{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "75234",
      "recalling_firm": "Zimmer Biomet, Inc.",
      "address_1": "56 E Bell Dr",
      "address_2": "N/A",
      "postal_code": "46582-6989",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution in the states of: MI, UT, SD, OH, WA, KS, MO, NJ, GA, and TX.",
      "recall_number": "Z-0083-2017",
      "product_description": "Phoenix Recon Soft Tissue Sleeve  Guide, Surgical, Instrument for the sleeve and Orthopedic Manual Surgical Instruments for the Arm    Product Usage:  The Phoenix Femoral Nail System is to be implanted into the femur for alignment, stabilization and fixation of fractures caused by trauma or disease, and the fixation of femurs that have been surgically prepared (osteotomy) for correction of deformity, and for arthrodesis. These instruments are used in support of the surgery.",
      "product_quantity": "31",
      "reason_for_recall": "Old revisions and new revisions of the soft tissue sleeve and recon targeting arm are not interchangeable, as the new revision tissue sleeve will not fit into the old revision targeting arm. The old revision tissue sleeve has too much clearance with new revision targeting arm.  Risks of encountering the affected product include: the correct instrument may not be readily available for the procedure causing the surgeon to complete the procedure by hand; or a delay in surgery greater than 30 minutes may occur while another Recon Targeting Arm or Recon Soft tissue Sleeve is located or while the surgeon completes the procedure by hand.",
      "recall_initiation_date": "20160809",
      "center_classification_date": "20161013",
      "termination_date": "20170323",
      "report_date": "20161019",
      "code_info": "Item number: 14-442008  Lot number: 2095181"
    }
  ]
}