{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Grand Rapids",
      "state": "MI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86474",
      "recalling_firm": "Skytron, Div. The KMW Group, Inc",
      "address_1": "5085 Corporate Exchange Blvd SE",
      "address_2": "N/A",
      "postal_code": "49512-5515",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "AZ,  MI, MS, NJ, PA,",
      "recall_number": "Z-0071-2021",
      "product_description": "VGDF-SKY (R4-010-06) VGA TO FIBER CONVERTER-Single Format Faceplate, a component of the SkyVision Linx 300 Integration Control System  Model: VGDF-SKY (version 2).  For communication and visual image management in the operating room suite.",
      "product_quantity": "11 units",
      "reason_for_recall": "VGDF (VGA) Faceplate failure results in failure of video signal for SkyVision Linx 300, a secondary display. Video would continue on the primary source device display",
      "recall_initiation_date": "20200911",
      "center_classification_date": "20201008",
      "termination_date": "20240924",
      "report_date": "20201014",
      "code_info": "Serial Numbers:  MXDM200501769  MXDM200501773  MXDM191100421  MXDM191100422  MXDM200200161  MXDM200200162  MXDM200200163  MXDM200200164  MXDM200501770  MXDM200501772  MXDM200501771"
    }
  ]
}