  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Trumbull",
      "state": "CT",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86479",
      "recalling_firm": "CooperSurgical, Inc.",
      "address_1": "95 Corporate Dr",
      "address_2": "N/A",
      "postal_code": "06611-1350",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "The products were distributed to China.",
      "recall_number": "Z-0100-2021",
      "product_description": "MediCult Vitrification Cooling Media, Model Number 12284001F",
      "product_quantity": "531 total boxes",
      "reason_for_recall": "There is a discrepancy between the printing on the Product Tyvek label of the individual component packaging. The Chinese characters on the label indicates that it contains an Equilibration Medium (P/N: 12220001) rather than the Vitrification Medium (P/N: 12230001) as indicated on the remainder of the label.",
      "recall_initiation_date": "20200908",
      "center_classification_date": "20201013",
      "termination_date": "20210111",
      "report_date": "20201021",
      "code_info": "Lot: 20230080"
    }
  ]
}