  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Dallas",
      "address_1": "4144 N Central Expy Ste 260",
      "reason_for_recall": "Estill has discovered that a fully discharged Ultra Battery 1 (TA-UB1) will reach 96.9% of its capacity during a 60 minute charging period.  If you have a fully discharged Ultra Battery 1 (TA-UB1), in order to achieve a 100% of charge capacity you should charge it for a period of 90 minutes.  In addition, the charging voltage range of 10-36V specified on the Ultra Battery 1 (TA-UB1) label and in t",
      "address_2": "",
      "product_quantity": "6,886",
      "code_info": "06-2011; 07-2011; 08-2011; 09-2011; 03-2012; 04-2012; 05-2012; 06-2012; 07-2012; 08-2012; 01-2013; 03-2013; 08-2013; 09-2013; 10-2013; 11-2013; 12-2013; 01-2014; 02-2014; 04-2014; 07-2012; 02-2014",
      "center_classification_date": "20141021",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) and Internationally to Serbia.",
      "state": "TX",
      "product_description": "Ultra Battery 1 Model: TA-UB1 NSN: 6515-01-592-5803     The Ultra Battery 1 (TA-UB1) is a rechargeable power source for the Thermal Angel Blood and IV Fluid Infusion Warmer. The TA-UB1 battery was designed to provide sufficient power for the Thermal Angel to warm and deliver one liter of blood or IV fluid infusion to a trauma patient.",
      "report_date": "20141029",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Estill Medical Technologies, Inc",
      "recall_number": "Z-0101-2015",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "69242",
      "termination_date": "20150821",
      "more_code_info": "",
      "recall_initiation_date": "20140725",
      "postal_code": "75204-3100",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}