{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Eatontown",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "62569",
      "recalling_firm": "Osteotech Inc",
      "address_1": "51 James Way",
      "address_2": "N/A",
      "postal_code": "07724-2272",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) and Internationally.",
      "recall_number": "Z-0089-2013",
      "product_description": "Grafton DBM Matrix Strips  Multiple Sizes    Grafton is a human bone allograft product containing demineralized human bone tissue combined with inert additive to yield a demineralized bone matrix (DBM) allograft. It is intended for use as a bone graft extender, bone graft substitute and bone void filler in bony voids or gaps of the skeletal system (ie spine, pelvis and extremeties) not intrinsic to the stability of the bony structure. The voids or gaps may be surgically created defects or effects created by traumatic injury to the bone.",
      "product_quantity": "4,117 units",
      "reason_for_recall": "Multiple lots of GRAFTON ¿ and XPANSE ¿, which are used for bone  grafting procedures are being recalled because there is a possibility that sterility of the outer surface of the inner pouch may become  compromised.",
      "recall_initiation_date": "20120420",
      "center_classification_date": "20121018",
      "termination_date": "20140117",
      "report_date": "20121024",
      "code_info": "Multiple Lots - please see attachment"
    }
  ]
}