  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Lugano",
      "state": "N/A",
      "country": "Switzerland",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "88700",
      "recalling_firm": "Inpeco S.A.",
      "address_1": "Via San Gottardo 10",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "N/A",
      "distribution_pattern": "Domestic distribution nationwide . Foreign distribution worldwide.",
      "recall_number": "Z-0087-2022",
      "product_description": "FlexLab (FLX), Accelerator a3600 (ACP), Aptio Automation (AP2). (Different brand names of same product.)",
      "product_quantity": "262 total, 52 US",
      "reason_for_recall": "The Firmware (FW) of the Automation System Interface Module to ADVIA Centaur XP/XPT may lead to an incorrect association of test results to sample ID. Improper management of the sample may bring an incorrect patient result.",
      "recall_initiation_date": "20210910",
      "center_classification_date": "20211012",
      "report_date": "20211020",
      "code_info": "Impacted Module: ADVIA Centaur XP/XPT Interface Module, Part No. FLX-021-00; FLX-021-10; APT-021-00    Serial numbers:  ***FLEXLAB***  FLX.0109  FLX.0127  FLX.0128  FLX.0129  FLX.0132  FLX.0133  FLX.0139  FLX.0146  FLX.0147  FLX.0149  FLX.0151  FLX.0153  FLX.0154  FLX.0156  FLX.0157  FLX.0159  FLX.0160  FLX.0162  FLX.0165  FLX.0166  FLX.0167  FLX.0169  FLX.0170  FLX.0172  FLX.0173  FLX.0174  FLX.0179  FLX.0180  FLX.0182  FLX.0191  FLX.0192  FLX.0193  FLX.0194  FLX.043  FLX.069  FLX.085  FLX.086  FLX.087  FLX.093  FLX.097    ***ACCELERATOR a3600***  ACP.334      ***APTIO AUTOMATION***  AP2.0002  AP2.0003  AP2.0004  AP2.0005  AP2.0006  AP2.0007  AP2.0008  AP2.009  AP2.0010  AP2.0011  AP2.0012  AP2.0014  AP2.0015  AP2.0016  AP2.0017  AP2.0018  AP2.0019  AP2.0022  AP2.0023  AP2.0024  AP2.0025  AP2.0026  AP2.0027  AP2.0028  AP2.0029  AP2.0031  AP2.0032  AP2.0033  AP2.0035  AP2.0037  AP2.0038  AP2.0039  AP2.0040  AP2.0041  AP2.0043  AP2.0045  AP2.0046  AP2.0047  AP2.0048  AP2.0049  AP2.0050  AP2.0051  AP2.0052  AP2.0053  AP2.0054  AP2.0055  AP2.0057  AP2.0058  AP2.0059  AP2.0061  AP2.0062  AP2.0063  AP2.0064  AP2.0065  AP2.0066  AP2.0067  AP2.0070  AP2.0075  AP2.0076  AP2.0077  AP2.0078  AP2.0079  AP2.0080  AP2.0081  AP2.0082  AP2.0083  AP2.0084  AP2.0085  AP2.0087  AP2.0088  AP2.0089  AP2.0090  AP2.0091  AP2.0092  AP2.0093  AP2.0094  AP2.0095  AP2.0097  AP2.0098  AP2.0100  AP2.0102  AP2.0103  AP2.0104  AP2.0105  AP2.0106  AP2.0107  AP2.0108  AP2.0109  AP2.0110  AP2.0111  AP2.0112  AP2.0115  AP2.0117  AP2.0118  AP2.0119  AP2.0122  AP2.0123  AP2.0124  AP2.0126  AP2.0127  AP2.0128  AP2.0129  AP2.0131  AP2.0132  AP2.0133  AP2.0135  AP2.0136  AP2.0137  AP2.0138  AP2.0141  AP2.0142  AP2.0144  AP2.0145  AP2.0147  AP2.0149  AP2.0150  AP2.0153  AP2.0155  AP2.0156  AP2.0160  AP2.0161  AP2.0162  AP2.0165  AP2.0166  AP2.0168  AP2.0173  AP2.0178  AP2.0179  AP2.0180  AP2.0182  AP2.0186  AP2.0190  AP2.0191  AP2.0192  AP2.0193  AP2.0195  AP2.0198  AP2.0199  AP2.0200  AP2.0204  AP2.0205  AP2.0206  AP2.0207  AP2.0208  AP2.0212  AP2.0216  AP2.0219  AP2.0221  AP2.0222  AP2.0223  AP2.0226  AP2.0233  AP2.0235  AP2.0236  AP2.0237  AP2.0238  AP2.0239  AP2.0243  AP2.0244  AP2.0245  AP2.0247  AP2.0249  AP2.0251  AP2.0252  AP2.0253  AP2.0257  AP2.0267  AP2.0269  AP2.0271  AP2.0272  AP2.0273  AP2.0279  AP2.0280  AP2.0281  AP2.0282  AP2.0284  AP2.0287  AP2.0288  AP2.0301  AP2.0305  AP2.0307  AP2.0309  AP2.0312  AP2.0315  AP2.0327  AP2.0333  AP2.0337  AP2.0341  AP2.0342  AP2.0343  AP2.0345  AP2.0351  AP2.0355  AP2.0357  AP2.0358  AP2.0370  AP2.0371  AP2.0399  AP2.0407  AP2.0409  AP2.0426  AP2.0457  AP2.0474  AP2.0475  AP2.0476  AP2.0486  AP2.0491  AP2.0494  AP2.0498  AP2.0503  AP2.0506  AP2.0509  AP2.0512  AP2.0554  AP2.0563  APT.0009C  APT.0058C  APT.0065C  APT.0093C  APT.0125C  APT.0091C"
    }
  ]
}