      "status": "Terminated",
      "city": "Rochester",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "74950",
      "recalling_firm": "Ortho-Clinical Diagnostics",
      "address_1": "100 Indigo Creek Dr",
      "address_2": "N/A",
      "postal_code": "14626-5101",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide Distribution - US (Nationwide) Puerto Rico and Bermuda), Internationally to Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy,  Japan, Mexico, Republic of Panama, Poland, Singapore, Spain, and Venezuela.",
      "recall_number": "Z-0092-2017",
      "product_description": "VITROS Chemistry Products URIC Slides, REF/Product Code 194 3927 (300 slides per sales unit),  Unique Device Identifier No. 10758750000340, Rx ONLY, IVD ---   For in vitro diagnostic use only.",
      "product_quantity": "US: 18,749 sales units; Foreign: 94,928 sales units",
      "reason_for_recall": "Ortho Clinical Diagnostics confirmed that N-Acetylcysteine (NAC), a drug used for acetaminophen overdose, mucolytic therapy and is used as nutritional supplement, can interfere with VITROS GLU, LAC, TRIG, and URIC Slides assays.",
      "recall_initiation_date": "20160811",
      "center_classification_date": "20161017",
      "termination_date": "20180611",
      "report_date": "20161026",
      "code_info": "Current/In-date GENs 26, 27, 28, 29, 31"
    }
  ]
}