{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Raynham",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "65897",
      "recalling_firm": "DePuy Spine, Inc.",
      "address_1": "325 Paramount Dr",
      "address_2": "N/A",
      "postal_code": "02767-5199",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide Distribution including the states of AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MT, NC, ND, NE, NH, NJ, NY, OH, OK, OR, PA, SC, SD, TN, TX, VA, WA, WI, WV, and WY.",
      "recall_number": "Z-0097-2014",
      "product_description": "PIPELINE LS ACCESS SYSTEM, Holding Clip    The PIPELINE LS Dilators and Holding Clip are intended for sequential dilation of soft tissue and enabling proper placement of the Pipeline LS Retractor.",
      "product_quantity": "198 units",
      "reason_for_recall": "Dilators and Holding Clip are being used for a secondary function of neuromonitoring, for which it is not 510(k) cleared.",
      "recall_initiation_date": "20130731",
      "center_classification_date": "20131028",
      "termination_date": "20150629",
      "report_date": "20131106",
      "code_info": "287105090"
    }
  ]
}