  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Palo Alto",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "75296",
      "recalling_firm": "Varian Medical Systems, Inc.",
      "address_1": "911 Hansen Way",
      "address_2": "N/A",
      "postal_code": "94304-1028",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Distribution and Internationally to 1 site in Canada, India and Morocco",
      "recall_number": "Z-0077-2017",
      "product_description": "Varian Head Frame - Model number HHF -ALL Head Frame Posts (PN1008016 ) and Post Kits (PN1008016).      The Varian Head Frame System is for use with a computed tomography scanner to perform imaging for treatment planning and a charged particle accelerator to perform immobilization of the treatment target for stereotactic radiosurgery or radiotherapy treatments on cranial lesions, tumors and conditions where radiation treatment is indicated.",
      "product_quantity": "75 devices",
      "reason_for_recall": "Varian Medical Systems has received a report that a user was able to easily rotate the head frame posts when attached to the mounting cam on the metal head ring. There was no report of serious injury due to this issue",
      "recall_initiation_date": "20160913",
      "center_classification_date": "20161013",
      "termination_date": "20170224",
      "report_date": "20161019",
      "code_info": "All Head Frame Posts and Post kits"
    }
  ]
}