  "results": [
    {
      "status": "Terminated",
      "city": "Zurich",
      "state": "N/A",
      "country": "Switzerland",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80917",
      "recalling_firm": "Thoratec Switzerland GMBH",
      "address_1": "Technoparkstrasse 1",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "U.S.: OR, AZ, CA, ID, WA, WI, NY, MA, CT, NJ, ME, RI, IL, MI, TN, OH, IN, KY, IA, AL, FL, NC, GA, SC, VA, MD, DC, PA, WV, DE, MN, MT, NE, CO, NM, UT, AR, MO, MC, TX, OK, LA, MS, ND",
      "recall_number": "Z-0103-2019",
      "product_description": "Thoratec CentriMag Motor, Model: 102956, a component of the Thoratec CentriMag Blood Pumping System and PediMag Blood Pumping System.",
      "product_quantity": "1569",
      "reason_for_recall": "Reports of circulatory support system motor stopping. Motor stop can be attributed to a break of a wire or short circuit between wires within the electrical cable where the cable exits the motor cable bend protection.",
      "recall_initiation_date": "20180905",
      "center_classification_date": "20181012",
      "termination_date": "20210428",
      "report_date": "20181024",
      "code_info": "UDI: 07640135140078. All serial numbers until design change is implemented to make the motor cable more resistant to damage."
    }
  ]
}