  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Brea",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "78256",
      "recalling_firm": "Beckman Coulter Inc.",
      "address_1": "250 S Kraemer Blvd",
      "address_2": "N/A",
      "postal_code": "92821-6232",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Distribution to the states of :  AZ, CA, CO, FL, GA, IL, IN, KS, NC, NY, OR, PA, SC, TX, and WA, including  Puerto Rico and government distribution, and to the countries of : Canada, Angola, Australia, Belgium, Botswana, Brazil, Chile, Colombia, Czech Republic, Denmark, Finland, France, Germany, Ghana, Hong Kong, India, Israel, Italy, Kenya, Lesotho, Malawi, Malaysia, Mexico, Namibia, Netherlands, New Zealand, Norway, Pakistan, Panama, Poland, Portugal, Romania, Slovakia, South Africa, Spain, Sweden, Switzerland, Turkey, Uganda, Ukraine, United Kingdom, and Zambia.",
      "recall_number": "Z-0091-2018",
      "product_description": "AQUIOS CL Flow Cytometer, Catalog No. B30166, Rx.   The firm name on the label is Beckman Coulter Ireland Inc., Ireland.    The AQUIOS CL Flow Cytometer is intended for use as an in vitro diagnostic device for immunophenotyping. It is used in conjunction with the following reagents and software package: AQUIOS Tetra-1 Panel monoclonal antibody reagents, AQUIOS Tetra-2+ Panel monoclonal antibody reagents, and AQUIOS CL Flow Cytometry Software.",
      "product_quantity": "268 units",
      "reason_for_recall": "The system may duplicate sample requests leading to sample misidentification when connected to a Laboratory Information System (LIS)",
      "recall_initiation_date": "20171002",
      "center_classification_date": "20171115",
      "termination_date": "20190508",
      "report_date": "20171122",
      "code_info": "All existing software versions (2.0, 2.0.1, and 2.1)."
    }
  ]
}