  "results": [
    {
      "status": "Terminated",
      "city": "Salt Lake City",
      "state": "UT",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "75330",
      "recalling_firm": "GE OEC Medical Systems, Inc",
      "address_1": "384 N Wright Brothers Dr",
      "address_2": "N/A",
      "postal_code": "84116-2862",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide  in the states of AL, AZ, CA, FL, KS, MS, NJ, NV, NY, PA, TX, UT, and WV. No Candaian or Va/govt/military consignees. There are 133 other foreign consignees.",
      "recall_number": "Z-0071-2017",
      "product_description": "OEC Brivo 715 Prime, OEC Brivo 785 Essential, OEC Brivo 865 Advance    Product Usage:  The OEC Brivo Mobile C-Arm X-Ray Products are designed to provide digital spot film imaging and fluoroscopic image guidance for all adult and pediatric populations for orientations between patient anatomy and surgical instruments. The product is used for general surgical applications and musculoskeletal procedures to visualize, for example, implant localizations or needle positions for aspirations, injections or biopsy. The OEC Brivo is not indicated for interventional use.",
      "product_quantity": "148 total",
      "reason_for_recall": "GE Healthcare Surgery announces a voluntary field action for the OEC Brivo 715 Prime, OEC Brivo 685 Essential, and OEC Brivo 865 Advance because the circuit board that controls vertical motion of the C-arm may fail, resulting in the C-Arm moving up or down without command.",
      "recall_initiation_date": "20160923",
      "center_classification_date": "20161011",
      "termination_date": "20170424",
      "report_date": "20161019",
      "code_info": "GE Brivo OEC715: Serial Numbers B2S15158 to B2S15164; B2S15166 to B2S15171; B2S15173 to B2S15195; B2S15197; B2S15200; B2S15202; B2S16046; B2S16050; B2S16051    GE Brivo OEC785: Serial Numbers B3S15236; B3S15241; B3S15242; B3S15246; B3S15248; B3S15250; B3S15253 to B3S15257; B3S15259 to B3S15338; B3S15353; B3S15357 to B3S15359; B3S16152    GE Brivo OEC865: Serial Numbers B4S15063 to B4S15070; B4S15072; B4S15073; B4S15075; B4S16029"
    }
  ]
}