  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Indianapolis",
      "address_1": "9115 Hague Rd",
      "reason_for_recall": "Roche has received four reports of injury caused by the sharp edge of the center guide rail of an  analyzer\u0019s sample rack tray. The cleaning procedure for the sample rack tray has been revised to clarify the cleaning process.",
      "address_2": "",
      "product_quantity": "9014 in total",
      "code_info": "Not applicable",
      "center_classification_date": "20161025",
      "distribution_pattern": "US Nationwide Distribution",
      "state": "IN",
      "product_description": "Modular Pre Analytics ( EVO/Plus)    Product Usage:  The RSD 800A is a computer-controlled stand-online system for decapping, aliquoting and sorting of primary tubes and for producing and sorting bar coded secondary tubes.. The system operates with an online connection to the host.",
      "report_date": "20161102",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Roche Diagnostics Operations, Inc.",
      "recall_number": "Z-0098-2017",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "75186",
      "termination_date": "20170605",
      "more_code_info": "",
      "recall_initiation_date": "20160715",
      "postal_code": "46256-1025",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}