  "results": [
    {
      "country": "United States",
      "city": "Saint Paul",
      "address_1": "1 Saint Jude Medical Dr",
      "reason_for_recall": "St. Jude Medical is conducting a voluntary recall of SMJ Batch # 4641580 of the Fast-Cath Trio\" Hemostasis Introducer, Catalog number 406308. The inner diameter of the 10F dilator is too small, such that the guidewire cannot advance through the device which may result in a procedural delay while another dilator of the desired size is obtained.",
      "address_2": "",
      "product_quantity": "220",
      "code_info": "Batch 4641580",
      "center_classification_date": "20141021",
      "distribution_pattern": "Nationwide Distribution including FL, IN, ME, MA, NY, NY, OH, RI,TX, and WV.",
      "state": "MN",
      "product_description": "St Jude Medical, 10F Fast-Cath Trio\" Hemostasis Introducer, REF #  406308, Sterile EO.    The Fast-Cath Trio\" Hemostasis Introducer consists of the following components: Port hemostasis adapter (3), Sheath (10F Cath-Lock\"), dilator (10F) and 0.038\u001d guidewire.  See product labeling for diagram of contents.",
      "report_date": "20141029",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "St Jude Medical",
      "recall_number": "Z-0108-2015",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "69499",
      "termination_date": "20150325",
      "more_code_info": "",
      "recall_initiation_date": "20141014",
      "postal_code": "55117-1789",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}