  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Research Triangle Park",
      "state": "NC",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "63365",
      "recalling_firm": "Teleflex Medical",
      "address_1": "2917 Weck Dr.",
      "address_2": "N/A",
      "postal_code": "27709",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide Distribution including the state of Nevada.",
      "recall_number": "Z-0078-2013",
      "product_description": "Hem-o-lok Large Polymer Ligating Clips, REF #544240, Rx only, Sterile, Teleflex Medical, Research Triangle Park, NC 27709.    Teleflex Medical\u0019s Hem-o-lok polymer ligating clips are single use, sterile, non-absorbable medial devices that have been designed to ligate vessels and tissue structures during various types of surgical procedures. Vessels and tissue structures in the range of 0.5 to 16 mm can be effectively ligated with Hem-o-lok clips.",
      "product_quantity": "126 units",
      "reason_for_recall": "Product is being recalled due to the possibility that the tray may contain pinholes, compromising product sterility.",
      "recall_initiation_date": "20120925",
      "center_classification_date": "20121018",
      "termination_date": "20121108",
      "report_date": "20121024",
      "code_info": "Catalog No. 544240, Lot #01E1200585"
    }
  ]
}