{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-10-14",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Waukegan",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "99801",
      "recalling_firm": "Cardinal Health 200, LLC",
      "address_1": "3651 Birchwood Dr",
      "address_2": "N/A",
      "postal_code": "60085-8337",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Pending",
      "recall_number": "Z-0079-2027",
      "product_description": "Presource Packs containing Medtronic Microstream Components:    Intubation Kit  Catalog Numbers   AN23366084  AN23366085  AN2336608H  AN2336608J    Intubating Bag Kit  Catalog Number OMSVM4166A    Intubation UH Kit  Catalog Number AN2336608G",
      "product_quantity": "210 total kits",
      "reason_for_recall": "Kits contain Microstream components. Medtronic issued an update to the Instructions for Use of these components due reports of difficulty in disconnecting the adapter from the patient's endotracheal tube to perform a procedure. The issue can result in a delay to treatment and unintended extubation. The potential exists for respiratory failure, hypoxia, aspiration, or low oxygen saturation.",
      "recall_initiation_date": "20260819",
      "center_classification_date": "20261002",
      "report_date": "20261014",
      "code_info": "AN23366084    UDI-DI 10197106950074    Kit lot 40711  AN23366085    UDI-DI 10198956177987    Kit lot 46520  AN2336608H    UDI-DI 10197106159828    Kit lots                 20866   23139   24264   25795   27111   27899   29023   29488   30527   31037   31698   32777   33326  AN2336608J    UDI-DI 10197106704035    Kit lots                 33461   33624   35231   35535   35689   36554   36600   36825   38327   38891   40511   41864   43975   44933   45972   47547   47897   48088   48528   49440    OMSVM4166A    UDI-DI 10198956162259    Kit Lots                 47765   50521   53221   53929   54565   54644    AN2336608G    UDI-DI 10197106052105    Kit lots                 18464   18569   19224"
    }
  ]
}