      "country": "United States",
      "city": "Johnstown",
      "address_1": "141 Sal Landrio Dr",
      "reason_for_recall": "There is a possibility that the needles may have the potential to skive a catheter\u0019s outer coating.  If this were to occur, there exists the risk that a portion of the catheter could remain in the patient, which may require otherwise unnecessary medical invention.",
      "address_2": "Crossroads Business Park",
      "product_quantity": "576 units",
      "code_info": "1) Spectra Lot# GC091702 (Epimed manufacturing lot # 2801501, Sterile Lot # 12157181)  2) Spectra Lot# GC101871 (Epimed manufacturing lot # 3211503, Sterile Lot # 12157445)  3) Spectra Lot# GC101897 (Epimed manufacturing lot # 3161501, Sterile Lot # 12157325)",
      "center_classification_date": "20161031",
      "distribution_pattern": "US Distribution to the states of : AL, AK, FL, IN, TN & TX",
      "state": "NY",
      "product_description": "15 g R.K. Epidural Needle (TW), Catalog #/REF #100-1415   Intended for the epidural placement, directly or through an epidural catheter, of anesthetic agents to elicit regional anesthesia",
      "report_date": "20161109",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Epimed International, Inc.",
      "recall_number": "Z-0086-2017",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "74018",
      "termination_date": "20180614",
      "more_code_info": "",
      "recall_initiation_date": "20160422",
      "postal_code": "12095-3835",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}