      "status": "Terminated",
      "city": "Wayne",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "66336",
      "recalling_firm": "Maquet Medical Systems USA",
      "address_1": "45 Barbour Pond Dr",
      "address_2": "N/A",
      "postal_code": "07470-2094",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide Distribution including AZ, FL, CO, TX, OH, VA, MN, WI, SC, IL, ND, UT, CA, OR, NY, IA, CT, SD, MI, HI, AL, MD, KS, MO",
      "recall_number": "Z-0082-2014",
      "product_description": "MAQUET Getinge Group CARDIOHELP Base Unit REF 70104.7999 REF 70104.8012 Manufactured for:  MAQUET Cardiopulmonary AG Made in Germany.    Cardiopulmonary support system.",
      "product_quantity": "63",
      "reason_for_recall": "It has come to the attention of MAQUET that in certain rare instances, the CARDIOHELP-i device may display an error message with audible alarm, indicating \"battery defective.\"  The alarm will occur-upon start up or when the main power supply is unplugged from the main power source.  The human machine interface touch screen of the CARDIOHELP-i device may gradually become black due to the human machine interface (HMI) back light shutting down.",
      "recall_initiation_date": "20130912",
      "center_classification_date": "20131025",
      "termination_date": "20170607",
      "report_date": "20131106",
      "code_info": "Part No. 70104.8012 Serial Nos. 90410440; 90410452; 90410455; 90410456; 90410457; 90410459; 90410469; 90410474; 90410480; 904104781; 90410482; 90410484; 90410498; 90410499; 90410500; 90410501; 90410502; 90410503; 90410521; 90410522; 90410524; 90410526; 90410527; 90410531; 90410533;   90410534; 90410535; 90410550; 90410551; 90410552; 90410568; 90410571; 90410572; 90410573; 90410575; 90410576; 90410577; 90410580; 90410581; 90410582; 90410584; 90410585; 90410592; 90410596; 90410599; 90410600; 90410605; 90410606; 90410607; 90410608; 90410609; 90410610; 90410611; 90410612; 90410626; 90410638; 90410640; 90410642; 90410644; 90410646; 90410647; 90410650; 90410651; 90410652; 90410653; 90410656; 90410659; 90410660; 90410661; 90410662; 90410663; 90410664; 90410665; 90410666; 90410667; 90410668; 90410670; 90410700; 90410708; 90410846; 90410750; 90410786"
    }
  ]
}