  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Waltham",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86410",
      "recalling_firm": "Fresenius Medical Care Holdings, Inc.",
      "address_1": "920 Winter St Bld 950",
      "address_2": "N/A",
      "postal_code": "02451-1521",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US distribution to DE, NC, OK, PA, and TX.",
      "recall_number": "Z-0098-2021",
      "product_description": "AquaBplus and AquaBplus B2 Water Purification System.  For use with a hemodialysis system to remove organic and inorganic substances and microbial contaminants from the water used for hemodialysis or related therapies.    AquaBplus 1500 - Typ USA - 3 x 208V / 60Hz -Fresenius Part Number: 24-1500-1  AquaBplus B2 2000 / B1 2500 (Bplus DUO 2000) - 3 x 208V / 60Hz- Fresenius Part  Number: 24-2000-0",
      "product_quantity": "15 units",
      "reason_for_recall": "Incorrectly crimped connectors were installed on the main protective earth conductor on these machines. If the machine is moved or maintenance is performed, these crimp connections could become loose and may adversely affect the electrical safety of the  products, potentially resulting in electrical shock to the Biomed Technician or clinic staff.",
      "recall_initiation_date": "20200901",
      "center_classification_date": "20201013",
      "termination_date": "20210120",
      "report_date": "20201021",
      "code_info": "9BDS1515      08D51544      OBDS1596      OBPS2864  9BDS1529      OBDS1545      OBPS2799      OBPS2884  9BPS2735      OBDS1551      OBPS2800      OBPS2903  9BPS2767      08D51590     OBPS2810"
    }
  ]
}