      "country": "United States",
      "city": "Indianapolis",
      "address_1": "9115 Hague Rd",
      "reason_for_recall": "Roche has received four reports of injury caused by the sharp edge of the center guide rail of an  analyzer\u0019s sample rack tray. The cleaning procedure for the sample rack tray has been revised to clarify the cleaning process.",
      "address_2": "",
      "product_quantity": "9014 in total",
      "code_info": "Not applicable",
      "center_classification_date": "20161025",
      "distribution_pattern": "US Nationwide Distribution",
      "state": "IN",
      "product_description": "cobas p 701 Post-Analytical Unit  calculator/data processing module, for clinical use    Product Usage:  The URISYS 2400 Analyzer is a fully automated urinalysis system intended for in vitro qualitative or semi-quantitative determination of urine analytes, including pH, leukocytes, nitrite, protein, glucose, ketones, urobilinogen, bilirubin, and erythrocytes, In addition, the analyzer determines specific gravity, color, and clarity. The URISYS 2400 is intended for professional use only.",
      "report_date": "20161102",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Roche Diagnostics Operations, Inc.",
      "recall_number": "Z-0101-2017",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "75186",
      "termination_date": "20170605",
      "more_code_info": "",
      "recall_initiation_date": "20160715",
      "postal_code": "46256-1025",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}