classification": "Class III",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "63130",
      "recalling_firm": "Beckman Coulter Inc.",
      "address_1": "250 S Kraemer Blvd",
      "address_2": "N/A",
      "postal_code": "92821-6232",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution: USA (Nationwide) and Internationally to: Afghanistan Angola Australia Bangladesh Belgium Bhutan Brazil Brunei Darussalam Bulgaria Canada China Colombia Costa Rica Cote d'Ivoire Czech Republic Denmark Ecuador Egypt El Salvador Eritrea Ethiopia Finland France French Polynesia Germany Ghana Greece Guatemala Hong Kong India Indonesia Ireland Israel Italy Japan Jordan Kenya Kuwait Lebanon Lesotho Libyan Arab Jamahiriya Lithuania Macao Malawi Malaysia Mexico Morocco Namibia Netherlands New Zealand Nigeria Oman Pakistan Panama Paraguay Philippines Poland Puerto Rico Qatar Reunion Romania Russian Federation Rwanda Saudi Arabia Serbia Singapore Slovakia South Africa Spain Swaziland Sweden Switzerland Taiwan Thailand Tunisia Turkey Uganda United Arab Emirates United Kingdom Viet Nam Zambia Zimbabwe",
      "recall_number": "Z-0110-2013",
      "product_description": "Synchron Systems Synchron Calibrator Multi Calibrator Diskettes, Part Number 442600.    The SYNCHRON MULTI-CALIBRATOR, used in conjunction with SYNCHRON reagents, is intended for use on SYNCHRON Systems for the calibration of Albumin, Blood Urea Nitrogen (Urea), Calcium, Cholesterol, Glucose, Inorganic Phosphorus, Lactate,Magnesium, Total Protein, Triglycerides, and Uric Acid.",
      "product_quantity": "5746 units total (2729 units in US)",
      "reason_for_recall": "The recall was initiated because Beckman Coulter has confirmed a mislabeling error, some Synchron Systems Multi Calibrator Diskettes (REF 442600, lot M002642G) may actually be Thyroxine (T4) Reagent Calibrator Diskettes.",
      "recall_initiation_date": "20111215",
      "center_classification_date": "20121022",
      "termination_date": "20121228",
      "report_date": "20121031",
      "code_info": "Lot M002642"
    }
  ]
}