{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Rochester",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "74950",
      "recalling_firm": "Ortho-Clinical Diagnostics",
      "address_1": "100 Indigo Creek Dr",
      "address_2": "N/A",
      "postal_code": "14626-5101",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide Distribution - US (Nationwide) Puerto Rico and Bermuda), Internationally to Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy,  Japan, Mexico, Republic of Panama, Poland, Singapore, Spain, and Venezuela.",
      "recall_number": "Z-0091-2017",
      "product_description": "VITROS Chemistry Products TRIG Slides, REF/Product Codes 133 6544 (300 slides per sales unit) and 832 9930 (90 slides per sales unit), Unique Device Identifier Nos. 10758750008889 and 10758750004768, Rx ONLY, IVD ---   For in vitro diagnostic use only. VITROS Chemistry Products TRIG Slides quantitatively measure triglyceride (TRIG) concentration in serum and plasma using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System. Triglyceride measurements are used in the diagnosis and treatment of patients with diabetes mellitus, nephrosis, liver obstruction, other diseases involving lipid metabolism, or various endocrine disorders",
      "product_quantity": "US: 60,372 sales units; Foreign: 145,611 sales units",
      "reason_for_recall": "Ortho Clinical Diagnostics confirmed that N-Acetylcysteine (NAC), a drug used for acetaminophen overdose, mucolytic therapy and is used as nutritional supplement, can interfere with VITROS GLU, LAC, TRIG, and URIC Slides assays.",
      "recall_initiation_date": "20160811",
      "center_classification_date": "20161017",
      "termination_date": "20180611",
      "report_date": "20161026",
      "code_info": "Current/In-date GENs 16, 17, 19, 20, 22, 23, 24, 25, 27 through 32"
    }
  ]
}