  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Rochester",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "74950",
      "recalling_firm": "Ortho-Clinical Diagnostics",
      "address_1": "100 Indigo Creek Dr",
      "address_2": "N/A",
      "postal_code": "14626-5101",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide Distribution - US (Nationwide) Puerto Rico and Bermuda), Internationally to Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy,  Japan, Mexico, Republic of Panama, Poland, Singapore, Spain, and Venezuela.",
      "recall_number": "Z-0089-2017",
      "product_description": "VITROS Chemistry Products GLU Slides, REF/Product Code  170 7801 (300 slides per sales unit), Unique Device Identifier No. 10758750009572, Rx ONLY, IVD ---    For in vitro diagnostic use only. VITROS Chemistry Products GLU Slides quantitatively measure glucose (GLU) concentration in serum, plasma, urine, and cerebrospinal fluid using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System",
      "product_quantity": "US: 192,459 sales units; Foreign: 231,710 sales units",
      "reason_for_recall": "Ortho Clinical Diagnostics confirmed that N-Acetylcysteine (NAC), a drug used for acetaminophen overdose, mucolytic therapy and is used as nutritional supplement, can interfere with VITROS GLU, LAC, TRIG, and URIC Slides assays.  Ortho also tested several compounds following industry standard CLSI EP7-A2 for potential interference on VITROS GLU Slides during validation testing for a proposed change and confirmed Dextran 40 and Glutathione may interfere with current VITROS GLU Slides.",
      "recall_initiation_date": "20160811",
      "center_classification_date": "20161017",
      "termination_date": "20180611",
      "report_date": "20161026",
      "code_info": "Current/In-date GENs 10, 11, 20 through 31"
    }
  ]
}