{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "San Jose",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "83603",
      "recalling_firm": "Becton, Dickinson and Company, BD Biosciences",
      "address_1": "2350 Qume Dr",
      "address_2": "N/A",
      "postal_code": "95131-1812",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "U.S.: CA, KS, NJ, MI, NY, PA, IL, IN, VA, WA, MD, IA, OK, RI, FL, CO, OH, TX, TN, GA, MO, MN, HI, VT, MA, NE.    O.U.S. (Foreign): Germany, Italy, South Korea, Japan, United Kingdom, Poland, France, Belgium, Canada, Latvia, India, Taiwan, Ireland, Algeria, Israel, Iraq, Netherlands, Malaysia, Australia, Sweden, Colombia, Switzerland, Slovakia, Spain, Brazil, Austria, Peru, Iran, Croatia, Lithuania, Czech Republic, Saudi Arabia, Finland, Qatar, South Africa, Denmark, Turkey, Mexico, Indonesia, Ecuador, Vietnam, Singapore",
      "recall_number": "Z-0074-2020",
      "product_description": "BD FACSLyric Flow Cytometer, 3-Laser 12 Color Instrument, REF: 663029, when using BD Trucount Tubes",
      "product_quantity": "22",
      "reason_for_recall": "Flow cytometers, when used with specific tubes may experience excessive electronic abort counts (electronic abort count of beads exceeds 1.0% of the total processed events), which may affect the ratio of cell population events to bead events, potentially resulting in falsely high absolute counts. This may affect the following: initiation of prophylactic therapy for Opportunistic Infections (HIV patients), decisions regarding care/management (PID/immune suppressed patients), laboratory developed tests/experiments, specimen engraftment potency, and use of leucoreduced blood products for transfusion.",
      "recall_initiation_date": "20190819",
      "center_classification_date": "20191009",
      "termination_date": "20240412",
      "report_date": "20191016",
      "code_info": "BD FACSLyric Flow Cytometer serial numbers:  R663029000001,  R663029000002,  R663029000003,  R663029000004,  R663029000005,  R663029000006,  R663029000007,  R663029000008,  R663029000010,  R663029000011,  R663029000012,  R663029000014,  R663029000015,  R663029000016,  R663029000017,  R663029000018,  R663029000021,  R663029000025,  R663029000031,  R659180000136,  R659180000160,  R659180000238."
    }
  ]
}