{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Ballina Tipperary",
      "state": "N/A",
      "country": "Ireland",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "93017",
      "recalling_firm": "TECHNO-PATH MANUFACTURING LTD.",
      "address_1": "Fort Henry Business Park",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "US Nationwide distribution in the state of Illinois.",
      "recall_number": "Z-0106-2024",
      "product_description": "Multichem P, Part Code 08P90-10",
      "product_quantity": "2964 units",
      "reason_for_recall": "Technopath Manufacturing Ltd. has identified an issue with the glass vial commodity used to fill the Multichem S Plus Level 1, Multichem S Plus Level 2, Multichem S Plus Level 3, Multichem IA Plus and Multichem P Controls. Some customers have experienced broken vials when receiving or thawing Technopath Multichem IA Plus and Multichem S Plus Controls. Breakages have been observed upon delivery, while thawing or while opening the vial. The breakages can result in the bottom of the vial breaking off, breakage in the neck thread area or cracking on the sides of the vial. This may expose the user to biohazardous material if cut and may cause a delay in patient results.",
      "recall_initiation_date": "20230825",
      "center_classification_date": "20231012",
      "report_date": "20231018",
      "code_info": "UDI/DI 05391523440769, Lot Numbers:  04530621P, 04561021P, 04610523P"
    }
  ]
}