  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Utica",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "75056",
      "recalling_firm": "ConMed Corporation",
      "address_1": "525 French Rd",
      "address_2": "N/A",
      "postal_code": "13502-5945",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide in the states of:  CA, HI, IL, IN, KS, LA, MA, NC, NM, NY, OH, OK, PA, TN, TX, UT, WA and WI; plus Foreign distribution to Australia, Canada, France, Germany, Italy, Spain, Czech Republic, Lithuania, Netherlands, Poland, Korea,Morocco, Malaysia, Mexico, Taiwan, and South Africa.",
      "recall_number": "Z-0085-2017",
      "product_description": "CONCEPT HEATWAVE Electrode 130 mm x 2.3 mm diameter, REF/Catalog Number CS-023, STERILE, Rx Only   Product Usage:  The HeatWave Electrode is intended to be used for electro-coagulation of soft tissue in shoulder, ankle, wrist, elbow and knee arthroscopic procedures.",
      "product_quantity": "Domestic: 1740; Foreign: 4770",
      "reason_for_recall": "CONMED received complaints of damage to the CS-023 electrode\u0019s shaft insulation which could result in burn marks on the shaft.  If the shaft insulation is damaged in any way, it could increase the risk of patient burn.",
      "recall_initiation_date": "20160829",
      "center_classification_date": "20161013",
      "termination_date": "20180607",
      "report_date": "20161019",
      "code_info": "All lots codes manufactured from 09/01/2011 through 11/17/2015:  267221, 301001, 309889, 318652, 319649, 325019, 354759, 334627, 336041, 341704, 343927, 349952, 384294, 354768, 365239, 369773, 374480, 378521, 406137, 385695, 388538, 392382, 394178, 396542, 436808, 413702, 456341, 423877, 427704, 433503, 478513, 451923, 488101, 464062, 468012, 471297, 511327, 482366, 522335, 494840, 499981, 505321, 544240, 515060, 551152, 527734, 531802, 536700, 581072, 550988, 590037, 557461, 570449, 574167, 611804, 588892, 621304, 597982, 598682, 605515, 643396, 619686, 656909, 629202, 634911, 643395, 676005, 654991, 692617, 667118, 671471, 674410, 683144, 695062, 695063, 297982."
    }
  ]
}