  "results": [
    {
      "status": "Terminated",
      "city": "Lake Forest",
      "state": "IL",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "68047",
      "recalling_firm": "Hospira Inc.",
      "address_1": "275 N Field Dr",
      "address_2": "N/A",
      "postal_code": "60045-2579",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution -- USA, including the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY; and, the countries of Austria, Belgium, Brazil, Columbia, France, Germany, Greece, Netherlands, Saudi Arabia, South Africa, Spain, Sweden, Switzerland, Turkey, and United Kingdom.",
      "recall_number": "Z-0096-2015",
      "product_description": "The GemStar Power Supply is an accessory for the GemStar Infusion Pump. It is a two piece assembled switching power supply. The Power Supply converts alternating current (AC) line input voltage of 120 or 240 V to an output to the device of 3.3 Volts DC.",
      "product_quantity": "US Distribution: 4,709 power supplies; Foreign Distribution: 7,425 power supplies",
      "reason_for_recall": "The GemStar Infusion pump may not receive direct current (DC) power from the power supply. In one instance, smoke was observed and the GemStar pump was operating on battery power while connected to the 3 volt DC power supply.",
      "recall_initiation_date": "20140203",
      "center_classification_date": "20141024",
      "termination_date": "20170307",
      "report_date": "20141105",
      "code_info": "Desktop GemStar Power Supply List Number: 13072-05; 1) US Distribution Lot Numbers: 10115G1, 11121G1, 11122G1, 1123G1, 11124G1, 12143G1, 12144G1, 12178G1, 15185G1, 12195G1, 17214G1, 18217G1, 18219G1, 19225G1, 20233G1, 21242G1, 22251G1, 23259G1, 24271G1, 25280G1, 26289G1, 26290G1, 27293G1, 27297G1, 27300G1; 2) Foreign Distribution Lot Numbers: 11128G1, 12140G1, 12142G1, 12148G1, 14176G1, 15187G1, 15196G1, 16207G1, 16208G1, 18218G1, 19226G1, 19229G1, 20234G1, 21241G1, 22255G1, 23262G1, 24270G1, 25276G1, 26285G1, 27299G1, 27301G1",
      "more_code_info": ""
    }
  ]
}